Ballistocardiogram as Proximal Timing Reference for Pulse Transit Time Measurement: Potential for Cuffless Blood Pressure Monitoring

被引:99
|
作者
Kim, Chang-Sei [1 ]
Carek, Andrew M. [2 ]
Mukkamala, Ramakrishna [3 ]
Inan, Omer T. [2 ]
Hahn, Jin-Oh [1 ]
机构
[1] Univ Maryland, Dept Mech Engn, College Pk, MD 20742 USA
[2] Georgia Inst Technol, Sch Elect & Comp Engn, Atlanta, GA 30332 USA
[3] Michigan State Univ, Dept Elect & Comp Engn, E Lansing, MI 48824 USA
基金
美国国家卫生研究院;
关键词
Ballistocardiogram (BCG); blood pressure; electrocardiogram (ECG); pulse arrival time (PAT); pulse transit time (PTT); ARRIVAL-TIME; PREEJECTION PERIOD; HEART-RATE; CONTRACTILITY; CALIBRATION; MARKER; ECG;
D O I
10.1109/TBME.2015.2440291
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Goal: We tested the hypothesis that the ballistocardiogram (BCG) waveform could yield a viable proximal timing reference for measuring pulse transit time (PTT). Methods: From 15 healthy volunteers, we measured PTT as the time interval between BCG and a noninvasively measured finger blood pressure (BP) waveform. To evaluate the efficacy of the BCG-based PTT in estimating BP, we likewise measured pulse arrival time (PAT) using the electrocardiogram (ECG) as proximal timing reference and compared their correlations to BP. Results: BCG-based PTT was correlated with BP reasonably well: the mean correlation coefficient (r) was 0.62 for diastolic (DP), 0.65 for mean (MP), and 0.66 for systolic (SP) pressures when the intersecting tangent method was used as distal timing reference. Comparing four distal timing references (intersecting tangent, maximum second derivative, diastolic minimum, and systolic maximum), PTT exhibited the best correlation with BP when the systolic maximum method was used (mean r value was 0.66 for DP, 0.67 for MP, and 0.70 for SP). PTT was more strongly correlated with DP than PAT regardless of the distal timing reference: mean r value was 0.62 versus 0.51 (p = 0.07) for intersecting tangent, 0.54 versus 0.49 (p = 0.17) for maximum second derivative, 0.58 versus 0.52 (p = 0.37) for diastolic minimum, and 0.66 versus 0.60 (p = 0.10) for systolic maximum methods. The difference between PTT and PAT in estimating DP was significant (p = 0.01) when the r values associated with all the distal timing references were compared altogether. However, PAT appeared to outperform PTT in estimating SP (p = 0.31 when the r values associated with all the distal timing references were compared altogether). Conclusion: We conclude that BCG is an adequate proximal timing reference in deriving PTT, and that BCG-based PTT may be superior to ECG-based PAT in estimating DP. Significance: PTT with BCG as proximal timing reference has potential to enable convenient and ubiquitous cuffless BP monitoring.
引用
收藏
页码:2657 / 2664
页数:8
相关论文
共 50 条
  • [31] PULSE TRANSIT TIME FOR BLOOD PRESSURE MEASUREMENT UNDER POSITIVE AIRWAY PRESSURE VENTILATION
    Schmalgemeier, Heidi
    Bitter, Thomas
    Bartsch, Stephan
    Bullert, Kevin
    Eckert, Siegfried
    Horstkotte, Dieter
    Oldenburg, Olaf
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2011, 57 (14) : E619 - E619
  • [32] Towards Ubiquitous Blood Pressure Monitoring in an Armband using Pulse Transit Time
    Toreyin, Hakan
    Javaid, Abdul Q.
    Ashouri, Hazar
    Ode, Oludotun
    Inan, Omer T.
    2015 IEEE BIOMEDICAL CIRCUITS AND SYSTEMS CONFERENCE (BIOCAS), 2015, : 560 - 563
  • [33] Adaptive Algorithm for Continuous Monitoring of Blood Pressure Using a Pulse Transit Time
    Jadooei, Ali
    Zaderykhin, O.
    Shulgin, V. I.
    2013 IEEE XXXIII INTERNATIONAL SCIENTIFIC CONFERENCE ELECTRONICS AND NANOTECHNOLOGY (ELNANO), 2013, : 297 - 301
  • [34] A Chair-Based Unobtrusive Cuffless Blood Pressure Monitoring System Based on Pulse Arrival Time
    Tang, Zunyi
    Tamura, Toshiyo
    Sekine, Masaki
    Huang, Ming
    Chen, Wenxi
    Yoshida, Masaki
    Sakatani, Kaoru
    Kobayashi, Hiroshi
    Kanaya, Shigehiko
    IEEE JOURNAL OF BIOMEDICAL AND HEALTH INFORMATICS, 2017, 21 (05) : 1194 - 1205
  • [35] Blood Pressure Measurement from Photo-plethysmography to Pulse Transit Time
    Myint, C. Z.
    Lim, King Hann
    Wong, Kiing Ing
    Gopalai, Alpha Agape
    Oo, Min Zin
    2014 IEEE CONFERENCE ON BIOMEDICAL ENGINEERING AND SCIENCES (IECBES), 2014, : 496 - 501
  • [36] Continuous Systolic Blood Pressure Measurement Based on Improved Pulse Transit Time
    Xie, Meng
    Yang, Bo
    Chen, Cailian
    Guan, Xinping
    Hong, Peijun
    PROCEEDINGS OF THE 35TH CHINESE CONTROL CONFERENCE 2016, 2016, : 5173 - 5178
  • [37] Non-contact Blood Pressure Measurement Based on Pulse Transit Time
    Chen Pengyong
    Zhou Zuofeng
    Zhao Ting
    Cao Jianzhong
    Yang Hongtao
    TENTH INTERNATIONAL CONFERENCE ON DIGITAL IMAGE PROCESSING (ICDIP 2018), 2018, 10806
  • [38] Pulse Arrival Time and Ballistocardiogram Application to Blood Pressure Variability Estimation
    Pinheiro, Eduardo
    Postolache, Octavian
    Girao, Pedro
    MEMEA: 2009 IEEE INTERNATIONAL WORKSHOP ON MEDICAL MEASUREMENTS AND APPLICATIONS, 2009, : 132 - +
  • [39] An Evaluation of the Cuffless Blood Pressure Estimation Based on Pulse Transit Time Technique: a Half Year Study on Normotensive Subjects
    Wong, Mico Yee-Man
    Poon, Carmen Chung-Yan
    Zhang, Yuan-Ting
    CARDIOVASCULAR ENGINEERING, 2009, 9 (01) : 32 - 38
  • [40] Pulse transit time: validation of blood pressure measurement under positive airway pressure ventilation
    Schmalgemeier, Heidi
    Bitter, Thomas
    Bartsch, Stephan
    Bullert, Kevin
    Fischbach, Thomas
    Eckert, Siegfried
    Horstkotte, Dieter
    Oldenburg, Olaf
    SLEEP AND BREATHING, 2012, 16 (04) : 1105 - 1112