Evaluation of 4 Point-of-Care Units for the Determination of Blood L-Lactate Concentration in Cattle

被引:27
|
作者
Karapinar, T. [1 ]
Kaynar, O.
Hayirli, A. [2 ]
Kom, M. [3 ]
机构
[1] Firat Univ, Fac Vet Med, Dept Internal Med, TR-23119 Elazig, Turkey
[2] Ataturk Univ, Dept Anim Nutr & Nutrit Disorders, Fac Vet Med, Erzurum, Turkey
[3] Firat Univ, Fac Vet Med, Dept Surg, TR-23169 Elazig, Turkey
关键词
Accuracy; Bovine; Portable lactate meter; Precision; Sensitivity; Specificity; PLASMA LACTATE; PROGNOSTIC VALUE; DAIRY-COWS; REGRESSION PROCEDURES; LEAST-SQUARES; WHOLE-BLOOD; DOGS; VOLVULUS; CALVES; ANALYZER;
D O I
10.1111/jvim.12204
中图分类号
S85 [动物医学(兽医学)];
学科分类号
0906 ;
摘要
BackgroundDespite being used commonly in bovine medicine, information on reliability of point-of-care (POC) lactate meters is limited. ObjectiveTo determine the validity of 4 commercially available POC lactate meters in cattle. AnimalsCattle with various diseases (n=120). MethodsBlood samples collected from the jugular vein were processed immediately on POC lactate meters. Plasma l-lactate concentration was measured by the enzymatic-colorimetric method (ELISA). Data were subjected to Friedman's test for comparison, Passing-Bablok regression and Bland-Altman plot analyses for reliability, and receiver operating characteristics analysis for sensitivity (Se) and specificity (Sp). ResultsThe POC lactate meters were highly correlated with ELISA (r=0.98-0.99) despite disagreements among units. When regressed on ELISA, blood l-lactate concentrations generated from Accutrend Plus and Lactate Pro were linear up to 16.6 and 15.7mmol/L, respectively, whereas those generated from i-STAT and Lactate Scout were linear up to 19.5 and 19.7mmol/L, respectively. All POC lactate meters had a Se of 100% with Sp of 95.7-98.6% at a plasma l-lactate cut-off concentration of 4mmol/L. i-STAT had the best accuracy (99.0%) and precision (99.8%), the best linear fit (y=-0.13+1.04X) yielding the lowest bias (-6.6%) as well as the highest Se (100%) and Sp (98.6%). Conclusions and Clinical ImportanceDespite high correlation with the reference method, dilution is needed for Accutrend Plus/Lactate Pro and i-STAT/Lactate Scout if concentrations >15 and 20mmol/L, respectively. i-STAT provided the most accurate and precise results.
引用
收藏
页码:1596 / 1603
页数:8
相关论文
共 50 条
  • [31] Comparison of point-of-care and central laboratory analyzers for blood gas and lactate measurements
    Indrasari, Nuri Dyah
    Wonohutomo, Jessica Purwanti
    Sukartini, Ninik
    JOURNAL OF CLINICAL LABORATORY ANALYSIS, 2019, 33 (05)
  • [32] Technical note: Evaluation of a colorimetric point-of-care test for measuring urine ammonium concentration in periparturient dairy cattle
    Megahed, A. A.
    Constable, P. D.
    JOURNAL OF DAIRY SCIENCE, 2020, 103 (09) : 8655 - 8660
  • [33] Multicentre evaluation of a new point-of-care test for the determination of NT-proBNP in whole blood
    Zugck, Christian
    Nelles, Manfred
    Katus, Hugo A.
    Collinson, Paul O.
    Gaze, David C.
    Dikkeschei, Bert
    Gurr, Eberhard
    Hayen, Wiebke
    Haass, Markus
    Hechler, Christoph
    van Hoof, Viviane
    Guerti, Khadija
    van Waes, Carl
    Printzen, Gert
    Klopprogge, Kai
    Schulz, Ilse
    Zerback, Rainer
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2006, 44 (10) : 1269 - 1277
  • [34] Multicentre evaluation of a new point-of-care test for the determination of CK-MB in whole blood
    Giannitsis, Evangelos
    Baum, Hannsjoerg
    Bertsch, Thomas
    Juchum, Martin
    Mueller-Bardorff, Margit
    Jorgensen, Bo
    Boehmer, Agathe
    Rebmann, Martina
    Schaeffler, Juergen
    Schwab, Michael
    Zerback, Rainer
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2008, 46 (05) : 630 - 638
  • [35] Evaluation of a Point-of-Care Blood Analyzer for Use in Determination of Select Hematological Indices in the Seminole Killifish
    DiMaggio, Matthew A.
    Ohs, Cortney L.
    Petty, B. Denise
    NORTH AMERICAN JOURNAL OF AQUACULTURE, 2010, 72 (03) : 261 - 268
  • [36] Can we use a point-of-care blood gas analyzer to measure the lactate concentration in cerebrospinal fluid of patients with suspected meningitis?
    Rousseau, Geoffroy
    Asmolov, Romain
    Auvet, Adrien
    Grammatico-Guillon, Leslie
    Guillon, Antoine
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2018, 56 (09) : E247 - E248
  • [37] L-LACTATE ENZYME ELECTRODE OBTAINED WITH IMMOBILIZED RESPIRATORY-CHAIN FROM ESCHERICHIA-COLI AND OXYGEN PROBE FOR SPECIFIC DETERMINATION OF L-LACTATE IN YOGURT, WINE AND BLOOD
    ADAMOWICZ, E
    BURSTEIN, C
    BIOSENSORS, 1987, 3 (01) : 27 - 43
  • [38] Performance evaluation of the epoc® point-of-care blood analysis system
    Chen, Yu
    Gorman, Monique
    O'Reilly, Bill
    CLINICAL BIOCHEMISTRY, 2015, 48 (15) : 1013 - 1013
  • [39] A Photochemical Dongle for Point-of-Care Evaluation of Blood Lipid Level
    Huang, Xiwei
    Guo, Jinhong
    2017 IEEE BIOMEDICAL CIRCUITS AND SYSTEMS CONFERENCE (BIOCAS), 2017,
  • [40] Evaluation of a whole blood point-of-care coagulation analyzer in dogs
    Conroy, Elizabeth M.
    Lyons, Bridget M.
    Koenig, Amie
    JOURNAL OF VETERINARY EMERGENCY AND CRITICAL CARE, 2024, 34 (05) : 446 - 454