Evaluation of 4 Point-of-Care Units for the Determination of Blood L-Lactate Concentration in Cattle

被引:27
|
作者
Karapinar, T. [1 ]
Kaynar, O.
Hayirli, A. [2 ]
Kom, M. [3 ]
机构
[1] Firat Univ, Fac Vet Med, Dept Internal Med, TR-23119 Elazig, Turkey
[2] Ataturk Univ, Dept Anim Nutr & Nutrit Disorders, Fac Vet Med, Erzurum, Turkey
[3] Firat Univ, Fac Vet Med, Dept Surg, TR-23169 Elazig, Turkey
关键词
Accuracy; Bovine; Portable lactate meter; Precision; Sensitivity; Specificity; PLASMA LACTATE; PROGNOSTIC VALUE; DAIRY-COWS; REGRESSION PROCEDURES; LEAST-SQUARES; WHOLE-BLOOD; DOGS; VOLVULUS; CALVES; ANALYZER;
D O I
10.1111/jvim.12204
中图分类号
S85 [动物医学(兽医学)];
学科分类号
0906 ;
摘要
BackgroundDespite being used commonly in bovine medicine, information on reliability of point-of-care (POC) lactate meters is limited. ObjectiveTo determine the validity of 4 commercially available POC lactate meters in cattle. AnimalsCattle with various diseases (n=120). MethodsBlood samples collected from the jugular vein were processed immediately on POC lactate meters. Plasma l-lactate concentration was measured by the enzymatic-colorimetric method (ELISA). Data were subjected to Friedman's test for comparison, Passing-Bablok regression and Bland-Altman plot analyses for reliability, and receiver operating characteristics analysis for sensitivity (Se) and specificity (Sp). ResultsThe POC lactate meters were highly correlated with ELISA (r=0.98-0.99) despite disagreements among units. When regressed on ELISA, blood l-lactate concentrations generated from Accutrend Plus and Lactate Pro were linear up to 16.6 and 15.7mmol/L, respectively, whereas those generated from i-STAT and Lactate Scout were linear up to 19.5 and 19.7mmol/L, respectively. All POC lactate meters had a Se of 100% with Sp of 95.7-98.6% at a plasma l-lactate cut-off concentration of 4mmol/L. i-STAT had the best accuracy (99.0%) and precision (99.8%), the best linear fit (y=-0.13+1.04X) yielding the lowest bias (-6.6%) as well as the highest Se (100%) and Sp (98.6%). Conclusions and Clinical ImportanceDespite high correlation with the reference method, dilution is needed for Accutrend Plus/Lactate Pro and i-STAT/Lactate Scout if concentrations >15 and 20mmol/L, respectively. i-STAT provided the most accurate and precise results.
引用
收藏
页码:1596 / 1603
页数:8
相关论文
共 50 条
  • [41] Evaluation of a point-of-care blood glucose monitor in healthy goats
    Quandt, Jane E.
    Barletta, Michele
    Cornell, Karen K.
    Giguere, Steeve
    Hofmeister, Erik H.
    JOURNAL OF VETERINARY EMERGENCY AND CRITICAL CARE, 2018, 28 (01) : 45 - 53
  • [42] Reliability of a point-of-care blood gas analyzer for measuring lactate concentrations in cerebrospinal fluid
    Cardenas-Gamez, R.
    Perez-Canadas, P.
    Rodriguez-Rojas, C.
    Ramos-Arenas, V.
    Gonzalez-Morales, M.
    Garcia-Martin, E.
    Martinez-Gascon, L. E.
    Orgaz-Morales, T.
    Moreno-Fuentes, A.
    Albaladejo-Oton, M. D.
    Garcia De Guadiana-Romualdo, L.
    CLINICA CHIMICA ACTA, 2019, 493 : S769 - S769
  • [43] Identification of hemolytic anemia in Korean indigenous cattle with a criteria value of reticulocyte count, indirect bilirubin, and L-lactate concentration
    Jung, Youngwoo
    Ku, Ji-Yeong
    Kim, Youngjun
    Kim, Woochan
    So, Hyungjae
    Piao, Lianfu
    Yoon, Ji-Seon
    Park, Jinho
    FRONTIERS IN VETERINARY SCIENCE, 2024, 11
  • [44] Point-of-Care Ultrasound for Evaluation of Systolic Heart Function in Outpatient Hemodialysis Units
    Lohani, Sadichhya
    Annadanam, Surekha
    Martire, Carol
    Zakaria, Sammy
    Hwang, Seungyoung
    Jaar, Bernard G.
    Sozio, Stephen M.
    Shafi, Tariq
    KIDNEY MEDICINE, 2021, 3 (02) : 317 - 319
  • [45] ECONOMIC EVALUATION OF BLOOD GLUCOSE POINT-OF-CARE TESTING IN THE INTENSIVE CARE UNIT
    Steuten, L. M. G.
    Kip, M.
    Hooijdonk, R.
    Monteban, H.
    Spronk, P.
    Schultz, M.
    VALUE IN HEALTH, 2014, 17 (03) : A250 - A250
  • [46] Invasive and non-invasive point-of-care testing and point-of-care monitoring of the hemoglobin concentration in human blood - how accurate are the data?
    Dietzel, Frank
    Dieterich, Peter
    Doerries, Frank
    Gehring, Hartmut
    Wegerich, Philipp
    BIOMEDICAL ENGINEERING-BIOMEDIZINISCHE TECHNIK, 2019, 64 (05): : 495 - 506
  • [47] Characterization of the analytic performance of an electrochemical point-of-care meter for measuring -hydroxybutyrate concentration in blood and plasma from periparturient dairy cattle
    Megahed, Ameer A.
    Hiew, Mark W. H.
    Townsend, Jonathan R.
    Constable, Peter D.
    VETERINARY CLINICAL PATHOLOGY, 2017, 46 (02) : 314 - 325
  • [48] Clinical indicators of blood gas disturbances, elevated L-lactate concentration and other abnormal blood parameters in newborn beef calves
    Homerosky, E. R.
    Caulkett, N. A.
    Timsit, E.
    Pajor, E. A.
    Kastelic, J. P.
    Windeyer, M. C.
    VETERINARY JOURNAL, 2017, 219 : 49 - 57
  • [49] Blood L-Lactate Concentration as an Indicator of Outcome in Roe Deer (Capreolus capreolus) Admitted to a Wildlife Rescue Center
    Di Lorenzo, Elena
    Rossi, Riccardo
    Ferrari, Fabiana
    Martini, Valeria
    Comazzi, Stefano
    ANIMALS, 2020, 10 (06): : 1 - 8
  • [50] EVALUATION OF THREE POINT OF CARE LACTATE METHODS FOR ASSESSMENT OF WHOLE BLOOD LACTATE IN AN OBSTETRICS SETTING
    Orsonneau, J. L.
    Fraissinet, F.
    Bigot-Corbel, E. E.
    Sebille-Rivain, V.
    Dudouet, D.
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2011, 49 : S684 - S684