A novel professional automated auscultatory blood pressure monitor with visual display of Korotkoff sounds: InBody BPBIO480KV validation according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization Universal Standard

被引:4
|
作者
Ntineri, Angeliki [1 ]
Theodosiadi, Aikaterini [1 ]
Menti, Ariadni [1 ]
Kyriakoulis, Konstantinos G. [1 ]
Ntousopoulos, Vasileios [1 ]
Kollias, Anastasios [1 ]
Stergiou, George S. [1 ,2 ]
机构
[1] Natl & Kapodistrian Univ Athens, Sotiria Hosp, Hypertens Ctr STRIDE 7, Sch Med,Dept Med 3, Athens, Greece
[2] Natl & Kapodistrian Univ Athens, Sotiria Hosp, Sch Med, Dept Med 3,Hypertens Ctr STRIDE 7, 152 Mesog Ave, Athens 11527, Greece
关键词
accuracy; auscultatory; blood pressure measurement; clinic; validation; MEASURING DEVICES;
D O I
10.1097/HJH.0000000000003341
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
Objective:A novel automated auscultatory upper arm-cuff blood pressure (BP) monitor (InBody BPBIO480KV) for office use was developed. An electronic stethoscope embedded in the device cuff records the Korotkoff sounds, which are audible to the user and graphically displayed during cuff deflation. Automated BP measurements are provided, while allowing the user to assess the Korotkoff sounds. The device accuracy was tested using the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO) Universal Standard (ISO 81060-2:2018) and its Amendment 1.2020-01.Methods:Participants were recruited to fulfil the age, sex, BP, arm circumference and cuff distribution criteria of the Universal Standard in general population using the same arm sequential measurement method. Three cuffs of the test device were used for arm circumference 23-28, 28-35 and 33-42 cm.Results:Data from 85 individuals were analysed [mean age 57.3 +/- 15.0 (SD) years, 53 men, arm circumference 23-42 cm]. For validation criterion 1, the mean +/- SD of the differences between the test device and reference BP readings (N = 255) was 0.3 +/- 5.5/0.6 +/- 4.7 mmHg (systolic/diastolic; threshold <= 5 +/- 8 mmHg). For criterion 2, the SD of the averaged BP differences per individual (N = 85) was 3.76/3.61 mmHg (systolic/diastolic; threshold <= 6.95/6.91 mmHg).Conclusion:The InBody BPBIO480KV device for office use, which provides automated auscultatory measurements while reproducing and displaying the Korotkoff sounds, comfortably fulfilled the AAMI/ESH/ISO Universal Standard requirements in general population and can be recommended for clinical use. The assessment of Korotkoff sounds by healthcare professionals for evaluating the quality of automated measurements requires further evaluation.
引用
收藏
页码:356 / 361
页数:6
相关论文
共 29 条
  • [1] A NOVEL PROFESSIONAL AUTOMATED AUSCULTATORY BLOOD PRESSURE MONITOR FOR OFFICE USE WITH VISUAL DISPLAY OF KOROTKOFF SOUNDS: INBODY BPBIO480KV VALIDATION
    Kyriakoulis, Konstantinos G.
    Ntineri, Angeliki
    Theodosiadi, Aikaterini
    Menti, Ariadni
    Ntousopoulos, Vasileios
    Kollias, Anastasios
    Stergiou, George S.
    JOURNAL OF HYPERTENSION, 2023, 41 : E213 - E213
  • [2] VISUAL ASSESSMENT OF KOROTKOFF SOUNDS FURTHER IMPROVES THE ACCURACY OF A VALIDATED PROFESSIONAL AUTOMATED AUSCULTATORY BLOOD PRESSURE MONITOR - INBODY BPBIO480KV
    Ntineri, Angeliki
    Menti, Ariadni
    Theodosiadi, Aikaterini
    Ntousopoulos, Vasileios
    Stathopoulou, Panagiota
    Kyriakoulis, Konstantinos
    Kollias, Anastasios
    Stergiou, George S.
