Calorimeter for Real-Time Dosimetry of Pulsed Ultra-High Dose Rate Electron Beams

被引:25
|
作者
Bourgouin, Alexandra [1 ,2 ,3 ]
Schueller, Andreas [3 ]
Hackel, Thomas [3 ]
Kranzer, Rafael [4 ,5 ]
Poppinga, Daniela [5 ]
Kapsch, Ralf-Peter [3 ]
McEwen, Malcolm [1 ]
机构
[1] Natl Res Council Canada, Ionizing Radiat Stand, Ottawa, ON, Canada
[2] Carleton Univ, Dept Phys, Ottawa, ON, Canada
[3] Phys Tech Bundesanstalt PTB, Braunschweig, Germany
[4] Carl von Ossietzky Univ Oldenburg, Univ Clin Med Radiat Phys, Pius Hosp, Med Campus, Oldenburg, Germany
[5] Phys Tech Werkstatten Freibug, Freiburg, Germany
基金
欧盟地平线“2020”;
关键词
FLASH; dosimetry; ultra-high dose per pulse; calorimeter; alanine; ionization chamber; ion recombination; HIGH-ENERGY PHOTON; IONIZATION CHAMBERS; RECOMBINATION CORRECTION; GRAPHITE CALORIMETER; LINEAR-ACCELERATOR; ALANINE DOSIMETERS; ION RECOMBINATION; RELATIVE RESPONSE; RADIOTHERAPY; CALIBRATION;
D O I
10.3389/fphy.2020.567340
中图分类号
O4 [物理学];
学科分类号
0702 ;
摘要
An aluminum calorimeter was investigated as a possible real-time dosimeter for electron beams with an ultra-high dose per pulse (DPP), as used in FLASH radiation therapy (a few Gy/pulse). Ionization chambers, the most widely used active dosimeter type in conventional external beam radiation therapy, suffer from large ion recombination losses at these conditions. Passive dosimeters, such as alanine, are independent of dose rate but do not provide real-time read-out. In this work it is shown that the response of alanine is independent of the DPP in the investigated ultra-high DPP range (up to 2.3 Gy/pulse). Alanine dose measurements were then used to determine the ion recombination correction for an Advanced Markus plane-parallel ionization chamber at ultra-high DPP. Ion collection losses larger than 50% were observed. Therefore, ionization chambers are not considered suitable for accurate dosimetry in FLASH radiation therapy. As an alternative, in a second (independent) experiment an aluminum open-to-atmosphere calorimeter, operated in the quasi-adiabatic mode was investigated at ultra-high DPP electron radiation. The beam pulse charge, and thus the DPP, was varied to evaluate the linearity of the calorimeter response in the DPP range between 0.3 and 1.8 Gy/pulse. On average, the standard deviation of the calorimeter response was 0.1%. The response was proportional to the DPP in the investigated range. The average deviation of the linear fit of the calorimeter dose as a function of the beam pulse charge was <0.5%. This preliminary investigation suggests that a simplified calorimeter design is suitable as a dosimeter with real-time read-out for clinical FLASH radiation therapy beams.
引用
收藏
页数:10
相关论文
共 50 条
  • [1] Absolute Dosimetry of Ultra-High Dose Rate Pulsed Electron Beams Using a Graphite Probe Calorimeter
    Keszti, F.
    Bourgouin, A.
    Schueller, A.
    Simon, W.
    Schoenfeld, A.
    Seuntjens, J.
    Renaud, J.
    [J]. MEDICAL PHYSICS, 2021, 48 (06)
  • [2] Investigation of a Simple Calorimeter Design for the Dosimetry of Ultra-High Pulse Dose Rate (FLASH) Electron Beams
    McEwen, M.
    Bourgouin, A.
    Schueller, A.
    [J]. MEDICAL PHYSICS, 2020, 47 (06) : E294 - E294
  • [3] Real-time monitoring of ultra-high dose rate electron beams using bremsstrahlung photons
    Kim, Hyun
    Jeong, Dong Hyeok
    Kang, Sang Koo
    Lee, Manwoo
    Lim, Heuijin
    Lee, Sang Jin
    Jang, Kyoung Won
    [J]. NUCLEAR ENGINEERING AND TECHNOLOGY, 2023, 55 (09) : 3417 - 3422
  • [4] The PTB water calorimeter for determining the absolute absorbed dose to water in ultra-high pulse dose rate electron beams
    Bourgouin, Alexandra
    Hackel, Thomas
    Kapsch, Ralf-Peter
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2023, 68 (11):
  • [5] Dosimetry of ultra-high dose rate electron beams using thermoluminescence and optically stimulated luminescence detectors
    Motta, S.
    Dal Bello, R.
    Christensen, J. B.
    Bossin, L.
    Yukihara, E. G.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2024, 69 (03):
  • [6] Radiobiological response to ultra-short pulsed megavoltage electron beams of ultra-high pulse dose rate
    Beyreuther, Elke
    Karsch, Leonhard
    Laschinsky, Lydia
    Lessmann, Elisabeth
    Naumburger, Doreen
    Oppelt, Melanie
    Richter, Christian
    Schuerer, Michael
    Woithe, Julia
    Pawelke, Joerg
    [J]. INTERNATIONAL JOURNAL OF RADIATION BIOLOGY, 2015, 91 (08) : 643 - 652
  • [7] Silicon carbide detectors for dosimetry and monitoring of ultra-high dose rate beams
    Okpuwe, C.
    Amato, A.
    D'Amico, I.
    De Liso, V.
    De Napoli, M.
    Di Martino, F.
    Felici, G.
    Galluzzo, L.
    Medina, E.
    Vignati, A.
    Camarda, M.
    Romano, F.
    Milluzzo, G.
    [J]. JOURNAL OF INSTRUMENTATION, 2024, 19 (03)
  • [8] A prototype scintillator real-time beam monitor for ultra-high dose rate radiotherapy
    Levin, Daniel S.
    Friedman, Peter S.
    Ferretti, Claudio
    Ristow, Nicholas
    Tecchio, Monica
    Litzenberg, Dale W.
    Bashkirov, Vladimir
    Schulte, Reinhard
    [J]. MEDICAL PHYSICS, 2024, 51 (04) : 2905 - 2923
  • [9] Investigation of TL and OSL detectors in ultra-high dose rate electron beams
    Motta, S.
    Christensen, J. B.
    Frei, F.
    Peier, P.
    Yukihara, E. G.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2023, 68 (14):
  • [10] The challenge of ionisation chamber dosimetry in ultra-short pulsed high dose-rate Very High Energy Electron beams
    McManus, M.
    Romano, F.
    Lee, N. D.
    Farabolini, W.
    Gilardi, A.
    Royle, G.
    Palmans, H.
    Subiel, A.
    [J]. SCIENTIFIC REPORTS, 2020, 10 (01)