Analytical validation of the modified Westergren method on the automated erythrocyte sedimentation rate analyzer CUBE 30 touch

被引:1
|
作者
Lapic, Ivana [1 ]
Rade, Anamarija [2 ]
Kraljevic, Andela [3 ]
Milos, Marija [4 ]
Herak, Desiree Coen [3 ]
Daskijevic, Lucija [3 ]
Cerovac, Paula [3 ]
Rogic, Dunja [3 ,5 ]
机构
[1] Univ Hosp Ctr Zagreb, Dept Lab Diagnost, Kispaticeva 12, Zagreb 10000, Croatia
[2] Gen Hosp Varazdin, Lab Med Biochem, Varazhdin, Croatia
[3] Univ Hosp Ctr Zagreb, Dept Lab Diagnost, Zagreb, Croatia
[4] Univ Mostar, Fac Pharm, Mostar, Bosnia & Herceg
[5] Univ Zagreb, Fac Pharm & Biochem, Zagreb, Croatia
关键词
automation; erythrocyte sedimentation rate; validation study; Westergren method; SYSTEM;
D O I
10.1515/cclm-2023-0033
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Objectives: Analytical validation of automated erythrocyte sedimentation rate (ESR) analyzers is necessary prior to their implementation into routine practice. Our aim was to perform the analytical validation of the modified Westergren method applied on the CUBE 30 touch analyzer (Diesse, Siena, Italy). Methods: Validation included determination of within-run and between-run precision following the Clinical and Laboratory Standards Institute EP15-A3 protocol, comparison with the reference Westergren method, sample stability assessment at both room temperature and 4 degrees C, after 4, 8 and 24-h storage, and checking the extent of hemolysis and lipemia interference. Results: Coefficients of variation (CVs) for within-run precision were 5.2% for the normal and 2.6% for the abnormal range, while between-run CVs were 9.4 and 2.2%, respectively. Comparison with the Westergren method (n=191) yielded Spearman's correlation coefficient of 0.93, no constant nor proportional difference [y=0.4 (95% CI: -1.7-1.0) + 1.06 (95% CI: 1.00-1.14)x] and a non-significant mean absolute bias of -2.6 mm (95% CI: -5.3-0.2). Lower comparability was evidenced with increasing ESR values, with both constant and proportional differences for ESR values between 40 and 80 mm, and above 80 mm. Sample stability was not compromised up to 8-h storage both at room temperature (p=0.054) and 4 degrees C (p=0.421). Hemolysis did not affect ESR measurement up to 1.0 g/L of free hemoglobin (p=0.089), while lipemia index above 5.0 g/L affects the ESR result (p=0.004). Conclusions: This study proved that CUBE 30 touch provides reliable ESR measurement and satisfactory comparability with the reference Westergren methods, with minor variation related to methodological differences.
引用
收藏
页码:1463 / 1469
页数:7
相关论文
共 49 条
  • [21] Comparison of the StaRRsed Interliner device with Westergren method in erythrocyte sedimentation rate measurement
    Orkmez, Mustafa
    Orhan, Seren
    Bozdayi, Mehmet Akif
    Tarakcioglu, Mehmet
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2021, 43 (04) : 616 - 622
  • [22] Measurement of Erythrocyte Sedimentation Rate and its Comparison by Microsed System and Westergren Method
    Rehman, Asma
    Farrukh, Ambreen
    Choudary, Munazza
    PAKISTAN JOURNAL OF MEDICAL & HEALTH SCIENCES, 2019, 13 (02): : 243 - 245
  • [23] Comparison of Erythrocyte Sedimentation Rate Results of Test-1 and Automatic Westergren Device with Reference Westergren Method
    Sonmez, Cigdem
    Guntas, Gulcan
    Kaymak, Aysegul Ozturk
    Nedim, Akkaya
    Akin, Kadir Okhan
    GAZI MEDICAL JOURNAL, 2014, 25 (02): : 51 - 53
  • [24] Validation of the Diesse Mini-ves automated system using the Westergren erythrocyte sedimentation rate in patients with systemic inflammatory conditions.
    Happe, MR
    Rennie, TA
    Murphy, FT
    Dooley, DP
    Battafarano, DF
    Casey, TJ
    ARTHRITIS AND RHEUMATISM, 2000, 43 (09): : S149 - S149
  • [25] Evaluation of the Ves-Matic Cube 200 Erythrocyte Sedimentation Method Comparison With Westergren-Based Methods
    Curvers, Joyce
    Kooren, Jurgen
    Laan, Maartje
    van Lierop, Edwin
    van de Kerkhof, Daan
    Scharnhorst, Volkher
    Herruer, Martien
    AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 2010, 134 (04) : 653 - 660
  • [26] COMPARATIVE ANALYSIS OF TEST-1 ERYTHROCYTE SEDIMENTATION RATE ANALYZER VERSUS WESTERGREN METHOD IN EVALUATING NORMAL AND ANEMIC BLOOD SAMPLES
    Prompetchara, Eakachai
    Chaiwong, Kwanlada
    Ketloy, Chutitorn
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2024, 46 : 82 - 82
  • [27] Validation of the New MINI-CUBE for Clinic Determination of Erythrocyte Sedimentation Rate
    Tomassetti, Flaminia
    Pelagalli, Martina
    Nicolai, Eleonora
    Sarubbi, Serena
    Calabrese, Cinzia
    Giovannelli, Alfredo
    Codella, Silvia
    Viola, Gemma
    Diamanti, Daniela
    Massoud, Renato
    Bernardini, Sergio
    Pieri, Massimo
    JOURNAL OF HEMATOLOGY, 2023, 12 (05) : 208 - 214
  • [28] Method comparison of automated systems for the erythrocyte sedimentation rate
    Giavarina, D
    Dall'Olio, G
    Soffiati, G
    AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 1999, 112 (05) : 721 - 722
  • [29] Evaluation of the InteRRliner automated erythrocyte sedimentation rate analyzer for a large academic laboratory
    Schapkaitz, E.
    RabuRabu, S.
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2017, 39 (03) : E66 - E69
  • [30] The Role of Erythrocyte Sedimentation Rate (ESR) in Myeloproliferative and Lymphoproliferative Diseases: Comparison between DIESSE CUBE 30 TOUCH and Alifax Test 1
    Pelagalli, Martina
    Tomassetti, Flaminia
    Nicolai, Eleonora
    Giovannelli, Alfredo
    Codella, Silvia
    Iozzo, Mariannina
    Massoud, Renato
    Secchi, Roberto
    Venditti, Adriano
    Pieri, Massimo
    Bernardini, Sergio
    Hatzimichael, Eleftheria
    DISEASES, 2023, 11 (04)