Extraction of pacemaker and implantable cardioverter defibrillator leads: Patient and lead characteristics in relation to the requirement of extraction tools

被引:73
|
作者
Bracke, F [1 ]
Meijer, A [1 ]
Van Gelder, B [1 ]
机构
[1] Catharina Hosp, Dept Cardiol, NL-5602 ZA Eindhoven, Netherlands
来源
关键词
lead extraction; technique; laser; pacemaker; implantable defibrillator;
D O I
10.1046/j.1460-9592.2002.01037.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Effective tools for extraction of pacemaker and ICD leads have been developed in the past decennium. This study investigated the necessity of using these tools in addition to direct traction in relation to patient and lead characteristics. The study encompasses first attempts at extraction of consecutive pacemaker and ICD leads from the subpectoral area. A stepwise extraction protocol was used with traction first (directly or with a locking stylet) followed by laser sheath extraction if not successful, The indication, patient age, time from implant, fixation mechanism, location, and insertion site of the leads were studied in relation to the outcome of traction. A total of 145 leads in 83 patients were extracted. Leads were implanted for 71 +/- 61 months. Indication for extraction was infection in 96 leads and malfunction in 49 leads. There were 90 ventricular leads including 16 ICD leads. Forty-nine (34%) leads were extracted with traction; in 96 (66%) leads a laser sheath was necessary. All leads implanted for < 6 months could be removed with traction alone. In a multivariate logistic regression model, time from implant was the main factor determining success of traction (P < 0.001), but in case of infection the success rate increased (P = 0.004). In conclusion, time from implant is the decisive factor to judge the potential efficacy of lead extraction with direct traction, If leads are implanted for 6 months, the availability of additional extraction tools is necessary when lead extraction is considered. In addition to time from implant, infected leads have a better chance to be removed with traction although it is a much weaker predictor.
引用
收藏
页码:1037 / 1040
页数:4
相关论文
共 50 条
  • [21] Pacemaker and defibrillator lead extraction
    Kutalek, SP
    CURRENT OPINION IN CARDIOLOGY, 2004, 19 (01) : 19 - 22
  • [22] Automated Implantable Cardioverter Defibrillator/Pacemaker Lead Fracture
    Rubero, Lose
    Ramlatchan, Sidhartha R.
    Solomon, Matthew
    George, Andrew
    Ganti, Latha
    CUREUS JOURNAL OF MEDICAL SCIENCE, 2020, 12 (11)
  • [23] Impact of coils and fixation mechanism on implantable cardioverter defibrillator lead extraction
    Segreti, L.
    Bongiorni, M. G.
    Di Cori, A.
    Zucchelli, G.
    Viani, S.
    Paperini, L.
    De Lucia, R.
    Coluccia, G.
    Romano, S. L.
    Soldati, E.
    EUROPEAN HEART JOURNAL, 2014, 35 : 375 - 375
  • [24] Prophylactic Lead Extraction at Implantable Cardioverter-Defibrillator Generator Change
    Mendenhall, G. Stuart
    Saba, Samir
    CIRCULATION-ARRHYTHMIA AND ELECTROPHYSIOLOGY, 2014, 7 (02): : 330 - 336
  • [25] Percutaneous extraction of pacemaker and defibrillator leads
    Centella, Tomasa
    Oliva, Enrique
    Garcia-Andrade, Ignacio
    Martin-Davila, Pilar
    Cobo, Javier
    Moya, Jose L.
    Hernandez-Madrid, Antonio
    Epeldegui, Antonio
    REVISTA ESPANOLA DE CARDIOLOGIA, 2007, 60 (06): : 607 - 615
  • [26] Deaths and cardiovascular injuries due to device-assisted implantable cardioverter-defibrillator and pacemaker lead extraction
    Hauser, Robert G.
    Katsiyiannis, William T.
    Gornick, Charles C.
    Almquist, Adrian K.
    Kallinen, Linda M.
    EUROPACE, 2010, 12 (03): : 395 - 401
  • [27] Initial experience of pacemaker and implantable cardioverter defibrillator lead extraction with the new GlideLight 80 Hz laser sheaths
    Hakmi, Samer
    Pecha, Simon
    Sill, Bjoern
    Reiter, Beate
    Willems, Stephan
    Aydin, Muhammet Ali
    Yildirim, Yalin
    Reichenspurner, Hermann
    Treede, Hendrik
    INTERACTIVE CARDIOVASCULAR AND THORACIC SURGERY, 2014, 18 (01) : 56 - 60
  • [28] From lead management to implanted patient management: indications to lead extraction in pacemaker and cardioverter-defibrillator systems
    Diemberger, Igor
    Biffi, Mauro
    Martignani, Cristian
    Boriani, Giuseppe
    EXPERT REVIEW OF MEDICAL DEVICES, 2011, 8 (02) : 235 - 255
  • [29] Initial experience using Excimer laser for the extraction of chronically implanted pacemaker and implantable cardioverter defibrillator leads in Japanese patients
    Okamura, Hideo
    Yasuda, Satoshi
    Sato, Shunsuke
    Ogawa, Koji
    Nakajima, Ikutaro
    Noda, Takashi
    Shimahara, Yusuke
    Hayashi, Teruyuki
    Onishi, Yoshihiko
    Kobayashi, Junjiro
    Kamakura, Shiro
    Ogawa, Hisao
    Shimizu, Wataru
    JOURNAL OF CARDIOLOGY, 2013, 62 (3-4) : 195 - 200
  • [30] Yes we can! But should we? Lead extraction for superfluous pacemaker and implanted cardioverter-defibrillator leads
    Bracke, Frank A.
    EUROPACE, 2009, 11 (05): : 546 - 547