Electrical Interaction between Implantable Vagus Nerve Stimulation Device and Implantable Cardiac Rhythm Management Device

被引:0
|
作者
Libbus, Imad [1 ]
Mazar, Scott T. [1 ]
Stubbs, Scott R. [1 ]
KenKnight, Bruce H. [1 ]
机构
[1] LivaNova, Minneapolis, MN 55417 USA
关键词
CARDIOVERTER-DEFIBRILLATOR; HEART-FAILURE; PATIENT; PACEMAKER; UNIT;
D O I
暂无
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Background: Autonomic regulation therapy via vagus nerve stimulation (VNS) was recently approved as a therapy for chronic heart failure, and will likely be utilized in patients who are also indicated for cardiac rhythm management device implantation. This study is designed to assess the degree to which VNS is likely to cause interference in the cardiac sensing of an implantable cardiac rhythm management device. Methods: A VNS stimulation lead and a cardiac sensing lead were placed in a simulated biological medium. A nonconductive carrier frame was used to position the leads at a precise electrode spacing. Stimulation was delivered through the VNS Therapy lead at a maximum output current and a variety of combinations of stimulation frequencies from 5-30 Hz and stimulation pulse widths from 130-1000 mu s. The electrode spacing began at 0 cm and was increased in 1 cm increments until the measured signal dropped below the cardiac rhythm management device noise floor for sensing. The test was conducted with both bipolar and unipolar sensing. Results: In the bipolar sensing configuration, the maximum sensed signal amplitude was 687 mu V at an electrode separation of 0 cm, signal frequency of 30 Hz, pulse width of 1000 mu s, and output current of 3.5 mA. In the unipolar sensing configuration, the maximum amplitude was 406 mu V. In both configurations, the measured signal with maximum stimulation intensity decreased significantly with electrode separation, and dropped below the noise floor at an electrode spacing of 3.0 cm. The sensed signal amplitude was further attenuated at lower stimulation amplitudes and pulse widths. Conclusion: Even at maximum neural stimulation intensity of 3.5 mA, at an electrode separation of at least 3.0 cm, neural stimulation did not result in a detectable level of interference with either bipolar or unipolar sensing. Because this separation is significantly smaller than the minimum electrode separation of 15 cm in clinical practice, VNS Therapy is not expected to interfere with the function of implantable cardiac devices.
引用
收藏
页码:3681 / 3684
页数:4
相关论文
共 50 条
  • [31] Adaptive Thermal Management of Implantable Device
    Chai, Ruizhi
    Zhang, Ying
    IEEE SENSORS JOURNAL, 2019, 19 (03) : 1176 - 1185
  • [32] The diagnostic evolution of the cardiac implantable electronic device: The implantable monitor of ischemia
    Bunch, T. Jared
    Day, John D.
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2008, 19 (03) : 282 - 284
  • [33] An implantable device for stimulation of denervated muscles in rats
    Dennis, RG
    Dow, DE
    Faulkner, JA
    MEDICAL ENGINEERING & PHYSICS, 2003, 25 (03) : 239 - 253
  • [34] Effect of Defibrillation on the Performance of an Implantable Vagus Nerve Stimulation System
    Libbus, Imad
    Faltaous, Mary
    DiCarlo, Lorenzo
    JOURNAL OF CARDIAC FAILURE, 2020, 26 (10) : S157 - S157
  • [35] Effect of defibrillation on the performance of an implantable vagus nerve stimulation system
    Libbus I.
    Stubbs S.R.
    Mazar S.T.
    Mindrebo S.
    KenKnight B.H.
    DiCarlo L.A.
    Bioelectronic Medicine, 2021, 7 (01)
  • [36] On the design of implantable electrodes for electrical nerve stimulation
    Hanekom, T
    Hanekom, JJ
    SOUTH AFRICAN JOURNAL OF SCIENCE, 1998, 94 (07) : 307 - 310
  • [37] Two Implantable Devices for Cardiac Rhythm Management
    Tang, Jonathan E.
    Ryu, Jasmine N.
    Essandoh, Michael K.
    Savona, Salvatore J.
    JOURNAL OF CARDIOTHORACIC AND VASCULAR ANESTHESIA, 2023, 37 (02) : 326 - 329
  • [38] Transcutaneous auricular vagus nerve stimulation would be an alternative to implantable cervical vagus nerve stimulation in some situation
    HE Jiakai
    ZHANG Jinling
    WANG Yu
    LI Shaoyuan
    FANG Jiliang
    ZHANG Shuai
    ZHAO Yanan
    ZHAI Weihang
    GAO Deqiang
    LI Ran
    JIANG Yuhang
    CHEN Zehao
    JIA Baohui
    RONG Peijing
    Journal of Traditional Chinese Medicine, 2023, 43 (03) : 627 - 630
  • [39] Implantable device for long-term electrical stimulation of denervated muscles in rabbits
    Lanmüller, H
    Ashley, Z
    Unger, E
    Sutherland, H
    Reichel, M
    Russold, M
    Jarvis, J
    Mayr, W
    Salmons, S
    MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 2005, 43 (04) : 535 - 540
  • [40] A mm-Sized Wireless Implantable Device for Electrical Stimulation of Peripheral Nerves
    Charthad, Jayant
    Chang, Ting Chia
    Liu, Zhaokai
    Sawaby, Ahmed
    Weber, Marcus J.
    Baker, Sam
    Gore, Felicity
    Felt, Stephen A.
    Arbabian, Amin
    IEEE TRANSACTIONS ON BIOMEDICAL CIRCUITS AND SYSTEMS, 2018, 12 (02) : 257 - 270