Evaluation of various methods of point-of-care testing of haemoglobin concentration in blood donors

被引:29
|
作者
Singh, Abhay [1 ]
Dubey, Anju [2 ]
Sonker, Atul [1 ]
Chaudhary, Rajendra [1 ]
机构
[1] Sanjay Gandhi Post Grad Inst Med Sci, Dept Transfus Med, Lucknow, Uttar Pradesh, India
[2] All India Inst Med Sci, Dept Transfus Med, Rishikesh 249201, Uttarakhand, India
关键词
point-of-care testing; haemoglobin; blood donors; non-invasive; NONINVASIVE HEMOGLOBIN; ACCURACY; VALIDATION; ANALYZER; HEMOCUE;
D O I
10.2450/2014.0085-14
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background. Point-of-care testing (POCT) devices for determining pre-donation haemoglobin (Hb) concentrations mark the advent of advanced technology for blood banks. POCT devices have undergone several improvements including changes in testing methodology and Size of device, befitting the needs of blood donors and blood banks in terms of safety and quality of blood components. This study was planned to evaluate the suitability of non-invasive and invasive POCT devices for blood donor Hb screening. Material and methods. Pre-donation Hb in apparently healthy blood donors was measured by a non-invasive spectrophotometric based method (NBM-200, OrSense) and an invasive method utilizing reagent free cuvettes (DiaSpect) along with a device using sodium azide-coated cuvettes (HemoControl, EKF diagnostic GmbH). The performance of the devices was evaluated by comparison with the reference method, i.e. an automated cell counter (KX-21). Results. Hb was measured in 485 prospective blood donors. DiaSpect hemoglobin T system was found to be the most sensitive method of POCT for Hb (sensitivity 98.1%) followed by HemoControl (sensitivity 86.8%). NBM-200 was the least sensitive method (sensitivity 71.7%). The intraclass correlation coefficient was highest for DiaSpect (0.78), followed by HemoControl (0.77) and NBM-200 (0.43). The variation of results on repeat testing was high for NBM-200 with a coefficient of variation of 4.28%, compared to 2.19% for DiaSpect. On comparing the mean testing time, DiaSpect (1.9 seconds) was found to be significantly quicker than the other two POCT devices (p<0.001). Discussion. NBM-200 has the apparent advantage of eliminating pain but also a substantial possibility of causing ineligible donors to be accepted. DiaSpect was fast and accurate, with its results showing perfect agreement with those of the standard method. It is, therefore, aptly suited for screening donors in blood banks
引用
收藏
页码:233 / 239
页数:7
相关论文
共 50 条
  • [1] Haemoglobin point-of-care testing
    Scicluna, M.
    Wirth, F.
    Azzopardi, L. M.
    [J]. INTERNATIONAL JOURNAL OF CLINICAL PHARMACY, 2020, 42 (01) : 239 - 239
  • [2] Portable point-of-care device for haemoglobin testing
    Mohan, Alladi
    [J]. NATIONAL MEDICAL JOURNAL OF INDIA, 2014, 27 (03): : 186 - 187
  • [3] Evaluation of two devices for point-of-care testing of haemoglobin in neonatal pigs
    Kutter, Annette P. N.
    Mauch, Jacqueline Y.
    Riond, Barbara
    Martin-Jurado, Olga
    Spielmann, Nelly
    Weiss, Markus
    Bettschart-Wolfensberger, Regula
    [J]. LABORATORY ANIMALS, 2012, 46 (01) : 65 - 70
  • [4] Evaluation and quality assessment of glucose concentration measurement in blood by point-of-care testing devices
    Wehmeier, Michael
    Arndt, Burkhard T.
    Schumann, Gerhard
    Kulpmann, Wolf R.
    [J]. CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2006, 44 (07) : 888 - 893
  • [5] Point-of-care estimation of haemoglobin concentration in neonates and infants
    Schapkaitz, Elise
    Mahlangu, Johnny Ndoni
    Letsoalo, Matime
    [J]. SOUTH AFRICAN JOURNAL OF CHILD HEALTH, 2012, 6 (01) : 10 - 11
  • [6] Point-of-Care Blood Testing in Severe Mental Illness: A Mixed-Methods Evaluation
    Lyman, Monty
    Fanshawe, Jack
    Brewin, Josh
    Fanshawe, Thomas
    Turner, Philip
    [J]. BJPSYCH OPEN, 2024, 10 : S58 - S59
  • [7] Point-of-Care Blood Testing in Severe Mental Illness: A Mixed-Methods Evaluation
    Lyman, Monty
    Fanshawe, Jack
    Brewin, Josh
    Fanshawe, Thomas
    Turner, Philip
    [J]. BJPSYCH OPEN, 2024, 10 : S58 - S59
  • [8] Point-of-care LeadCarea® blood lead testing system:: Potential utility for lead testing of blood donors
    Criss, V. R.
    Gallagher, M.
    Mattero, A.
    Soldin, S.
    Luban, N. L.
    Wong, E. C.
    [J]. TRANSFUSION, 2006, 46 (09) : 121A - 121A
  • [9] Capillary blood for point-of-care testing
    Tang, Ruihua
    Yang, Hui
    Choi, Jane Ru
    Gong, Yan
    You, MinLi
    Wen, Ting
    Li, Ang
    Li, XiuJun
    Xu, Bo
    Zhang, Sufeng
    Mei, Qibing
    Xu, Feng
    [J]. CRITICAL REVIEWS IN CLINICAL LABORATORY SCIENCES, 2017, 54 (05) : 294 - 308
  • [10] Portable liquid chromatography for point-of-care testing of glycated haemoglobin
    Li, Ding
    Chen, Haiyan
    Ren, Shuang
    Zhang, Yanfang
    Yang, Yang
    Chang, Honglong
    [J]. SENSORS AND ACTUATORS B-CHEMICAL, 2020, 305