On the effective point of measurement in megavoltage photon beams

被引:72
|
作者
Kawrakow, Iwan [1 ]
机构
[1] NRC, Ionizing Radiat Stand, Ottawa, ON K1A 0R6, Canada
关键词
D O I
10.1118/1.2201147
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
This paper presents a numerical investigation of the effective point of measurement of thimble ionization chambers in megavoltage photon beams using Monte Carlo simulations with the EGSNRC system. It is shown that the effective point of measurement for relative photon beam dosimetry depends on every detail of the chamber design, including the cavity length, the mass density of the wall material, and the size of the central electrode, in addition to the cavity radius. Moreover. the effective point of measurement also depends on the beam quality and the field size. The paper therefore argues that the upstream shift of 0.6 times the cavity radius, recommended in current dosimetry protocols, is inadequate for accurate relative photon beam dosimetry, particularly in the build-up region. On the other hand, once the effective point of measurement is selected appropriately, measured depth-ionization curves can be equated to measured depth-dose curves for all depths within +/-0.5%.
引用
收藏
页码:1829 / 1839
页数:11
相关论文
共 50 条
  • [21] Absorbed dose conversion factors for megavoltage photon beams
    Nutbrown, R
    Duane, S
    Shipley, D
    Thomas, R
    ADVANCED MONTE CARLO FOR RADIATION PHYSICS, PARTICLE TRANSPORT SIMULATION AND APPLICATIONS, 2001, : 299 - 304
  • [22] The effective point of measurement for depth-dose measurements in small MV photon beams with different detectors
    Wegener, Sonja
    Sauer, Otto A.
    MEDICAL PHYSICS, 2019, 46 (11) : 5209 - 5215
  • [24] Experimental determination of the effective point of measurement of cylindrical ionization chambers for high-energy photon and electron beams
    Huang, Yanxiao
    Willomitzer, Christian
    Abu Zakaria, Golam
    Hartmann, Guenther H.
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2010, 26 (03): : 126 - 131
  • [25] SURFACE DOSE FOR MEGAVOLTAGE PHOTON BEAMS OUTSIDE THE TREATMENT FIELD
    STARKSCHALL, G
    STGEORGE, FJ
    ZELLMER, DL
    MEDICAL PHYSICS, 1983, 10 (06) : 906 - 910
  • [26] Monte Carlo Investigation of Focal Spot in Megavoltage Photon Beams
    Ding, G. X.
    Zhang, H.
    Wang, J.
    Homann, K. L.
    Lim, S. B. G.
    Das, I. J.
    MEDICAL PHYSICS, 2024, 51 (10) : 7898 - 7898
  • [27] SURFACE DOSE FOR MEGAVOLTAGE PHOTON BEAMS OUTSIDE THE TREATMENT FIELD
    STARKSCHALL, G
    STGEORGE, FJ
    ZELLMER, DL
    MEDICAL PHYSICS, 1982, 9 (04) : 632 - 632
  • [28] A DOSIMETRIC INTERCOMPARISON OF MEGAVOLTAGE PHOTON BEAMS IN UK RADIOTHERAPY CENTERS
    THWAITES, DI
    WILLIAMS, JR
    AIRD, EG
    KLEVENHAGEN, SC
    WILLIAMS, PC
    PHYSICS IN MEDICINE AND BIOLOGY, 1992, 37 (02): : 445 - 461
  • [29] Characteristics of mobile MOSFET dosimetry system for megavoltage photon beams
    Kumar, A. Sathish
    Sharma, S. D.
    Ravindran, B. Paul
    JOURNAL OF MEDICAL PHYSICS, 2014, 39 (03) : 142 - 149
  • [30] Response of a NanoDot OSLD System in Megavoltage Photon and Electron Beams
    Hoshida, K.
    Araki, F.
    Ohno, T.
    Shimohigashi, Y.
    MEDICAL PHYSICS, 2018, 45 (06) : E314 - E314