Monte Carlo studies of the exit photon spectra and dose to a metal/phosphor portal imaging screen

被引:26
|
作者
Yeboah, C [1 ]
Pistorius, S
机构
[1] Tom Baker Canc Ctr, Dept Med Phys, Calgary, AB T2N 4NA, Canada
[2] Univ Calgary, Dept Phys & Astron, Calgary, AB T2N 2N4, Canada
[3] Canc Care Manitoba, Winnipeg, MB R3E 0V9, Canada
[4] Univ Manitoba, Dept Phys & Astron, Winnipeg, MB R3T 2N2, Canada
[5] Univ Manitoba, Dept Radiol, Winnipeg, MB R3T 0W3, Canada
关键词
exit dose; portal imaging; energy spectrum; mean energy; effective energy;
D O I
10.1118/1.598835
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The energy spectra and the dose to a Cu plate/Gd2O2S phosphor portal imaging detector were investigated for monoenergetic incident beams of photons (1.25, 2, and 5 MeV). The Monte Carlo method was used to characterize the influence of the patient/detector geometry, detector material and design, and incident beam energy on the spectral distribution and the dose, at the imaging detector plane. of a photon beam scattered from a water phantom. The results show that radiation equilibrium is lost in the air gap and that, for the geometries studied, this effect led to a reduction in the exit dose of up to 40%. The finding that the effects of the air gap and held size are roughly complementary has led to the hypothesis that an equivalent field size concept may be used to account for intensity and spectral changes arising from air gap variations. The copper plate preferentially attenuates the low-energy scattered photons incident on it, while producing additional annihilation, bremsstrahlung, and scattered photons. As a result, the scatter spectra at the copper surface entrance of the detector differs significantly from that at the Cu/phosphor interface. In addition, the mean scattered photon energy at the interface was observed to be roughly 0.4 MeV higher than the corresponding effective energy for 2 MeV incident beams. A comparison of the dose to various detector materials showed that exit dosimetry errors of up to 24% will occur if it is assumed that the Cu plate/Gd2O2S phosphor detector is water equivalent. (C) 2000 American Association of Physicists in Medicine. [S0094-2405(00)01002-6].
引用
收藏
页码:330 / 339
页数:10
相关论文
共 50 条
  • [31] AN IMPROVED ACCELERATOR MODEL FOR 3D DOSE RECONSTRUCTION, BASED ON PORTAL IMAGING AND MONTE CARLO SIMULATION.
    Donato, K.
    van Elmpt, W.
    Nijsten, S.
    Lambin, P.
    Verhaegen, F.
    RADIOTHERAPY AND ONCOLOGY, 2009, 92 : S33 - S33
  • [32] Monte Carlo PENRADIO software for dose calculation in medical imaging
    Adrien, Camille
    Noriega, Mercedes Lopez
    Bonniaud, Guillaume
    Bordy, Jean-Marc
    Le Loirec, Cindy
    Poumarede, Benedicte
    SNA + MC 2013 - JOINT INTERNATIONAL CONFERENCE ON SUPERCOMPUTING IN NUCLEAR APPLICATIONS + MONTE CARLO, 2014,
  • [33] Monte Carlo PENRADIO software for dose calculation in medical imaging
    Adrien, Camille
    Noriega, Mercedes Lopez
    Bonniaud, Guillaume
    Bordy, Jean-Marc
    Le Loirec, Cindy
    Poumarede, Benedicte
    2013 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2013,
  • [34] Performance of three-photon PET imaging: Monte Carlo simulations
    Kacperski, K
    Spyrou, NM
    PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (23): : 5679 - 5695
  • [35] Photon propagation model of In Vivo bioluminescent imaging based on Monte Carlo
    Li, Hui
    Tian, Jie
    Wang, Ge
    Ruan Jian Xue Bao/Journal of Software, 2004, 15 (11): : 1709 - 1719
  • [36] Quantum Monte Carlo Studies of Transition Metal Oxides
    Mitas, Lubos
    Kolorenc, Jindrich
    THEORETICAL AND COMPUTATIONAL METHODS IN MINERAL PHYSICS: GEOPHYSICAL APPLICATIONS, 2010, 71 : 137 - 145
  • [37] Monte Carlo dose calculation for photon and electron radiotherapy on dynamically deforming anatomy
    Zobrist, Bjorn
    Bertholet, Jenny
    Frei, Daniel
    Volken, Werner
    Amstutz, Florian
    Stampanoni, Marco F. M.
    Manser, Peter
    Fix, Michael K.
    Loebner, Hannes A.
    MEDICAL PHYSICS, 2025, 52 (02) : 1281 - 1292
  • [38] Simple beam models for Monte Carlo photon beam dose calculations in radiotherapy
    Fix, MK
    Keller, H
    Rüegsegger, P
    Born, EJ
    MEDICAL PHYSICS, 2000, 27 (12) : 2739 - 2747
  • [39] Modelling of electron contamination in clinical photon beams for Monte Carlo dose calculation
    Yang, J
    Li, JS
    Qin, L
    Xiong, W
    Ma, CM
    PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (12): : 2657 - 2673
  • [40] Measurements and Monte Carlo simulations of the fluence and dose characteristics of microplanar photon beams
    Univ of Western Sydney Nepean, Kingswood
    Australas Phys Eng Sci Med, 4 ([d]217-224):