Evaluation of 4 Point-of-Care Units for the Determination of Blood L-Lactate Concentration in Cattle

被引:27
|
作者
Karapinar, T. [1 ]
Kaynar, O.
Hayirli, A. [2 ]
Kom, M. [3 ]
机构
[1] Firat Univ, Fac Vet Med, Dept Internal Med, TR-23119 Elazig, Turkey
[2] Ataturk Univ, Dept Anim Nutr & Nutrit Disorders, Fac Vet Med, Erzurum, Turkey
[3] Firat Univ, Fac Vet Med, Dept Surg, TR-23169 Elazig, Turkey
关键词
Accuracy; Bovine; Portable lactate meter; Precision; Sensitivity; Specificity; PLASMA LACTATE; PROGNOSTIC VALUE; DAIRY-COWS; REGRESSION PROCEDURES; LEAST-SQUARES; WHOLE-BLOOD; DOGS; VOLVULUS; CALVES; ANALYZER;
D O I
10.1111/jvim.12204
中图分类号
S85 [动物医学(兽医学)];
学科分类号
0906 ;
摘要
BackgroundDespite being used commonly in bovine medicine, information on reliability of point-of-care (POC) lactate meters is limited. ObjectiveTo determine the validity of 4 commercially available POC lactate meters in cattle. AnimalsCattle with various diseases (n=120). MethodsBlood samples collected from the jugular vein were processed immediately on POC lactate meters. Plasma l-lactate concentration was measured by the enzymatic-colorimetric method (ELISA). Data were subjected to Friedman's test for comparison, Passing-Bablok regression and Bland-Altman plot analyses for reliability, and receiver operating characteristics analysis for sensitivity (Se) and specificity (Sp). ResultsThe POC lactate meters were highly correlated with ELISA (r=0.98-0.99) despite disagreements among units. When regressed on ELISA, blood l-lactate concentrations generated from Accutrend Plus and Lactate Pro were linear up to 16.6 and 15.7mmol/L, respectively, whereas those generated from i-STAT and Lactate Scout were linear up to 19.5 and 19.7mmol/L, respectively. All POC lactate meters had a Se of 100% with Sp of 95.7-98.6% at a plasma l-lactate cut-off concentration of 4mmol/L. i-STAT had the best accuracy (99.0%) and precision (99.8%), the best linear fit (y=-0.13+1.04X) yielding the lowest bias (-6.6%) as well as the highest Se (100%) and Sp (98.6%). Conclusions and Clinical ImportanceDespite high correlation with the reference method, dilution is needed for Accutrend Plus/Lactate Pro and i-STAT/Lactate Scout if concentrations >15 and 20mmol/L, respectively. i-STAT provided the most accurate and precise results.
引用
收藏
页码:1596 / 1603
页数:8
相关论文
共 50 条
  • [1] Reliability of the Lactate Scout point-of-care instrument for the determination of blood L-lactate concentration in sheep
    Kaynar, Ozgur
    Karapinar, Tolga
    Hayirli, Armagan
    Baydar, Ersoy
    VETERINARY CLINICAL PATHOLOGY, 2015, 44 (04) : 559 - 563
  • [2] Clinical Utilization of Point-of-Care Blood L-Lactate Concentrations in Naturally Occurring Respiratory Disease in Feedlot Cattle
    Zeineldin, Mohamed
    Ghanem, Mohamed
    Abd El-Raof, Yassein
    Elattar, Hossam
    PAKISTAN VETERINARY JOURNAL, 2017, 37 (02) : 210 - 214
  • [3] Validation of the handheld Lactate-Pro analyzer for measurement of blood L-lactate concentration in cattle
    Buczinski, Sebastien
    Dore, Elizabeth
    Boulay, Guillaume
    Francoz, David
    VETERINARY CLINICAL PATHOLOGY, 2014, 43 (04) : 567 - 572
  • [4] Evaluation of four point-of-care meters for rapid determination of blood lactate concentrations in dogs
    Acierno, Mark J.
    Mitchell, Mark A.
    JAVMA-JOURNAL OF THE AMERICAN VETERINARY MEDICAL ASSOCIATION, 2007, 230 (09): : 1315 - 1318
  • [5] FITC Embedded ZnO/Silica Nanocomposites as probe for detection of L-lactate: Point-of-Care diagnosis
    S. S. Joglekar
    P. V. Pimpliskar
    V. V. Sirdeshmukh
    P. S. Alegaonkar
    A. A. Kale
    MRS Advances, 2019, 4 : 2533 - 2540
  • [6] FITC Embedded ZnO/Silica Nanocomposites as probe for detection of L-lactate: Point-of-Care diagnosis
    Joglekar, S. S.
    Pimpliskar, P., V
    Sirdeshmukh, V. V.
    Alegaonkar, P. S.
    Kale, A. A.
    MRS ADVANCES, 2019, 4 (46-47) : 2533 - 2540
  • [7] Evaluation of an Electrochemical Point-of-Care Meter for Measuring Glucose Concentration in Blood from Periparturient Dairy Cattle
    Megahed, A. A.
    Hiew, M. W. H.
    Townsend, J. R.
    Messick, J. B.
    Constable, P. D.
    JOURNAL OF VETERINARY INTERNAL MEDICINE, 2015, 29 (06) : 1718 - 1727
  • [8] EVALUATION OF SERIAL BLOOD LACTATE AND THE USE OF A POINT-OF-CARE LACTATE METER IN LIVE-STRANDED PINNIPEDS
    Marinkovich, Matt
    Wack, Raymund F.
    Field, Cara L.
    Whoriskey, Sophie T.
    Kass, Philip H.
    Gjeltema, Jenessa
    JOURNAL OF ZOO AND WILDLIFE MEDICINE, 2019, 50 (01) : 137 - 146
  • [9] IMPROVED MICROMETHOD FOR DETERMINATION OF L-LACTATE C-14 IN BLOOD
    FREMINET, A
    BURSAUX, E
    POYART, CF
    BIOCHEMICAL MEDICINE, 1972, 6 (01): : 72 - &
  • [10] Evaluation of various methods of point-of-care testing of haemoglobin concentration in blood donors
    Singh, Abhay
    Dubey, Anju
    Sonker, Atul
    Chaudhary, Rajendra
    BLOOD TRANSFUSION, 2015, 13 (02) : 233 - 239