Characterization of a full field digital mammography detector based on direct x-ray conversion in selenium

被引:41
|
作者
Yorker, JG [1 ]
Jeromin, LS [1 ]
Lee, DLY [1 ]
Palecki, EF [1 ]
Golden, KP [1 ]
Jing, ZX [1 ]
机构
[1] Direct Radiog, Newark, DE 19702 USA
来源
MEDICAL IMAGING 2002: PHYSICS OF MEDICAL IMAGING | 2002年 / 4682卷
关键词
digital mammography; direct x-ray detector; selenium; MTF; DQE;
D O I
10.1117/12.465568
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
We describe a high-resolution digital x-ray detector suitable for producing high quality mammographic images. The detector consists of an array of 3584 by 4096 pixels on 70 micrometer centers covering an area of 25 cm by 29 cm. The conversion layer of the detector is 250 micrometer thick amorphous selenium. Each pixel of the array consists of a storage capacitor for collecting x-ray signals and an amorphous silicon switching transistor. The signal is read out by custom high-speed, low-noise electronics. The integration of this detector with a mammographic x-ray system and acquisition console is described, as well as algorithms for calibration of the full system. We review characterization of the imaging performance of our system based on quantitative analyses of MTF and DQE data, and compare experimental results with theoretical calculations. We compare the performance of our direct conversion system with that of screen/film analog systems and indirect conversion digital detectors, such as LORAD's CsI/CCD detector, operated under similar conditions. MTF degradation mechanisms and system noise sources and their effect on DQE are discussed. We review qualitative aspects of image quality from our detector and present preliminary observer performance characteristics on clinical studies run with our system.
引用
收藏
页码:21 / 29
页数:3
相关论文
共 50 条
  • [21] Evaluation of a digital X-ray detector based on a phosphor-coated CCD for mammography
    Taibi, A
    DelGuerra, A
    Gambaccini, M
    Marziani, M
    Tuffanelli, A
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1997, 392 (1-3): : 210 - 213
  • [22] X-ray imaging with amorphous selenium: Optimal spectra for digital mammography
    Fahrig, R
    Rowlands, JA
    Yaffe, MJ
    MEDICAL PHYSICS, 1996, 23 (04) : 557 - 567
  • [23] Investigation of Ghosting Recovery Mechanisms in Selenium X-ray Detector Structures for Mammography
    Mahmood, S. A.
    Kabir, M. Z.
    Tousignant, O.
    Greenspan, J.
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2012, 59 (03) : 597 - 604
  • [24] X-ray spectrum optimization of full-field digital mammography: Simulation and phantom study
    Bernhardt, Philipp
    Mertelmeier, Thomas
    Hoheisel, Martin
    MEDICAL PHYSICS, 2006, 33 (11) : 4337 - 4349
  • [25] Application of X-ray crystallographic detector to whole breast digital mammography
    Xie, JH
    Williams, MB
    Stanton, M
    Rosen, D
    Phillips, W
    Fajardo, LL
    DIGITAL MAMMOGRAPHY '96, 1996, 1119 : 139 - 144
  • [26] Direct-Conversion X-Ray Detector with 50-µm High-Gain Pixel Amplifiers for Low-X-Ray-Dose Digital Mammography
    Tsung-Ter Kuo
    Chien-Ming Wu
    Isaac Chan
    Hui-Hsin Lu
    Sun-Hua Pao
    Journal of Medical and Biological Engineering, 2015, 35 : 249 - 257
  • [27] Direct-Conversion X-Ray Detector with 50-Aμm High-Gain Pixel Amplifiers for Low-X-Ray-Dose Digital Mammography
    Kuo, Tsung-Ter
    Wu, Chien-Ming
    Chan, Isaac
    Lu, Hui-Hsin
    Pao, Sun-Hua
    JOURNAL OF MEDICAL AND BIOLOGICAL ENGINEERING, 2015, 35 (02) : 249 - 257
  • [28] CCD-based detector for full-field digital mammography
    Stanton, M
    Phillips, W
    Stewart, A
    Smilowitz, L
    Williams, MB
    Simoni, P
    Ingersoll, C
    McCauley, T
    Qian, H
    MEDICAL IMAGING 1999: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2, 1999, 3659 : 740 - 748
  • [29] Radiation exposure in full-field digital mammography with a selenium flat-panel detector
    Gosch, D.
    Jendrass, S.
    Scholz, M.
    Kahn, T.
    ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2006, 178 (07): : 693 - 697
  • [30] Radiation exposure in full-field digital mammography with a flat-panel x-ray detector based on amorphous silicon in comparison with conventional screen-film mammography.
    Hermann, KP
    Obenauer, S
    Grabbe, E
    ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2000, 172 (11): : 940 - 945