    JOURNAL OF HYPERTENSION, 2023, 41 : E104 - E105
  • [3] Validation of the InBody BPBIO250 oscillometric blood pressure monitor for professional office use in general population according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization Universal Standard
    Kollias, Anastasios
    Gravvani, Areti
    Anagnostopoulos, Ioannis
    Kyriakoulis, Konstantinos G.
    Bountzona, Ioanna
    Menti, Ariadni
    Stergiou, George S.
    BLOOD PRESSURE MONITORING, 2020, 25 (02) : 115 - 117
  • [4] Validation of the single-cuff oscillometric blood pressure monitor InBody BPBIO750 for public spaces according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization Universal Standard
    Ntineri, Angeliki
    Prapa, Sofia
    Bountzona, Ioanna
    Menti, Ariadni
    Stergiou, George S.
    BLOOD PRESSURE MONITORING, 2021, 26 (02) : 146 - 148
  • [5] Validation of the InBody BPBIO210 manual auscultatory hybrid device for professional office use in a general population according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization Universal Standard
    Ntineri, Angeliki
    Menti, Ariadni
    Kyriakoulis, Konstantinos G.
    Bountzona, Ioanna
    Prapa, Sofia
    Kollias, Anastasios
    Stergiou, George S.
    BLOOD PRESSURE MONITORING, 2022, 27 (02) : 135 - 138
  • [6] Validation of the InBody BP170 oscillometric home blood pressure monitor in general population according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization Universal Standard
    Kollias, Anastasios
    Anagnostopoulos, Ioannis
    Gravvani, Areti
    Stambolliu, Emelina
    Bountzona, Ioanna
    Menti, Ariadni
    Stergiou, George S.
    BLOOD PRESSURE MONITORING, 2020, 25 (01) : 50 - 52
  • [7] Validation of the RBP-9801 oscillometric blood pressure monitor in the general population according to the association for the Advancement of Medical Instrumentation/European Society of Hypertension/ International Organization for Standardization Universal Standard
    Yang, Shijie
    Zhou, Zhanyang
    Miao, Huanhuan
    Zhou, Qiong
    Zhai, Mei
    Zhang, Yuqing
    JOURNAL OF CLINICAL HYPERTENSION, 2024, 26 (06): : 703 - 707
  • [8] Validation of the EDAN SA-10 ambulatory blood pressure monitor in general population according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization Universal Standard
    Wang, Bing
    Liu, Kui
    Li, Linyi
    Yin, Shimin
    Ren, Jie
    BLOOD PRESSURE MONITORING, 2022, 27 (04) : 276 - 279
  • [9] A Universal Standard for the Validation of Blood Pressure Measuring Devices Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO) Collaboration Statement
    Stergiou, George S.
    Alpert, Bruce
    Mieke, Stephan
    Asmar, Roland
    Atkins, Neil
    Eckert, Siegfried
    Frick, Gerhard
    Friedman, Bruce
    Grassl, Thomas
    Ichikawa, Tsutomu
    Ioannidis, John P.
    Lacy, Peter
    McManus, Richard
    Murray, Alan
    Myers, Martin
    Palatini, Paolo
    Parati, Gianfranco
    Quinn, David
    Sarkis, Josh
    Shennan, Andrew
    Usuda, Takashi
    Wang, Jiguang
    Wu, Colin O.
    O'Brien, Eoin
    HYPERTENSION, 2018, 71 (03) : 368 - 374
  • [10] A universal standard for the validation of blood pressure measuring devices: Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO) Collaboration Statement
    Stergiou, George S.
    Alpert, Bruce
    Mieke, Stephan
    Asmar, Roland
    Atkins, Neil
    Eckert, Siegfried
    Frick, Gerhard
    Friedman, Bruce
    Grassl, Thomas
    Ichikawa, Tsutomu
    Ioannidis, John P.
    Lacy, Peter
    McManus, Richard
    Murray, Alan
    Myers, Martin
    Palatini, Paolo
    Parati, Gianfranco
    Quinn, David
    Sarkis, Josh
    Shennan, Andrew
    Usuda, Takashi
    Wang, Jiguang
    Wu, Colin O.
    O'Brien, Eoin
    JOURNAL OF HYPERTENSION, 2018, 36 (03) : 472 - 478