Validation of the professional device for blood pressure measurement Microlife WatchBP Office in adults and children according to the American National Standards Institute/Association for the Advancement of Medical Instrumentation/International Organization for Standardization standard

被引:25
|
作者
Kollias, Anastasios [1 ]
Ntineri, Angeliki [1 ]
Kyriakoulis, Konstantinos G. [1 ]
Stambolliu, Emelina [1 ]
Lagou, Styliani [1 ]
Boubouchairopoulou, Nadia [1 ]
Stergiou, George S. [1 ]
机构
[1] Natl & Kapodistrian Univ Athens, Sch Med, Hypertens Ctr STRIDE 7, Dept Med,Sotiria Hosp, 152 Mesogion Ave, Athens 11527, Greece
关键词
accuracy; automated; blood pressure measurement; office; oscillometric; professional; validation; SOCIETY;
D O I
10.1097/MBP.0000000000000307
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
ObjectiveTo assess the accuracy of the professional oscillometric blood pressure (BP) monitor Microlife WatchBP Office in adults and children according to the American National Standards Institute/Association for the Advancement of Medical Instrumentation/International Organization for Standardization (ANSI/AAMI/ISO) 81060-2:2013 standard.MethodsAdults and children (aged 3-12 years) were recruited to fulfil the age, sex, BP and cuff distribution criteria of the ANSI/AAMI/ISO standard using the same-arm sequential BP measurement method. Three cuffs of the test device were used for arm circumferences of 14-22, 22-32 and 32-42cm.ResultsA total of 115 participants were recruited and 88 were included in the analysis (51 adults and 37 children). For criterion 1, the meanSD of the differences between the test device and reference BP was -1.0 +/- 7.0/-4.7 +/- 5.4mmHg (systolic/diastolic). For criterion 2, the SD of the averaged BP differences between the test device and reference per participant was 5.15/4.77mmHg (systolic/diastolic).ConclusionThe professional Microlife WatchBP Office BP monitor fulfilled the requirements of the ANSI/AAMI/ISO validation standard for adults and children and can be recommended for clinical use.
引用
收藏
页码:112 / 114
页数:3
相关论文
共 38 条
  • [21] A universal standard for the validation of blood pressure measuring devices: Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO) Collaboration Statement
    Stergiou, George S.
    Alpert, Bruce
    Mieke, Stephan
    Asmar, Roland
    Atkins, Neil
    Eckert, Siegfried
    Frick, Gerhard
    Friedman, Bruce
    Grassl, Thomas
    Ichikawa, Tsutomu
    Ioannidis, John P.
    Lacy, Peter
    McManus, Richard
    Murray, Alan
    Myers, Martin
    Palatini, Paolo
    Parati, Gianfranco
    Quinn, David
    Sarkis, Josh
    Shennan, Andrew
    Usuda, Takashi
    Wang, Jiguang
    Wu, Colin O.
    O'Brien, Eoin
    JOURNAL OF HYPERTENSION, 2018, 36 (03) : 472 - 478
  • [22] Recommendations and Practical Guidance for performing and reporting validation studies according to the Universal Standard for the validation of blood pressure measuring devices by the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO)
    Stergiou, George S.
    Palatini, Paolo
    Asmar, Roland
    Ioannidis, John P.
    Kollias, Anastasios
    Lacy, Peter
    McManus, Richard J.
    Myers, Martin G.
    Parati, Gianfranco
    Shennan, Andrew
    Wang, Jiguang
    O'Brien, Eoin
    JOURNAL OF HYPERTENSION, 2019, 37 (03) : 459 - 466
  • [23] A novel professional automated auscultatory blood pressure monitor with visual display of Korotkoff sounds: InBody BPBIO480KV validation according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization Universal Standard
    Ntineri, Angeliki
    Theodosiadi, Aikaterini
    Menti, Ariadni
    Kyriakoulis, Konstantinos G.
    Ntousopoulos, Vasileios
    Kollias, Anastasios
    Stergiou, George S.
    JOURNAL OF HYPERTENSION, 2023, 41 (02) : 356 - 361
  • [24] Validation of the single-cuff oscillometric blood pressure monitor InBody BPBIO750 for public spaces according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization Universal Standard
    Ntineri, Angeliki
    Prapa, Sofia
    Bountzona, Ioanna
    Menti, Ariadni
    Stergiou, George S.
    BLOOD PRESSURE MONITORING, 2021, 26 (02) : 146 - 148
  • [25] Validation of Omron HBP-1100-E Professional Blood Pressure Measuring Device According to the American Association for the Advancement of Medical Instrumentation Protocol: The PERSIAN Guilan Cohort Study (PGCS)
    Joukar, Farahnaz
    Yeganeh, Sara
    Naghipour, Mohammadreza
    Hassanipour, Soheil
    Nikbakht, Hossein-Ali
    Mansour-Ghanaei, Fariborz
    MEDICAL DEVICES-EVIDENCE AND RESEARCH, 2020, 13 : 231 - 236
  • [26] Validation of the FM-800 Ambulatory Blood Pressure Monitor According to the Association for the Advancement of Medical Instrumentation Criteria and the International Protocol
    Nakamura, Kaori
    Kikuya, Masahiro
    Hara, Azusa
    Hirose, Takuo
    Obara, Taku
    Metoki, Hirohito
    Asayama, Kei
    Inoue, Ryusuke
    Ohkubo, Takayoshi
    Totsune, Kazuhito
    Imai, Yutaka
    CLINICAL AND EXPERIMENTAL HYPERTENSION, 2010, 32 (08) : 523 - 527
  • [27] Validation of the A&D UM-101 professional device for office blood pressure measurement according to the international protocol
    Stergiou, G.
    Giovas, P.
    Gkinos, C.
    Tzamouranis, D.
    JOURNAL OF HYPERTENSION, 2007, 25 : S164 - S164
  • [28] Clinical Accuracy of the Withings BPM Connect for Self-Blood Pressure Measurements in Pregnancy and Pre-Eclampsia: Validation According to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization Universal Standard
    Zelveian, Parounak
    Topouchian, Jirar
    Hakobyan, Zoya
    Asmar, Jennifer
    Gharibyan, Heghine
    Asmar, Roland
    VASCULAR HEALTH AND RISK MANAGEMENT, 2022, 18 : 181 - 189
  • [29] Validation of the KANG KC-2850 oscillometric upper-arm ambulatory blood pressure monitor in general population according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization universal standard
    Liu, Kui
    Ma, Yi-Fei
    Wang, Yan
    Zhang, Yang
    Li, Lin-Yi
    Zhang, Jie-Qiong
    Jiang, Jun
    Chen, Rui-Lin
    Wan, Yi
    BLOOD PRESSURE MONITORING, 2022, 27 (05) : 349 - 352
  • [30] Validation of the A&D UM-101 professional hybrid device for office blood pressure measurement according to the International Protocol
    Stergiou, George S.
    Giovas, Periklis P.
    Gkinos, Charilaos P.
    Tzamouranis, Dimitris G.
    BLOOD PRESSURE MONITORING, 2008, 13 (01) : 37 - 42