A novel approach to EPID-based 3D volumetric dosimetry for IMRT and VMAT QA

被引:22
|
作者
Alhazmi, Abdulaziz [1 ]
Gianoli, Chiara [1 ]
Neppl, Sebastian [2 ]
Martins, Juliana [1 ]
Veloza, Stella [1 ]
Podesta, Mark [3 ]
Verhaegen, Frank [3 ]
Reiner, Michael [2 ]
Belka, Claus [2 ]
Parodi, Katia [1 ]
机构
[1] Ludwig Maximilians Univ Munchen, Fac Phys, Dept Med Phys, Munich, Germany
[2] Univ Hosp, Dept Radiat Oncol, LMU Munich, Munich, Germany
[3] Maastricht Univ, GROW Sch Oncol & Dev Biol, Dept Radiat Oncol MAASTRO, Med Ctr, Maastricht, Netherlands
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2018年 / 63卷 / 11期
关键词
EPID dosimetry; pre-treatment QA; dose reconstruction; VMAT; IMRT; 2D DIODE-ARRAY; QUALITY-ASSURANCE; VERIFICATION; POSITION; MODE; FILM;
D O I
10.1088/1361-6560/aac1a6
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Intensity modulated radiation therapy (IMRT) and volumetric modulated arc therapy (VMAT) are relatively complex treatment delivery techniques and require quality assurance (QA) procedures. Pre-treatment dosimetric verification represents a fundamental QA procedure in daily clinical routine in radiation therapy. The purpose of this study is to develop an EPID-based approach to reconstruct a 3D dose distribution as imparted to a virtual cylindrical water phantom to be used for plan-specific pre-treatment dosimetric verification for IMRT and VMAT plans. For each depth, the planar 2D dose distributions acquired in air were back-projected and convolved by depth-specific scatter and attenuation kernels. The kernels were obtained by making use of scatter and attenuation models to iteratively estimate the parameters from a set of reference measurements. The derived parameters served as a look-up table for reconstruction of arbitrary measurements. The summation of the reconstructed 3D dose distributions resulted in the integrated 3D dose distribution of the treatment delivery. The accuracy of the proposed approach was validated in clinical IMRT and VMAT plans by means of gamma evaluation, comparing the reconstructed 3D dose distributions with Octavius measurement. The comparison was carried out using (3% , 3 mm) criteria scoring 99% and 96% passing rates for IMRT and VMAT, respectively. An accuracy comparable to the one of the commercial device for 3D volumetric dosimetry was demonstrated. In addition, five IMRT and five VMAT were validated against the 3D dose calculation performed by the TPS in a water phantom using the same passing rate criteria. The median passing rates within the ten treatment plans was 97.3%, whereas the lowest was 95%. Besides, the reconstructed 3D distribution is obtained without predictions relying on forward dose calculation and without external phantom or dosimetric devices. Thus, the approach provides a fully automated, fast and easy QA procedure for plan-specific pretreatment dosimetric verification.
引用
收藏
页数:13
相关论文
共 50 条
  • [21] EPID-based 3D in vivo dosimetry for SBRT lung and spine - Values and Challenges
    Van Uytven, E.
    McCowan, P.
    VanBeek, T.
    McCurdy, B. M.
    RADIOTHERAPY AND ONCOLOGY, 2018, 127 : S350 - S351
  • [22] 2D/3D EPID-based in-vivo dosimetry: Preliminary results
    Camilleri, J.
    Mazurier, J.
    Franck, D.
    Ducassou, M.
    Gallocher, O.
    Latorzeff, I.
    Marre, D.
    Mathy, N.
    Zarate, D.
    Franceries, X.
    RADIOTHERAPY AND ONCOLOGY, 2014, 111 : S167 - S167
  • [23] Towards real-time EPID-based 3D in vivo dosimetry for IMRT with Deep Neural Networks: A feasibility study
    Martins, Juliana Cristina
    Maier, Joscha
    Gianoli, Chiara
    Neppl, Sebastian
    Dedes, George
    Alhazmi, Abdulaziz
    Veloza, Stella
    Reiner, Michael
    Belka, Claus
    Kachelriess, Marc
    Parodi, Katia
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2023, 114
  • [24] Current status of 3D EPID-based in vivo dosimetry in The Netherlands Cancer Institute
    Mijnheer, B.
    Olaciregui-Ruiz, I.
    Rozendaal, R.
    Spreeuw, H.
    van Herk, M.
    Mans, A.
    8TH INTERNATIONAL CONFERENCE ON 3D RADIATION DOSIMETRY (IC3DDOSE), 2015, 573
  • [25] Establishing Institution-Specific Action Levels for EPID-Based QA for IMRT
    Gray, A.
    Saldan, A.
    Rosenzweig, D.
    Meiler, R.
    Clark, D.
    Abu-Aita, R.
    Schell, M.
    Cavanaugh, D.
    MEDICAL PHYSICS, 2013, 40 (06)
  • [26] Characterization of a new EPID-based system for in-vivo dosimetry in VMAT treatments
    Bresciani, S.
    Poli, M.
    Miranti, A.
    Maggio, A.
    Di Dia, A.
    Bracco, C.
    Stasi, M.
    RADIOTHERAPY AND ONCOLOGY, 2016, 119 : S736 - S737
  • [27] Epid-based in vivo dosimetry for SBRT-VMAT treatment dose verification
    Cilla, S.
    Ianiro, A.
    Craus, M.
    Viola, P.
    Fidanzio, A.
    Azario, L.
    Greco, F.
    Grusio, M.
    Deodato, F.
    Macchia, G.
    Valentini, V.
    Morganti, A.
    Piermattei, A.
    RADIOTHERAPY AND ONCOLOGY, 2017, 123 : S772 - S773
  • [28] Planar EPID-Based Dosimetry for SRS and SRT Patient-Specific QA
    Thongsawad, Sangutid
    Chanton, Tadchapong
    Saiyo, Nipon
    Udee, Nuntawat
    LIFE-BASEL, 2021, 11 (11):
  • [29] An EPID-based 3D patient dose verification method for SBRT-VMAT delivery
    McCowan, P.
    vanUytven, E.
    vanBeek, T.
    McCurdy, B.
    MEDICAL PHYSICS, 2014, 41 (08) : 2 - 2
  • [30] A novel approach to SBRT patient quality assurance using EPID-based real-time transit dosimetryA step to QA with in vivo EPID dosimetry
    Christos Moustakis
    Fatemeh Ebrahimi Tazehmahalleh
    Khaled Elsayad
    Francis Fezeu
    Sergiu Scobioala
    Strahlentherapie und Onkologie, 2020, 196 : 182 - 192