摘要
The energy spectra and the dose to a Cu plate/Gd2O2S phosphor portal imaging detector were investigated for monoenergetic incident beams of photons (1.25, 2, and 5 MeV). The Monte Carlo method was used to characterize the influence of the patient/detector geometry, detector material and design, and incident beam energy on the spectral distribution and the dose, at the imaging detector plane. of a photon beam scattered from a water phantom. The results show that radiation equilibrium is lost in the air gap and that, for the geometries studied, this effect led to a reduction in the exit dose of up to 40%. The finding that the effects of the air gap and held size are roughly complementary has led to the hypothesis that an equivalent field size concept may be used to account for intensity and spectral changes arising from air gap variations. The copper plate preferentially attenuates the low-energy scattered photons incident on it, while producing additional annihilation, bremsstrahlung, and scattered photons. As a result, the scatter spectra at the copper surface entrance of the detector differs significantly from that at the Cu/phosphor interface. In addition, the mean scattered photon energy at the interface was observed to be roughly 0.4 MeV higher than the corresponding effective energy for 2 MeV incident beams. A comparison of the dose to various detector materials showed that exit dosimetry errors of up to 24% will occur if it is assumed that the Cu plate/Gd2O2S phosphor detector is water equivalent. (C) 2000 American Association of Physicists in Medicine. [S0094-2405(00)01002-6].
引用
收藏
页码:330 / 339
页数:10
相关论文
共 50 条
  • [1] Monte Carlo studies of metal/phosphor screen in therapeutic X-ray imaging
    Kim, HK
    Cho, GS
    Chung, YH
    Lee, HK
    Yoon, SC
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1999, 422 (1-3): : 713 - 717
  • [2] Monte Carlo simulations of the imaging performance of metal plate/phosphor screens used in radiotherapy
    Kausch, C
    Schreiber, B
    Kreuder, F
    Schmidt, R
    Dössel, O
    MEDICAL PHYSICS, 1999, 26 (10) : 2113 - 2124
  • [3] Modeling Electronic Portal Imaging Device for Dose Reconstruction using Monte Carlo Method
    Jung, J. W.
    Kim, J. O.
    Yeo, I.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2010, 78 (03): : S724 - S724
  • [4] PHOTON SPECTRA IN STANDARD DOSIMETRIC OR IMAGING PHANTOMS CALCULATED WITH MONTE-CARLO METHODS
    PETOUSSI, N
    ZANKL, M
    PANZER, W
    DREXLER, G
    NETTE, P
    RADIATION PROTECTION DOSIMETRY, 1992, 43 (1-4) : 147 - 149
  • [5] EXIT DOSE STUDIES IN MEGAVOLTAGE PHOTON THERAPY
    LAMBERT, GD
    LIVERSAGE, WE
    HIRST, AM
    DOUGHTY, D
    BRITISH JOURNAL OF RADIOLOGY, 1983, 56 (665): : 329 - 334
  • [6] Monte Carlo validation of a portal imager scatter dose model
    Ozard, S
    El-Khatib, E
    PROCEEDINGS OF THE 22ND ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-4, 2000, 22 : 2889 - 2891
  • [7] OPTIMIZATION OF METAL/PHOSPHOR SCREENS FOR ONLINE PORTAL IMAGING
    WOWK, B
    RADCLIFFE, T
    LESZCZYNSKI, KW
    SHALEV, S
    RAJAPAKSHE, R
    MEDICAL PHYSICS, 1994, 21 (02) : 227 - 235
  • [8] MONTE-CARLO CALCULATIONS OF IRRADIATION TEST PHOTON SPECTRA
    WOOLF, S
    BURKE, EA
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1984, 31 (06) : 1089 - 1094
  • [9] Monte Carlo detective quantum efficiency and scatter studies of a metal/a-Se portal detector
    Lachaine, M
    Fallone, BG
    MEDICAL PHYSICS, 1998, 25 (07) : 1186 - 1194
  • [10] Monte Carlo detective quantum efficiency and scatter studies of a metal/a-Se portal detector
    McGill University, Medical Physics Unit, Montreal General Hospital, 1650 Cedar Ave., Montreal, Que. H3G 1A4, Canada
    Med. Phys., 7 I (1186-1194):