Validation of the professional device for blood pressure measurement Microlife WatchBP Office in adults and children according to the American National Standards Institute/Association for the Advancement of Medical Instrumentation/International Organization for Standardization standard

被引:25
|
作者
Kollias, Anastasios [1 ]
Ntineri, Angeliki [1 ]
Kyriakoulis, Konstantinos G. [1 ]
Stambolliu, Emelina [1 ]
Lagou, Styliani [1 ]
Boubouchairopoulou, Nadia [1 ]
Stergiou, George S. [1 ]
机构
[1] Natl & Kapodistrian Univ Athens, Sch Med, Hypertens Ctr STRIDE 7, Dept Med,Sotiria Hosp, 152 Mesogion Ave, Athens 11527, Greece
关键词
accuracy; automated; blood pressure measurement; office; oscillometric; professional; validation; SOCIETY;
D O I
10.1097/MBP.0000000000000307
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
ObjectiveTo assess the accuracy of the professional oscillometric blood pressure (BP) monitor Microlife WatchBP Office in adults and children according to the American National Standards Institute/Association for the Advancement of Medical Instrumentation/International Organization for Standardization (ANSI/AAMI/ISO) 81060-2:2013 standard.MethodsAdults and children (aged 3-12 years) were recruited to fulfil the age, sex, BP and cuff distribution criteria of the ANSI/AAMI/ISO standard using the same-arm sequential BP measurement method. Three cuffs of the test device were used for arm circumferences of 14-22, 22-32 and 32-42cm.ResultsA total of 115 participants were recruited and 88 were included in the analysis (51 adults and 37 children). For criterion 1, the meanSD of the differences between the test device and reference BP was -1.0 +/- 7.0/-4.7 +/- 5.4mmHg (systolic/diastolic). For criterion 2, the SD of the averaged BP differences between the test device and reference per participant was 5.15/4.77mmHg (systolic/diastolic).ConclusionThe professional Microlife WatchBP Office BP monitor fulfilled the requirements of the ANSI/AAMI/ISO validation standard for adults and children and can be recommended for clinical use.
引用
收藏
页码:112 / 114
页数:3
相关论文
共 38 条
  • [31] Simultaneous validation of the Grandway MD2301 digital automatic blood pressure monitor by the British Hypertension Society and the Association for the Advancement of Medical Instrumentation/the International Organization for Standardization protocols
    Huang, Jinhua
    Wang, Yun
    Liu, Zhaoying
    Wang, Yuling
    BLOOD PRESSURE MONITORING, 2017, 22 (01) : 44 - 47
  • [32] Validation of the Raycome model M2 ambulatory blood pressure monitor in the general population according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization Universal Standard (ISO 81060-2:2018)
    Yang, Shijie
    Zhou, Zhanyang
    Miao, Huanhuan
    Zhang, Hongye
    Zhou, Qiong
    Zhai, Mei
    Zhang, Yuqing
    BLOOD PRESSURE MONITORING, 2024, 29 (03) : 161 - 165
  • [33] Validation of the YuWell YE660D oscillometric upper-arm blood pressure monitor for clinic and home in general population according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization Universal Standard (International Organization for Standardization 81060-2:2018) and the International Organization for Standardization Amendment 1. 2020
    Zhang, Hui-Jie
    Peng, Yue
    Zhang, Juan
    Zhang, Jin
    Teng, Li-Ni
    Zhang, Shu-Juan
    Zhou, De-Jun
    Long, Ming-Zhi
    BLOOD PRESSURE MONITORING, 2023, 28 (05) : 276 - 279
  • [35] Validation of the oscillometric home blood pressure monitor Braun BUA4000 with wide-range cuff in a general population according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO) Universal Standard
    Ntousopoulos, Vasileios
    Ntineri, Angeliki
    Theodosiadi, Aikaterini
    Stathopoulou, Panagiota
    Menti, Ariadni
    Kyriakoulis, Konstantinos G.
    Kollias, Anastasios
    Stergiou, George S.
    BLOOD PRESSURE MONITORING, 2024, 29 (02) : 99 - 102
  • [36] Validation of the automated oscillometric upper-arm cuff home blood pressure monitor AVITA BPM82 in a general population according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO) Universal Standard
    Menti, Ariadni
    Ntousopoulos, Vasileios
    Theodosiadi, Aikaterini
    Stathopoulou, Panagiota
    Kollias, Anastasios
    Stergiou, George S.
    BLOOD PRESSURE MONITORING, 2024, 29 (02) : 89 - 92
  • [37] Validation of the YuWell YE900 oscillometric blood pressure monitor for professional office use in adults and children according to the AAMI/ESH/ISO Universal Standard (ISO 81060-2:2018)
    Zhang, Hui-Jie
    Zhang, Juan
    Wang, Su-Lan
    Zhang, Jin
    Teng, Li-Ni
    Zhang, Shu-Juan
    Zhou, De-Jun
    Long, Ming-Zhi
    BLOOD PRESSURE MONITORING, 2021, 26 (05) : 396 - 399
  • [38] Validation of the blood pressure measurement technology used in the Novacor Diasys 3 (DIS-0001-00) upper arm device for ambulatory blood pressure measurement, according to the requirements of the AAMI/ANSI/ISO 81060-2:2013 standard (for both a general study and a cardiac-stress study in adults) and of the European Society of Hypertension International Protocol revision 2010
    Atkins, Neil
    Fania, Claudio
    Palatini, Paolo
    BLOOD PRESSURE MONITORING, 2021, 26 (01) : 70 - 77