Analytical validation of the modified Westergren method on the automated erythrocyte sedimentation rate analyzer CUBE 30 touch

被引:1
|
作者
Lapic, Ivana [1 ]
Rade, Anamarija [2 ]
Kraljevic, Andela [3 ]
Milos, Marija [4 ]
Herak, Desiree Coen [3 ]
Daskijevic, Lucija [3 ]
Cerovac, Paula [3 ]
Rogic, Dunja [3 ,5 ]
机构
[1] Univ Hosp Ctr Zagreb, Dept Lab Diagnost, Kispaticeva 12, Zagreb 10000, Croatia
[2] Gen Hosp Varazdin, Lab Med Biochem, Varazhdin, Croatia
[3] Univ Hosp Ctr Zagreb, Dept Lab Diagnost, Zagreb, Croatia
[4] Univ Mostar, Fac Pharm, Mostar, Bosnia & Herceg
[5] Univ Zagreb, Fac Pharm & Biochem, Zagreb, Croatia
关键词
automation; erythrocyte sedimentation rate; validation study; Westergren method; SYSTEM;
D O I
10.1515/cclm-2023-0033
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Objectives: Analytical validation of automated erythrocyte sedimentation rate (ESR) analyzers is necessary prior to their implementation into routine practice. Our aim was to perform the analytical validation of the modified Westergren method applied on the CUBE 30 touch analyzer (Diesse, Siena, Italy). Methods: Validation included determination of within-run and between-run precision following the Clinical and Laboratory Standards Institute EP15-A3 protocol, comparison with the reference Westergren method, sample stability assessment at both room temperature and 4 degrees C, after 4, 8 and 24-h storage, and checking the extent of hemolysis and lipemia interference. Results: Coefficients of variation (CVs) for within-run precision were 5.2% for the normal and 2.6% for the abnormal range, while between-run CVs were 9.4 and 2.2%, respectively. Comparison with the Westergren method (n=191) yielded Spearman's correlation coefficient of 0.93, no constant nor proportional difference [y=0.4 (95% CI: -1.7-1.0) + 1.06 (95% CI: 1.00-1.14)x] and a non-significant mean absolute bias of -2.6 mm (95% CI: -5.3-0.2). Lower comparability was evidenced with increasing ESR values, with both constant and proportional differences for ESR values between 40 and 80 mm, and above 80 mm. Sample stability was not compromised up to 8-h storage both at room temperature (p=0.054) and 4 degrees C (p=0.421). Hemolysis did not affect ESR measurement up to 1.0 g/L of free hemoglobin (p=0.089), while lipemia index above 5.0 g/L affects the ESR result (p=0.004). Conclusions: This study proved that CUBE 30 touch provides reliable ESR measurement and satisfactory comparability with the reference Westergren methods, with minor variation related to methodological differences.
引用
收藏
页码:1463 / 1469
页数:7
相关论文
共 49 条
  • [1] VALIDATION OF AN ERYTHROCYTE SEDIMENTATION RATE ANALYZER WITH MODIFIED WESTERGREN METHOD
    Patil, Asha
    Nayak, Deepak M.
    Belurkar, Sushma V.
    Verma, Seemitr
    JOURNAL OF EVOLUTION OF MEDICAL AND DENTAL SCIENCES-JEMDS, 2013, 2 (03): : 284 - 289
  • [2] Evaluation of an automated erythrocyte sedimentation rate analyzer as compared to the Westergren manual method in measurement of erythrocyte sedimentation rate
    Subramanian, Arulselvi
    Rangarajan, Kanchana
    Pandey, Ravindra Mohan
    Gandhi, Jatin S.
    Sharma, Vijay
    Bhoi, Sanjeev Kumar
    INDIAN JOURNAL OF PATHOLOGY AND MICROBIOLOGY, 2011, 54 (01) : 70 - 74
  • [3] Analytical validation of the iSED automated analyzer for erythrocyte sedimentation rate
    Lapic, Ivana
    Milos, Marija
    Tosato, Francesca
    Piva, Elisa
    Zadro, Renata
    Rogic, Dunja
    Plebani, Mario
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2020, 42 (02) : 109 - 115
  • [4] ISED, CUBE 30 touch and westergren: Comparison of different methodologies for measuring erythrocyte sedimentation rate (ESR)
    Miletic, M.
    Barberic, M.
    CLINICA CHIMICA ACTA, 2024, 558
  • [5] AN AUTOMATED-METHOD FOR RECORDING THE WESTERGREN ERYTHROCYTE SEDIMENTATION-RATE
    KING, JF
    KENNEDY, K
    RIMMER, AR
    JOURNAL OF CLINICAL PATHOLOGY, 1981, 34 (04) : 449 - 452
  • [6] Evaluation of the Diesse Cube 30 touch erythrocyte sedimentation method in comparison with Alifax test 1 and the manual Westergren gold standard method
    Pieri, Massimo
    Pignalosa, Stefano
    Perrone, Marco Alfonso
    Russo, Carmelo
    Noce, Gianluca
    Perrone, Andrea
    Terrinoni, Alessandro
    Massoud, Renato
    Bernardini, Sergio
    SCANDINAVIAN JOURNAL OF CLINICAL & LABORATORY INVESTIGATION, 2021, 81 (03): : 181 - 186
  • [7] Is Vision C interchangeable with the modified Westergren method for the erythrocyte sedimentation rate?
    Erdogan, Serpil
    Firat, Ridvan
    Avcioglu, Gamze
    Yilmaz, Gulsen
    Erel, Ozcan
    Yilmaz, Fatma Meric
    TURKISH JOURNAL OF BIOCHEMISTRY-TURK BIYOKIMYA DERGISI, 2022, 47 (04): : 403 - 408
  • [8] Comparison of the erythrocyte sedimentation rate measured by the Micro Test 1 - Sedimentation Analyzer and the conventional Westergren method
    Arikan, Serap
    Akalin, Nalan
    ANNALS OF SAUDI MEDICINE, 2007, 27 (05) : 362 - 365
  • [9] Challenges of preanalytical variables in erythrocyte sedimentation rate: a CUBE 30 touch evaluation
    Tomassetti, Flaminia
    Guerranti, Roberto
    Leoncini, Roberto
    Pieroni, Carolina
    Diamanti, Daniela
    Cirianni, Michele
    Silvestrini, Caterina
    Galasso, Lucrezia
    Pelagalli, Martina
    Nicolai, Eleonora
    Giovannelli, Alfredo
    Pieri, Massimo
    Bernardini, Sergio
    SCANDINAVIAN JOURNAL OF CLINICAL & LABORATORY INVESTIGATION, 2024, 84 (7-8): : 477 - 485
  • [10] Ves-Matic CUBE 200: is modified Westergren method for erythrocyte sedimentation rate a valid alternative to the gold standard?
    Lapic, Ivana
    Piva, Elisa
    Spolaore, Federica
    Musso, Giulia
    Tosato, Francesca
    Pelloso, Michela
    Plebani, Mario
    JOURNAL OF CLINICAL PATHOLOGY, 2019, 72 (10) : 716 - 719