Imprinted Zeolite Modified Carbon Paste Electrode as a Selective Sensor for Blood Glucose Analysis by Potentiometry

被引:4
|
作者
Khasanah, Miratul [1 ]
Widati, Alfa Akustia [1 ]
Handajani, Usreg Sri [1 ]
Harsini, Muji [1 ]
Ilmiah, Bahrotul [1 ]
Oktavia, Irene Dinda [1 ]
机构
[1] Univ Airlangga, Fac Sci & Technol, Dept Chem, Campus C,Jl Dr Ir H Soekarno MERR, Surabaya 60115, Indonesia
关键词
blood glucose; carbon paste electrode; imprinted zeolite; potentiometric sensor; HYDROTHERMAL SYNTHESIS; PERFORMANCE; MEMBRANES; NAA;
D O I
10.22146/ijc.49820
中图分类号
O6 [化学];
学科分类号
0703 ;
摘要
Imprinted zeolite modified carbon paste (carbon paste-IZ) electrode had been developed as a sensor to analyze blood glucose content by potentiometry. The used zeolite was Lynde Type A (LTA) that synthesized with a mole ratio of Na2O, Al2O3, SiO2 and H2O of 4:1:1.8:270, respectively while non-imprinted zeolite was prepared with a mole ratio of glucose/Si of 0.0306. Glucose was then extracted from the zeolite framework using hot water (80 degrees C) to produce imprinted zeolite (IZ). The carbon paste-IZ electrode prepared from activated carbon, paraffin pastilles, and IZ with a mass ratio of 5:4:1 showed the best performance. The modified electrode demonstrated the measurement range of 10(-4)-10(-2) M, the Nernst factor of 29.55 mV/decade, the response time less than 120 s, and the detection limit of 5.62 x 10(-5) M. Ascorbic acid, uric acid, urea and creatinine did not interfere on the glucose analysis by potentiometry. Comparison test with spectrophotometry showed an accuracy of (90.7 +/- 1.4)% (n = 5), while the application of the electrode to analyze five spiked serum samples showed recovery of (92.2 +/- 1.3)% (n = 5). The electrode was stable for up to 9 weeks (168 times usage). Based on its performance, the developed electrode can be applied to analyze glucose in human serum sample and recommended for used in the medical field.
引用
收藏
页码:1301 / 1310
页数:10
相关论文
共 50 条
  • [1] Carbon Paste Electrode Modified Imprinted Zeolite as a Selective Sensor for Creatine Analysis by Potentiometry
    Athiroh, A.
    Fadillah, T.
    Damayanti, D. F.
    Widati, A. A.
    Abdulloh, A.
    Khasanah, M.
    [J]. 12TH CONGRESS OF INDONESIAN SOC. FOR BIOCHEMISTRY AND MOLECULAR BIOLOGY IN CONJUNCTION WITH THE 2ND INT. CONF COLLABORATION SEMINAR OF CHEMISTRY AND INDUSTRY (COSCI) AND ANMICRO WORKSHOP, 2018, 2019, 217
  • [2] Imprinted Zeolite Modified Carbon Paste Electrode as A Selective Potentiometric Sensor for Blood Glucose
    Khasanah, Miratul
    Widati, Alfa Akustia
    Handajani, Usreg Sri
    Shofiyyah, Masfah Raudlotus
    Rakhma, Sabrina Aulia
    Predianto, Herwin
    [J]. 14TH JOINT CONFERENCE ON CHEMISTRY 2019, 2020, 2237
  • [3] Imprinted Zeolite Modified Carbon Paste Electrode as a Potentiometric Sensor for Uric Acid
    Khasanah, Miratul
    Widati, Alfa Akustia
    Fitri, Sarita Aulia
    [J]. 5TH INTERNATIONAL CONFERENCE AND WORKSHOP ON BASIC AND APPLIED SCIENCES (ICOWOBAS 2015), 2016, 1718
  • [4] High Selective Methadone Sensor Based on Molecularly Imprinted Polymer Carbon Paste Electrode Modified with Carbon Nanotubes
    Soleimani, Majid
    Afshar, Majid Ghahraman
    Ganjali, Mohammad Reza
    [J]. SENSOR LETTERS, 2013, 11 (10) : 1983 - 1991
  • [5] Selective and sensitive voltammetric sensor for methocarbamol determination by molecularly imprinted polymer modified carbon paste electrode
    Payehghadr, Mahmood
    Taherkhani, Yeganeh
    Maleki, Abbas
    Nourifard, Farzaneh
    [J]. EURASIAN CHEMICAL COMMUNICATIONS, 2020, 2 (09): : 982 - 990
  • [6] Zeolite-modified carbon paste electrode for selective monitoring of dopamine
    Wang, J
    Walcarius, A
    [J]. JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 1996, 407 (1-2): : 183 - 187
  • [7] Carbon Paste Electrode Modified Molecularly Imprinted Polymer as a Sensor for Creatinine Analysis by Stripping Voltammetry
    Khasanah, M.
    Darmokoesoemo, H.
    Rizki, D. A.
    [J]. 6TH INTERNATIONAL CONFERENCE AND WORKSHOPS ON BASIC AND APPLIES SCIENCES, 2017, 1888
  • [8] A Novel and Selective Methylene Blue Imprinted Polymer Modified Carbon Paste Electrode
    Soysal, Mert
    Muti, Mihrican
    Esen, Cem
    Gencdag, Kubra
    Aslan, Alican
    Erdem, Arzum
    Karagoezler, A. Ersin
    [J]. ELECTROANALYSIS, 2013, 25 (05) : 1278 - 1285
  • [9] Imprinted-Zeolite-X-Based Sensor for Non-Enzymatic Detection of Blood Glucose by Potentiometry
    Khasanah, Miratul
    Widati, Alfa Akustia
    Handajani, Usreg Sri
    Mastura, Akhsin
    Sari, Eka Yunicha
    [J]. CHEMENGINEERING, 2022, 6 (05)
  • [10] THE ZEOLITE-MODIFIED CARBON PASTE ELECTRODE
    ELMURR, N
    KERKENI, M
    SELLAMI, A
    BENTAARIT, Y
    [J]. JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 1988, 246 (02): : 461 - 465