The Role of Erythrocyte Sedimentation Rate (ESR) in Myeloproliferative and Lymphoproliferative Diseases: Comparison between DIESSE CUBE 30 TOUCH and Alifax Test 1

被引:4
|
作者
Pelagalli, Martina [1 ,2 ]
Tomassetti, Flaminia [1 ,2 ]
Nicolai, Eleonora [1 ]
Giovannelli, Alfredo [1 ,2 ]
Codella, Silvia [1 ,2 ]
Iozzo, Mariannina [2 ]
Massoud, Renato [1 ,2 ]
Secchi, Roberto [3 ]
Venditti, Adriano [3 ]
Pieri, Massimo [1 ,2 ]
Bernardini, Sergio [1 ,2 ]
Hatzimichael, Eleftheria
机构
[1] Univ Roma Tor Vergata, Dept Expt Med, Via Montpellier 1, I-00133 Rome, Italy
[2] Tor Vergata Univ Hosp, Dept Lab Med, Viale Oxford 81, I-00133 Rome, Italy
[3] Univ Tor Vergata, Dept Biomed & Prevent, Hematol, I-00133 Rome, Italy
关键词
ESR; lymphoproliferative diseases; myeloproliferative diseases; cancer; C-REACTIVE PROTEIN; EARLY RELAPSE; INFLAMMATION; SURVIVAL; CANCER; EVOLUTION; UTILITY;
D O I
10.3390/diseases11040169
中图分类号
R-3 [医学研究方法]; R3 [基础医学];
学科分类号
1001 ;
摘要
(1) Background: The erythrocyte sedimentation rate (ESR) is widely diffused in hematology laboratories to monitor inflammatory statuses, response to therapies (such as antibiotics), and oncologic diseases. However, ESR is not a specific diagnostic marker but needs to be contextualized and compared with clinical and other laboratory findings. This study aimed to investigate the performance of two automated instruments, namely the DIESSE CUBE 30 TOUCH (DIESSE, Siena, Italy) and the Alifax Test 1 (Alifax Srl, Polverara, Italy), in comparison with the gold standard, the Westergren method, in lymphoproliferative and myeloproliferative patients. (2) Methods: 97 EDTA samples were selected from the hematology department of Roma Tor Vergata Hospital and analyzed. Statistical analysis was applied. (3) A good correlation between CUBE 30 TOUCH and the gold standard was observed in the overall sample (R2 = 0.90), as well as in patients with lymphoproliferative diseases (R2 = 0.90) and myeloproliferative diseases (R2 = 0.90). The correlation between Test 1 and the gold standard was observed in the overall sample (R2 = 0.68), as well as in patients with lymphoproliferative diseases (R2 = 0.79) and myeloproliferative diseases (R2 = 0.53). (4) Conclusions: The CUBE 30 TOUCH appears to be a more trustworthy tool for evaluating ESR in these pathologies.
引用
收藏
页数:12
相关论文
共 19 条
  • [1] Evaluation of the Diesse Cube 30 touch erythrocyte sedimentation method in comparison with Alifax test 1 and the manual Westergren gold standard method
    Pieri, Massimo
    Pignalosa, Stefano
    Perrone, Marco Alfonso
    Russo, Carmelo
    Noce, Gianluca
    Perrone, Andrea
    Terrinoni, Alessandro
    Massoud, Renato
    Bernardini, Sergio
    SCANDINAVIAN JOURNAL OF CLINICAL & LABORATORY INVESTIGATION, 2021, 81 (03): : 181 - 186
  • [2] ISED, CUBE 30 touch and westergren: Comparison of different methodologies for measuring erythrocyte sedimentation rate (ESR)
    Miletic, M.
    Barberic, M.
    CLINICA CHIMICA ACTA, 2024, 558
  • [3] Method Comparison of Erythrocyte Sedimentation Rate Automated Systems, the VES-MATIC 5 (DIESSE) and Test 1 (ALIFAX), with the Reference Method in Routine Practice
    Cennamo, Michele
    Giuliano, Loredana
    Arrigoni, Gloria
    Fardone, Valentina
    Russo, Roberta
    De Tomasi, Luca Maria
    Bertani, Fabio
    Cammarota, Gaetano
    Brunetti, Giovanni
    Del Vecchio, Lucia
    Partenope, Michelarcangelo
    JOURNAL OF CLINICAL MEDICINE, 2024, 13 (03)
  • [4] Challenges of preanalytical variables in erythrocyte sedimentation rate: a CUBE 30 touch evaluation
    Tomassetti, Flaminia
    Guerranti, Roberto
    Leoncini, Roberto
    Pieroni, Carolina
    Diamanti, Daniela
    Cirianni, Michele
    Silvestrini, Caterina
    Galasso, Lucrezia
    Pelagalli, Martina
    Nicolai, Eleonora
    Giovannelli, Alfredo
    Pieri, Massimo
    Bernardini, Sergio
    SCANDINAVIAN JOURNAL OF CLINICAL & LABORATORY INVESTIGATION, 2024, 84 (7-8): : 477 - 485
  • [5] MEASUREMENT OF ERYTHROCYTE SEDIMENTATION RATE (ESR) IN EDTA AND CITRATE BLOOD - COMPARISON OF VESMATIC 30, VESMATIC CUBE 30 AND VESMATIC CUBE 80 ANALYZERS
    Perovic, E.
    Sambunjak, J.
    Coza, M.
    Valcic, A.
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2011, 49 : S481 - S481
  • [6] COMPARISON OF ERYTHROCYTE SEDIMENTATION RATE METHODOLOGIES: ALIFAX TEST 1 THL, VES-MATIC 20 AND WESTERGREN
    Rodrigues, M. C.
    Stebbeds, W. J. D.
    Barracha, A. S.
    Lapa, A. S.
    Vassal, R.
    Silva, E.
    Freitas, A. M. M. S.
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2011, 49 : S473 - S473
  • [7] Analytical validation of the modified Westergren method on the automated erythrocyte sedimentation rate analyzer CUBE 30 touch
    Lapic, Ivana
    Rade, Anamarija
    Kraljevic, Andela
    Milos, Marija
    Herak, Desiree Coen
    Daskijevic, Lucija
    Cerovac, Paula
    Rogic, Dunja
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2023, 61 (08) : 1463 - 1469
  • [8] The effect of paraproteins on the erythrocyte sedimentation rate: a comparison between the StarrSed and TEST 1
    Raijmakers, Maarten T. M.
    Kuijper, Philip H. M.
    Bakkeren, Dirk L.
    Vader, Huib L.
    ANNALS OF CLINICAL BIOCHEMISTRY, 2008, 45 : 593 - 597
  • [9] Correlation Between ABO Blood Grouping and Erythrocyte Sedimentation Rate (ESR) in Periodontal Diseases
    Durge, Khushboo J.
    Patil, Ruchita T.
    Oza, Ranu R.
    Dhadse, Prasad, V
    Salian, Shrishti S.
    CUREUS JOURNAL OF MEDICAL SCIENCE, 2024, 16 (07)
  • [10] Comparison of Erythrocyte Sedimentation Rate (ESR) Between the Alcor iSED ESR Analyzer and the Streck Autoplus ESR Analyzer
    Biederman, Laura
    Gulati, Gene
    Gong, Jerald
    AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 2014, 142