The rapid evaluation of bacterial growth in blood cultures by selected ion flow tube-mass spectrometry (SIFT-MS) and comparison with the BacT/ALERT automated blood culture system
被引:57
|作者:
Scotter, J. M.
Allardyce, R. A.
Langford, Vs.
Hill, A.
Murdoch, D. R.
机构:
[1] Syft Technol Ltd, Med Applicat, Christchurch, New Zealand
[2] Univ Otago, Christchurch Sch Med & Hlth Sci, Dunedin, New Zealand
[3] Canterbury Hlth Labs, Microbiol Unit, Christchurch, New Zealand
关键词:
blood cultures;
selected ion flow tube-mass spectrometry;
BacT/ALERT;
D O I:
10.1016/j.mimet.2005.09.016
中图分类号:
Q5 [生物化学];
学科分类号:
071010 ;
081704 ;
摘要:
We compared the performance of the BacT/ALERT automated blood culture system with real-time, quantitative volatile organic compound (VOC) detection by selected ion flow tube-mass spectrometry (SIFT-MS). Blood samples from healthy donors were artificially infected with 5 or 100 CFU of organisms commonly causing bacteremia. Positive results by SIFT-MS analysis of headspace gases were recorded for 53/60 (88.3%) bottles at 8h, and 58/60 (96.6%) bottles at 24 h. We conclude that SIFT-MS is a sensitive method for the detection of microbial VOCs. Furthermore, profiles of the VOCs detected may allow simultaneous identification of infecting organisms. (c) 2005 Elsevier B.V. All rights reserved.