Calorimeter for Real-Time Dosimetry of Pulsed Ultra-High Dose Rate Electron Beams

被引:25
|
作者
Bourgouin, Alexandra [1 ,2 ,3 ]
Schueller, Andreas [3 ]
Hackel, Thomas [3 ]
Kranzer, Rafael [4 ,5 ]
Poppinga, Daniela [5 ]
Kapsch, Ralf-Peter [3 ]
McEwen, Malcolm [1 ]
机构
[1] Natl Res Council Canada, Ionizing Radiat Stand, Ottawa, ON, Canada
[2] Carleton Univ, Dept Phys, Ottawa, ON, Canada
[3] Phys Tech Bundesanstalt PTB, Braunschweig, Germany
[4] Carl von Ossietzky Univ Oldenburg, Univ Clin Med Radiat Phys, Pius Hosp, Med Campus, Oldenburg, Germany
[5] Phys Tech Werkstatten Freibug, Freiburg, Germany
基金
欧盟地平线“2020”;
关键词
FLASH; dosimetry; ultra-high dose per pulse; calorimeter; alanine; ionization chamber; ion recombination; HIGH-ENERGY PHOTON; IONIZATION CHAMBERS; RECOMBINATION CORRECTION; GRAPHITE CALORIMETER; LINEAR-ACCELERATOR; ALANINE DOSIMETERS; ION RECOMBINATION; RELATIVE RESPONSE; RADIOTHERAPY; CALIBRATION;
D O I
10.3389/fphy.2020.567340
中图分类号
O4 [物理学];
学科分类号
0702 ;
摘要
An aluminum calorimeter was investigated as a possible real-time dosimeter for electron beams with an ultra-high dose per pulse (DPP), as used in FLASH radiation therapy (a few Gy/pulse). Ionization chambers, the most widely used active dosimeter type in conventional external beam radiation therapy, suffer from large ion recombination losses at these conditions. Passive dosimeters, such as alanine, are independent of dose rate but do not provide real-time read-out. In this work it is shown that the response of alanine is independent of the DPP in the investigated ultra-high DPP range (up to 2.3 Gy/pulse). Alanine dose measurements were then used to determine the ion recombination correction for an Advanced Markus plane-parallel ionization chamber at ultra-high DPP. Ion collection losses larger than 50% were observed. Therefore, ionization chambers are not considered suitable for accurate dosimetry in FLASH radiation therapy. As an alternative, in a second (independent) experiment an aluminum open-to-atmosphere calorimeter, operated in the quasi-adiabatic mode was investigated at ultra-high DPP electron radiation. The beam pulse charge, and thus the DPP, was varied to evaluate the linearity of the calorimeter response in the DPP range between 0.3 and 1.8 Gy/pulse. On average, the standard deviation of the calorimeter response was 0.1%. The response was proportional to the DPP in the investigated range. The average deviation of the linear fit of the calorimeter dose as a function of the beam pulse charge was <0.5%. This preliminary investigation suggests that a simplified calorimeter design is suitable as a dosimeter with real-time read-out for clinical FLASH radiation therapy beams.
引用
收藏
页数:10
相关论文
共 50 条
  • [41] A GPU optimization workflow for real-time execution of ultra-high frame rate computer vision applications
    Nourazar, Mohsen
    Booth, Brian G.
    Goossens, Bart
    [J]. JOURNAL OF REAL-TIME IMAGE PROCESSING, 2024, 21 (01)
  • [42] An Ultra-High Frame Rate Ion Imaging Platform Using ISFET Arrays With Real-Time Compression
    Zeng, Junming
    Kuang, Lei
    Cacho-Soblechero, Miguel
    Georgiou, Pantelis
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL CIRCUITS AND SYSTEMS, 2021, 15 (04) : 820 - 833
  • [43] Characterization of the PTB ultra-high pulse dose rate reference electron beam
    Bourgouin, Alexandra
    Knyziak, Adrian
    Marinelli, Marco
    Kranzer, Rafael
    Schueller, Andreas
    Kapsch, Ralf-Peter
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2022, 67 (08):
  • [44] Radiation safety of ultra-high dose rate electron accelerators for FLASH radiotherapy
    Praestegaard, Lars Hjorth
    [J]. MEDICAL PHYSICS, 2024, : 6206 - 6219
  • [45] Characterization of Ultra-High-Dose Rate Electron Beams with ElectronFlash Linac
    Giuliano, Lucia
    Franciosini, Gaia
    Palumbo, Luigi
    Aggar, Lilia
    Dutreix, Marie
    Faillace, Luigi
    Favaudon, Vincent
    Felici, Giuseppe
    Galante, Federica
    Mostacci, Andrea
    Migliorati, Mauro
    Pacitti, Matteo
    Patriarca, Annalisa
    Heinrich, Sophie
    [J]. APPLIED SCIENCES-BASEL, 2023, 13 (01):
  • [46] Diamond detectors for dose and instantaneous dose-rate measurements for ultra-high dose-rate scanned helium ion beams
    Tessonnier, Thomas
    Verona-Rinati, Gianluca
    Rank, Luisa
    Kranzer, Rafael
    Mairani, Andrea
    Marinelli, Marco
    [J]. MEDICAL PHYSICS, 2024, 51 (02) : 1450 - 1459
  • [47] A dose rate independent 2D Ce-doped YAG scintillating dosimetry system for time resolved beam monitoring in ultra-high dose rate electron "FLASH"radiation therapy
    Vanreusel, Verdi
    Heinrich, Sophie
    De Kerf, Thomas
    Leblans, Paul
    Vandenbroucke, Dirk
    Vanlanduit, Steve
    Verellen, Dirk
    Gasparini, Alessia
    Nascimento, Luana de Freitas
    [J]. SENSORS AND ACTUATORS A-PHYSICAL, 2024, 371
  • [48] Response of diamond detectors in ultra-high dose-per-pulse electron beams for dosimetry at FLASH radiotherapy (vol 67, 075002, 2022)
    Kranzer, R.
    Schueller, A.
    Bourgouin, A.
    Hackel, T.
    Poppinga, D.
    Lapp, M.
    Looe, H. K.
    Poppe, B.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2022, 67 (14):
  • [49] Dosimetry of Ultra-Short High Dose-Per-Pulse Very High Energy Electron Beams
    McManus, M.
    Romano, F.
    Lee, N.
    Wilfrid, F.
    Antonio, G.
    Royle, Gary
    Palmans, H.
    Subiel, A.
    [J]. MEDICAL PHYSICS, 2020, 47 (06) : E372 - E372
  • [50] DOSIMETRY FOR THIN BIOLOGICAL SAMPLES IRRADIATED AT ULTRA-HIGH DOSE RATES BY ELECTRON PULSES OF NANOSECOND DURATIONS
    MCDONALD, J
    PINKERTON, A
    WEISS, H
    EPP, ER
    [J]. RADIATION RESEARCH, 1971, 47 (01) : 353 - +