Characterization of a full field digital mammography detector based on direct x-ray conversion in selenium

被引:41
|
作者
Yorker, JG [1 ]
Jeromin, LS [1 ]
Lee, DLY [1 ]
Palecki, EF [1 ]
Golden, KP [1 ]
Jing, ZX [1 ]
机构
[1] Direct Radiog, Newark, DE 19702 USA
关键词
digital mammography; direct x-ray detector; selenium; MTF; DQE;
D O I
10.1117/12.465568
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
We describe a high-resolution digital x-ray detector suitable for producing high quality mammographic images. The detector consists of an array of 3584 by 4096 pixels on 70 micrometer centers covering an area of 25 cm by 29 cm. The conversion layer of the detector is 250 micrometer thick amorphous selenium. Each pixel of the array consists of a storage capacitor for collecting x-ray signals and an amorphous silicon switching transistor. The signal is read out by custom high-speed, low-noise electronics. The integration of this detector with a mammographic x-ray system and acquisition console is described, as well as algorithms for calibration of the full system. We review characterization of the imaging performance of our system based on quantitative analyses of MTF and DQE data, and compare experimental results with theoretical calculations. We compare the performance of our direct conversion system with that of screen/film analog systems and indirect conversion digital detectors, such as LORAD's CsI/CCD detector, operated under similar conditions. MTF degradation mechanisms and system noise sources and their effect on DQE are discussed. We review qualitative aspects of image quality from our detector and present preliminary observer performance characteristics on clinical studies run with our system.
引用
收藏
页码:21 / 29
页数:3
相关论文
共 50 条
  • [1] Imaging characteristics of a direct conversion full-field digital mammography detector using selenium
    Lee, DL
    Yorker, JG
    Jing, ZX
    Jeromin, LS
    [J]. DIGITAL MAMMOGRAPHY, PROCEEDINGS, 2003, : 43 - 47
  • [2] Evaluation of a direct conversion full-field digital mammography detector
    Lee, DL
    Yorker, JG
    Jing, Z
    Hoffberg, MG
    Jeromin, LS
    [J]. RADIOLOGY, 2001, 221 : 463 - 464
  • [3] Physical characterization of a prototype selenium-based full field digital mammography detector
    Saunders, RS
    Samei, E
    Jesneck, JL
    Lo, JY
    [J]. MEDICAL PHYSICS, 2005, 32 (02) : 588 - 599
  • [4] An integrated CMOS-selenium x-ray detector for digital mammography
    Andre, MP
    Spivey, B
    Martin, P
    Morsell, L
    Atlas, E
    Pellegrino, T
    [J]. PHYSICS OF MEDICAL IMAGING, 1998, 3336 : 204 - 209
  • [5] Flexible x-ray imaging detector based on direct conversion in amorphous selenium
    Kuo, Tsung-Ter
    Wu, Chien-Ming
    Lu, Hui-Hsin
    Chan, Isaac
    Wang, Kai
    Leou, Keh-Chyang
    [J]. JOURNAL OF VACUUM SCIENCE & TECHNOLOGY A, 2014, 32 (04):
  • [6] Monoenergetic X-ray beam in digital full field mammography
    Diekmann, F
    Diekmann, SD
    Fischer, T
    Arkadiev, V
    Lawaczeck, R
    Hamm, BK
    [J]. RADIOLOGY, 2002, 225 : 417 - 417
  • [7] X-ray scattering in full-field digital mammography
    Nykänen, K
    Siltanen, S
    [J]. MEDICAL PHYSICS, 2003, 30 (07) : 1864 - 1873
  • [8] Characterization of an x-ray imaging setup for digital mammography based on a silicon microstrip detector
    Mali, T
    Cindro, V
    Mikuz, M
    Krizaj, D
    Vrtacnik, D
    Zdesar, U
    [J]. IWDM 2000: 5TH INTERNATIONAL WORKSHOP ON DIGITAL MAMMOGRAPHY, 2001, : 89 - 94
  • [9] Effective x-ray attenuation measurements with full field digital mammography
    Heine, John J.
    Behera, Madhusmita
    [J]. MEDICAL PHYSICS, 2006, 33 (11) : 4350 - 4366
  • [10] Characterization of a direct, full-field flat-panel digital mammography detector
    Zhao, B
    Zhao, W
    [J]. MEDICAL IMAGING 2003: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2, 2003, 5030 : 157 - 167