Magnetic resonance imaging (MRI) is one of the most powerful imaging modality in clinics and is essential for the diagnosis of strokes through carotid artery imaging. The limiting factor for high-quality MRI is the signal-to-noise ratio (SNR) performance of the radio frequency (RF) coils. The current RF surface coils, however, are made of rigid or semiflexible materials with very limited bending properties. As a result, their SNR is limited because they cannot be placed very close to the imaging area, thus receiving noises from parts of the human body, which are not intended to be imaged. Taking advantage of the computerized embroidery and laser cutting technology, in this paper, we utilize electrotextile to design, fabricate, and measure multilayer RF coil array system for 3 Tesla (3T) MRI to improve the SNR performance. The proposed RF coil array system provides an ergonomic and high-performance solution to the 3T MRI systems. A roadmap to systematically design electrotextile RF coil arrays is proposed. RF coil array is characterized to have the accurate resonant frequency, good impedance matching, and low mutual coupling. In addition, magnetic field distribution, bending effects, and human body effects are also discussed. A systematic method to characterize the performance of the electrotextile pattern is studied and used to assist the development and performance characterization. Finally, the high resolution and high SNR images of various kinds of phantoms are obtained using the University of California at Los Angeles (UCLA) Antenna Lab electrotextile coil array after its integration with the 3T MRI scanners at UCLA David Geffen School of Medicine Translational Research Imaging Center. Compared with the conventional surface coil, more than 10 dB SNR increase is observed at the depth of 0.5 cm and 3 dB increase at the depth of 3 cm.
引用
收藏
页码:5115 / 5125
页数:11
相关论文
共 50 条
  • [41] High-Resolution Uniform MR Imaging of Finger Joints Using a Dedicated RF Coil at 3T
    Kwok, Wingchi E.
    You, Zhigang
    Monu, Johnny
    Seo, Gwysuk
    Ritchlin, Christopher
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2010, 31 (01) : 240 - 247
  • [42] A Novel Three-Channel Endorectal Coil for Prostate Magnetic Resonance Imaging at 3T
    Mo, Zhiguang
    Zhang, Xiaoping
    Liang, Huageng
    Chen, Qiaoyan
    Tie, Changjun
    Xiao, Wen
    Cao, Qi
    Liu, Chenchen
    Zou, Chao
    Wan, Liwen
    Zhang, Xiaoliang
    Li, Ye
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2023, 70 (12) : 3381 - 3388
  • [43] Design of a novel antisymmetric coil array for parallel transmit cardiac MRI in pigs at 7 T
    Elabyad, Ibrahim A.
    Terekhov, M.
    Stefanescu, M. R.
    Lohr, D.
    Fischer, M.
    Schreiber, L. M.
    JOURNAL OF MAGNETIC RESONANCE, 2019, 305 : 195 - 208
  • [44] Size-Adaptive Occipital 18-Channel Receive-Only RF Coil for 3T MRI
    Mathieu, William
    Popovic, Milica
    Farivar, Reza
    IEEE JOURNAL OF ELECTROMAGNETICS RF AND MICROWAVES IN MEDICINE AND BIOLOGY, 2024,
  • [45] Impacts of RF Shimming on Local SAR Caused by MRI 3T Birdcage Coil near Femoral Plate Implants
    Zeng, Qi
    Guo, Ran
    Zheng, Jianfeng
    Chen, Ji
    2017 IEEE INTERNATIONAL SYMPOSIUM ON ANTENNAS AND PROPAGATION & USNC/URSI NATIONAL RADIO SCIENCE MEETING, 2017, : 1005 - 1006
  • [46] Dual Resonant RF Coil for 1.5T and 3T MRI Systems Employing FPCB Etched Sub-legs
    Ahmad, Sheikh Faisal
    Son, Hyeok Woo
    Choi, Ick Chang
    Kim, Young Cheol
    Chang, Yongmin
    Kim, Hyun Deok
    ASIA-PACIFIC MICROWAVE CONFERENCE 2011, 2011, : 1442 - 1445
  • [47] Design and Development of a General-Purpose Transmit/Receive (T/R) Switch for 3T MRI, Compatible for a Linear, Quadrature and Double-Tuned RF Coil
    Thapa, Bijaya
    Kaggie, Joshua
    Sapkota, Nabraj
    Frank, Deborah
    Jeong, Eun-Kee
    CONCEPTS IN MAGNETIC RESONANCE PART B-MAGNETIC RESONANCE ENGINEERING, 2016, 46 (02) : 57 - 65
  • [48] 3T MR imaging of the orbits, optic nerves and chiasm using a hybrid phased-array coil
    Lane, JI
    Kocharian, A
    Witte, RJ
    Felmlee, JP
    Watson, R
    RADIOLOGY, 2002, 225 : 564 - 564
  • [49] 3T Multiparametric MRI (3T mMRI) With a Phased-Array Torso Coil in Staging of Prostate Cancer Patients (PCP) Prior to Radical Prostatectomy: Preliminary Results
    Counago, F.
    Recio, M.
    Del Cerro, E.
    Diaz, A. A.
    Marcos, F. J.
    Murillo, R.
    Rodriguez, J. M.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2014, 90 : S409 - S410
  • [50] Development of a Solenoid RF Coil for Animal Imaging in 3 T High-Magnetic-Field MRI
    Lee, Hong-Seok
    Woo, Dong-Cheol
    Min, Kwang-Hong
    Kim, Yong-Kwon
    Lee, Heung-Kyu
    Choe, Bo-Young
    SCANNING, 2008, 30 (05) : 419 - 425