The PTB water calorimeter for determining the absolute absorbed dose to water in ultra-high pulse dose rate electron beams

被引:4
|
作者
Bourgouin, Alexandra [1 ]
Hackel, Thomas [1 ]
Kapsch, Ralf-Peter [1 ]
机构
[1] Phys Tech Bundesanstalt PTB, Dosimetry Radiat Therapy & Diagnost Radiol, D-38116 Braunschweig, Germany
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2023年 / 68卷 / 11期
关键词
water calorimetry; primary standard; ultra-high dose per pulse; electron beam; flash dosimetry; DOSIMETRY;
D O I
10.1088/1361-6560/acce1d
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Objective. The purpose of this investigation is to establish the water calorimeter as the primary standard in PTB's ultra-high pulse dose rate (UHPDR) 20 MeV reference electron beams. Approach. The calorimetric measurements were performed at the PTB research linac facility using the UHPDR reference electron beam setups that enable a dose per pulse between about 0.1 Gy and 6 Gy. The beam is monitored by an in-flange integrating current transformer. The correction factors required to determine the absorbed dose to water were evaluated using thermal and Monte Carlo simulations. Measurements were performed with different total doses delivered per pulse by modifying the instantaneous dose rate within a pulse and by changing the pulse length. The obtained temperature-time traces were compared to the simulated ones to validate the thermal simulations. In addition, absorbed-dose-to-water measurements obtained using the secondary standard alanine dosimeter system were compared to measurements performed with the primary standard. Main results. The simulated and measured temperature-time traces were shown to be consistent, within combined uncertainties, with one another. Measurements with alanine dosimeters proved to be consistent within k = 1 of the total combined uncertainty with the absorbed dose to water determined using the primary standard. Significance. The total relative standard uncertainty of absorbed dose to water determined using the PTB water calorimeter primary standard in UHPDR electron beams was estimated to be less than 0.5%, and the combined correction factors were found to deviate from 1 by less than 1% for both PTB UHPDR 20 MeV reference electron beams. The water calorimeter is therefore considered to be an established primary standard for the higher energy UHPDR reference electron beams.
引用
收藏
页数:15
相关论文
共 50 条
  • [1] Absorbed-dose-to-water measurement using alanine in ultra-high-pulse-dose-rate electron beams
    Bourgouin, Alexandra
    Hackel, Thomas
    Marinelli, Marco
    Kranzer, Rafael
    Schueller, Andreas
    Kapsch, Ralf-Peter
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2022, 67 (20):
  • [2] The PTB water calorimeter for the absolute determination of absorbed dose to water in 60Co radiation
    Krauss, A
    [J]. METROLOGIA, 2006, 43 (03) : 259 - 272
  • [3] Absolute Dosimetry of Ultra-High Dose Rate Pulsed Electron Beams Using a Graphite Probe Calorimeter
    Keszti, F.
    Bourgouin, A.
    Schueller, A.
    Simon, W.
    Schoenfeld, A.
    Seuntjens, J.
    Renaud, J.
    [J]. MEDICAL PHYSICS, 2021, 48 (06)
  • [4] Characterization of the PTB ultra-high pulse dose rate reference electron beam
    Bourgouin, Alexandra
    Knyziak, Adrian
    Marinelli, Marco
    Kranzer, Rafael
    Schueller, Andreas
    Kapsch, Ralf-Peter
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2022, 67 (08):
  • [5] Investigation of a Simple Calorimeter Design for the Dosimetry of Ultra-High Pulse Dose Rate (FLASH) Electron Beams
    McEwen, M.
    Bourgouin, A.
    Schueller, A.
    [J]. MEDICAL PHYSICS, 2020, 47 (06) : E294 - E294
  • [6] Calorimeter for Real-Time Dosimetry of Pulsed Ultra-High Dose Rate Electron Beams
    Bourgouin, Alexandra
    Schueller, Andreas
    Hackel, Thomas
    Kranzer, Rafael
    Poppinga, Daniela
    Kapsch, Ralf-Peter
    McEwen, Malcolm
    [J]. FRONTIERS IN PHYSICS, 2020, 8
  • [7] ABSORBED DOSE WATER CALORIMETER
    DOMEN, SR
    [J]. MEDICAL PHYSICS, 1980, 7 (02) : 157 - 159
  • [8] ABSORBED DOSE CALORIMETER FOR HIGH DOSE RATE
    PETREE, B
    [J]. RADIATION RESEARCH, 1958, 9 (01) : 166 - 166
  • [9] CALORIMETER FOR ABSOLUTE MEASUREMENT OF ABSORBED ELECTRON DOSE
    TSVETKOV, II
    FOMINYKH, VI
    [J]. MEASUREMENT TECHNIQUES, 1980, 23 (06) : 564 - 565
  • [10] AN IMPROVED ABSORBED DOSE WATER CALORIMETER
    DOMEN, SR
    [J]. MEDICAL PHYSICS, 1981, 8 (04) : 552 - 553