Assessment of image quality in abdominal CT: potential dose reduction with model-based iterative reconstruction

被引:42
|
作者
Kataria, Bharti [1 ]
Althen, Jonas Nilsson [2 ]
Smedby, Orjan [3 ]
Persson, Anders [1 ]
Sokjer, Hannibal [4 ]
Sandborg, Michael [2 ]
机构
[1] Linkoping Univ, Dept Radiol, Dept Med & Hlth Sci, Ctr Med Image Sci & Visualizat CMIV, S-58185 Linkoping, Sweden
[2] Linkoping Univ, Dept Med & Hlth Sci, Dept Med Phys, Linkoping, Sweden
[3] KTH Royal Inst, STH, Stockholm, Sweden
[4] Linkoping Univ, Dept Med & Hlth Sci, S-58183 Linkoping, Sweden
关键词
Dose; Computed tomography; Iterative reconstruction; Abdomen; FBP; FILTERED BACK-PROJECTION; COMPUTED-TOMOGRAPHY; ALGORITHM; PHANTOM; MBIR;
D O I
10.1007/s00330-017-5113-4
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
To estimate potential dose reduction in abdominal CT by visually comparing images reconstructed with filtered back projection (FBP) and strengths of 3 and 5 of a specific MBIR. A dual-source scanner was used to obtain three data sets each for 50 recruited patients with 30, 70 and 100% tube loads (mean CTDIvol 1.9, 3.4 and 6.2 mGy). Six image criteria were assessed independently by five radiologists. Potential dose reduction was estimated with Visual Grading Regression (VGR). Comparing 30 and 70% tube load, improved image quality was observed as a significant strong effect of log tube load and reconstruction method with potential dose reduction relative to FBP of 22-47% for MBIR strength 3 (p < 0.001). For MBIR strength 5 no dose reduction was possible for image criteria 1 (liver parenchyma), but dose reduction between 34 and 74% was achieved for other criteria. Interobserver reliability showed agreement of 71-76% (kappa (w) 0.201-0.286) and intra-observer reliability of 82-96% (kappa (w) 0.525-0.783). MBIR showed improved image quality compared to FBP with positive correlation between MBIR strength and increasing potential dose reduction for all but one image criterion. aEuro cent MBIR's main advantage is its de-noising properties, which facilitates dose reduction. aEuro cent MBIR allows for potential dose reduction in relation to FBP. aEuro cent Visual Grading Regression (VGR) produces direct numerical estimates of potential dose reduction. aEuro cent MBIR strengths 3 and 5 dose reductions were 22-34 and 34-74%. aEuro cent MBIR strength 5 demonstrates inferior performance for liver parenchyma.
引用
收藏
页码:2464 / 2473
页数:10
相关论文
共 50 条
  • [41] Model-based iterative reconstruction in pediatric chest CT: assessment of image quality in a prospective study of children with cystic fibrosis
    Mieville, Frederic A.
    Berteloot, Laureline
    Grandjean, Albane
    Ayestaran, Paul
    Gudinchet, Francois
    Schmidt, Sabine
    Brunelle, Francis
    Bochud, Francois O.
    Verdun, Francis R.
    [J]. PEDIATRIC RADIOLOGY, 2013, 43 (05) : 558 - 567
  • [42] Superior objective and subjective image quality of deep learning reconstruction for low-dose abdominal CT imaging in comparison with model-based iterative reconstruction and filtered back projection
    Tamura, Akio
    Mukaida, Eisuke
    Ota, Yoshitaka
    Kamata, Masayoshi
    Abe, Shun
    Yoshioka, Kunihiro
    [J]. BRITISH JOURNAL OF RADIOLOGY, 2021, 94 (1123):
  • [43] Impact of Model-Based Iterative Reconstruction on Image Quality of Contrast-Enhanced Neck CT
    Gaddikeri, S.
    Andre, J. B.
    Benjert, J.
    Hippe, D. S.
    Anzai, Y.
    [J]. AMERICAN JOURNAL OF NEURORADIOLOGY, 2015, 36 (02) : 391 - 396
  • [44] Image Quality and Dose Reduction Evaluation of a New CT Iterative Reconstruction Algorithm Using Model Observers
    Tseng, H.
    Fan, J.
    Kupinski, M.
    Sainath, P.
    Hsieh, J.
    [J]. MEDICAL PHYSICS, 2013, 40 (06)
  • [45] Initial Evaluation of Model-Based Iterative CT Reconstruction Using Standard Image Quality Phantoms
    Gingold, E.
    Dave, J.
    [J]. MEDICAL PHYSICS, 2014, 41 (06)
  • [46] Model-Based Iterative Reconstruction (MBIR) for the Reduction of Metal Artifacts on CT
    Boudabbous, Sana
    Arditi, Daniel
    Paulin, Emilie
    Syrogiannopoulou, Aphrodite
    Becker, Christoph
    Montet, Xavier
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 2015, 205 (02) : 380 - 385
  • [47] COMPARISON OF ADAPTIVE STATISTICAL ITERATIVE RECONSTRUCTION (ASIR™) AND MODEL-BASED ITERATIVE RECONSTRUCTION (VEO™) FOR PAEDIATRIC ABDOMINAL CT EXAMINATIONS: AN OBSERVER PERFORMANCE STUDY OF DIAGNOSTIC IMAGE QUALITY
    Hultenmo, Maria
    Caisander, Hakan
    Mack, Karsten
    Thilander-Klang, Anne
    [J]. RADIATION PROTECTION DOSIMETRY, 2016, 169 (1-4) : 107 - 114
  • [48] Clinical impact of model-based type iterative reconstruction with fast reconstruction time on image quality of low-dose screening chest CT
    Yuki, Hideaki
    Oda, Seitaro
    Utsunomiya, Daisuke
    Funama, Yoshinori
    Kidoh, Masafumi
    Namimoto, Tomohiro
    Katahira, Kazuhiro
    Honda, Keiichi
    Tokuyasu, Shinichi
    Yamashita, Yasuyuki
    [J]. ACTA RADIOLOGICA, 2016, 57 (03) : 295 - 302
  • [49] Reduced Radiation Dose with Model-based Iterative Reconstruction versus Standard Dose with Adaptive Statistical Iterative Reconstruction in Abdominal CT for Diagnosis of Acute Renal Colic
    Fontarensky, Mikael
    Alfidja, Agaicha
    Perignon, Renan
    Schoenig, Arnaud
    Perrier, Christophe
    Mulliez, Aurelien
    Guy, Laurent
    Boyer, Louis
    [J]. RADIOLOGY, 2015, 276 (01) : 156 - 166
  • [50] Image Quality Assessment of Standard- and Low-Dose Chest CT Using Filtered Back Projection, Adaptive Statistical Iterative Reconstruction, and Novel Model-Based Iterative Reconstruction Algorithms
    Vardhanabhuti, Varut
    Loader, Robert J.
    Mitchell, Grant R.
    Riordan, Richard D.
    Roobottom, Carl A.
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 2013, 200 (03) : 545 - 552