Evaluation of the Diesse Cube 30 touch erythrocyte sedimentation method in comparison with Alifax test 1 and the manual Westergren gold standard method

被引:10
|
作者
Pieri, Massimo [1 ]
Pignalosa, Stefano [1 ]
Perrone, Marco Alfonso [2 ]
Russo, Carmelo [1 ]
Noce, Gianluca [1 ]
Perrone, Andrea [3 ]
Terrinoni, Alessandro [1 ]
Massoud, Renato [1 ]
Bernardini, Sergio [1 ]
机构
[1] Univ Roma Tor Vergata, Dept Expt Med & Surg, Div Clin Biochem & Clin Mol Biol, Viale Montpellier 1, I-00133 Rome, Italy
[2] Univ Roma Tor Vergata, Dept Expt Med, Div Cardiol, Rome, Italy
[3] Univ Roma Tor Vergata, Dept Enterprise Engn, Rome, Italy
关键词
Erythrocyte sedimentation rate; Westergren method; ESR; evaluation; comparison;
D O I
10.1080/00365513.2021.1881996
中图分类号
R-3 [医学研究方法]; R3 [基础医学];
学科分类号
1001 ;
摘要
The erythrocyte sedimentation rate (ESR) is a traditional nonspecific laboratory test used for the assessment of inflammation. Even if its usefulness is nowadays being largely debated, it is still considered a valuable laboratory test in selected clinical conditions, such as rheumatoid diseases, orthopedic infections and Hodgkin's lymphoma, and it can be used for the infectious, inflammatory, malignancies, and autoimmune diseases follow-up. The introduction of new methodologies on semi-automated and automated analyzers started about four decades ago and opened a new era of ESR analysis characterized by shorter assay time, use of (EDTA) undiluted blood, that increases sample stability and allows using a single sample for also other hematologic tests, and greater safety for laboratory personnel. In this context, the aim of this study was to evaluate the performances of new device Diesse Cube 30 touch, comparing it with Alifax Test 1 and with the gold standard Westergren method. The new Diesse Cube 30 touch for determination of the ESR shows a good correlation with the manual Westergren gold standard method in a shorter time, and in a standardized way, since all the phases of the test are automatized. The Diesse Cube 30 touch respect the manual gold standard method, displayed a small bias to confirm that the new automated test system tended to have a small bias for ESR values (mean positive bias +0.2 mm/h). The findings of the present study show that the Diesse Cube 30 touch Westergren-based method can be a valid alternative in laboratory analysis for the determination of ESR.
引用
收藏
页码:181 / 186
页数:6
相关论文
共 46 条
  • [1] The Role of Erythrocyte Sedimentation Rate (ESR) in Myeloproliferative and Lymphoproliferative Diseases: Comparison between DIESSE CUBE 30 TOUCH and Alifax Test 1
    Pelagalli, Martina
    Tomassetti, Flaminia
    Nicolai, Eleonora
    Giovannelli, Alfredo
    Codella, Silvia
    Iozzo, Mariannina
    Massoud, Renato
    Secchi, Roberto
    Venditti, Adriano
    Pieri, Massimo
    Bernardini, Sergio
    Hatzimichael, Eleftheria
    DISEASES, 2023, 11 (04)
  • [2] Analytical validation of the modified Westergren method on the automated erythrocyte sedimentation rate analyzer CUBE 30 touch
    Lapic, Ivana
    Rade, Anamarija
    Kraljevic, Andela
    Milos, Marija
    Herak, Desiree Coen
    Daskijevic, Lucija
    Cerovac, Paula
    Rogic, Dunja
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2023, 61 (08) : 1463 - 1469
  • [3] Method Comparison of Erythrocyte Sedimentation Rate Automated Systems, the VES-MATIC 5 (DIESSE) and Test 1 (ALIFAX), with the Reference Method in Routine Practice
    Cennamo, Michele
    Giuliano, Loredana
    Arrigoni, Gloria
    Fardone, Valentina
    Russo, Roberta
    De Tomasi, Luca Maria
    Bertani, Fabio
    Cammarota, Gaetano
    Brunetti, Giovanni
    Del Vecchio, Lucia
    Partenope, Michelarcangelo
    JOURNAL OF CLINICAL MEDICINE, 2024, 13 (03)
  • [4] ISED, CUBE 30 touch and westergren: Comparison of different methodologies for measuring erythrocyte sedimentation rate (ESR)
    Miletic, M.
    Barberic, M.
    CLINICA CHIMICA ACTA, 2024, 558
  • [5] Ves-Matic CUBE 200: is modified Westergren method for erythrocyte sedimentation rate a valid alternative to the gold standard?
    Lapic, Ivana
    Piva, Elisa
    Spolaore, Federica
    Musso, Giulia
    Tosato, Francesca
    Pelloso, Michela
    Plebani, Mario
    JOURNAL OF CLINICAL PATHOLOGY, 2019, 72 (10) : 716 - 719
  • [6] COMPARISON OF ERYTHROCYTE SEDIMENTATION RATE METHODOLOGIES: ALIFAX TEST 1 THL, VES-MATIC 20 AND WESTERGREN
    Rodrigues, M. C.
    Stebbeds, W. J. D.
    Barracha, A. S.
    Lapa, A. S.
    Vassal, R.
    Silva, E.
    Freitas, A. M. M. S.
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2011, 49 : S473 - S473
  • [7] Comparison of Erythrocyte Sedimentation Rate Results of Test-1 and Automatic Westergren Device with Reference Westergren Method
    Sonmez, Cigdem
    Guntas, Gulcan
    Kaymak, Aysegul Ozturk
    Nedim, Akkaya
    Akin, Kadir Okhan
    GAZI MEDICAL JOURNAL, 2014, 25 (02): : 51 - 53
  • [8] Comparison of the erythrocyte sedimentation rate measured by the Micro Test 1 - Sedimentation Analyzer and the conventional Westergren method
    Arikan, Serap
    Akalin, Nalan
    ANNALS OF SAUDI MEDICINE, 2007, 27 (05) : 362 - 365
  • [9] Evaluation of an automated erythrocyte sedimentation rate analyzer as compared to the Westergren manual method in measurement of erythrocyte sedimentation rate
    Subramanian, Arulselvi
    Rangarajan, Kanchana
    Pandey, Ravindra Mohan
    Gandhi, Jatin S.
    Sharma, Vijay
    Bhoi, Sanjeev Kumar
    INDIAN JOURNAL OF PATHOLOGY AND MICROBIOLOGY, 2011, 54 (01) : 70 - 74
  • [10] Evaluation of the Ves-Matic Cube 200 Erythrocyte Sedimentation Method Comparison With Westergren-Based Methods
    Curvers, Joyce
    Kooren, Jurgen
    Laan, Maartje
    van Lierop, Edwin
    van de Kerkhof, Daan
    Scharnhorst, Volkher
    Herruer, Martien
    AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 2010, 134 (04) : 653 - 660