Time-resolved fluoroimmunoassay of plasma and urine O-desmethylangolensin

被引:17
|
作者
L'homme, R
Brouwers, E
Al-Maharik, N
Lapcík, O
Hampl, R
Mikola, H
Wähälä, K
Adlercreutz, H
机构
[1] Univ Helsinki, Folkhalsan Res Ctr, Div Clin Chem, FIN-00014 Helsinki, Finland
[2] Univ Helsinki, Folkhalsan Res Ctr, Inst Prevent Med, FIN-00014 Helsinki, Finland
[3] Univ Helsinki, Dept Chem, Organ Chem Lab, FIN-00014 Helsinki, Finland
[4] Inst Chem, Dept Chem Nat Subst, Prague 16628 6, Czech Republic
[5] Inst Endocrinol, Prague 11694 1, Czech Republic
[6] Wallac Oy, FIN-20101 Turku, Finland
关键词
immunoassay; phytoestrogens; isofavones; urine; O-desmethylangolensin; fluorometry;
D O I
10.1016/S0960-0760(02)00073-0
中图分类号
Q5 [生物化学]; Q7 [分子生物学];
学科分类号
071010 ; 081704 ;
摘要
We present a method for the determination of the phytoestrogen metabolite O-desmethylangolensin (O-DMA) in plasma (serum) and in urine. O-DMA is a metabolite of daidzein, which occurs in soybeans. It has been suggested that isoflavones may afford protection against breast and prostate cancer and therefore, also the metabolites are of interest. The method is based on time-resolved fluoroimmunoassay (TR-FIA) using a europium chelate as a label. After the synthesis of 4"-O-carboxymethyl-O-DMA, this compound is coupled to bovine serum albumin, and then used as antigen in immunization of rabbits. The tracers with the europium chelate are synthesized using the same 4"-O-derivative of the alpha-methyldeoxybenzoin. After enzymatic hydrolysis and ether extraction the immunoassay is carried out by time resolved fluoroimmunoassay (TR-FIA). Cross-reactivity was tested with angolensin, dihydrogenistein, dihydrodaidzein, equol, 6'-OH-angolensin, trans-4-OH-equol, 6'-OH-O-DMA, cis-4-OH-equol and 5-OH-equol. The antiserum cross-reacted only with angolensin. This cross-reactivity seems not to influence the results, which were highly specific. Plasma samples are hydrolyzed and extracted. Urine samples are analyzed directly after hydrolysis without extraction. The correlation coefficient between the plasma TR-FIA results and the GC-MS results was high; r value was 0.985. The correlation coefficient between the urine TR-FIA results and the GC-MS results was high over the entire range of concentrations (0-1500 nmol/l); r value was 0.976, but lower in the low concentration range (0-100 nmol/l), i.e. value was 0.631. The intra-assay coefficients of variation (CVs) for plasma O-DMA concentrations and for urine O-DMA concentrations at three different concentrations varied 2.8-7.7 and 3.0-6.0%, respectively and the inter-assay CVs varied 3.8-8.9 and 4.4-6.6%, respectively. The working range of the plasma and urine O-DMA assays was 0.5-512 nmol/l. (C) 2002 Elsevier Science Ltd. All rights reserved.
引用
收藏
页码:353 / 361
页数:9
相关论文
共 50 条
  • [1] Time-resolved fluoroimmunoassay for equol in plasma and urine
    Brouwers, E
    L'homme, R
    Al-Maharik, N
    Lapcík, O
    Hampl, R
    Wähälä, K
    Mikola, H
    Adlercreutz, H
    [J]. JOURNAL OF STEROID BIOCHEMISTRY AND MOLECULAR BIOLOGY, 2003, 84 (05): : 577 - 588
  • [2] Time-resolved fluoroimmunoassay for plasma enterolactone
    Adlercreutz, H
    Wang, GJJ
    Lapcík, O
    Hampl, R
    Wähälä, K
    Mäkelä, T
    Lusa, K
    Talme, M
    Mikola, H
    [J]. ANALYTICAL BIOCHEMISTRY, 1998, 265 (02) : 208 - 215
  • [3] EUROPIUM LABELED TIME-RESOLVED FLUOROIMMUNOASSAY FOR URINE ALBUMIN
    NISBET, JA
    HUTCHINSON, E
    MAZURKIEWICZ, J
    [J]. ANNALS OF CLINICAL BIOCHEMISTRY, 1993, 30 : 223 - 224
  • [4] TIME-RESOLVED FLUOROIMMUNOASSAY FOR SECRETIN IN HUMAN PLASMA
    ITO, K
    MAEDA, M
    TSUJI, A
    [J]. BUNSEKI KAGAKU, 1992, 41 (12) : 627 - 632
  • [5] Time-resolved fluoroimmunoassay of plasma daidzein and genistein
    Wang, GJ
    Lapcík, O
    Hampl, R
    Uehara, M
    Al-Maharik, N
    Stumpf, K
    Mikola, H
    Wähälä, K
    Adlercreutz, H
    [J]. STEROIDS, 2000, 65 (06) : 339 - 348
  • [6] Changes in the time-resolved fluoroimmunoassay of plasma enterolactone
    Stumpf, K
    Uehara, M
    Nurmi, T
    Adlercreutz, H
    [J]. ANALYTICAL BIOCHEMISTRY, 2000, 284 (01) : 153 - 157
  • [7] Time-resolved fluoroimmunoassay
    Degan, P
    Podestà, A
    Montagnoli, G
    [J]. MOLECULAR BIOTECHNOLOGY, 1999, 13 (03) : 215 - 222
  • [8] Time-resolved fluoroimmunoassay
    Paolo Degan
    Adriano Podestà
    Giorgio Montagnoli
    [J]. Molecular Biotechnology, 1999, 13 : 215 - 222
  • [9] Rapid analysis of phytoestrogens in human urine by time-resolved fluoroimmunoassay
    Uehara, M
    Lapcík, O
    Hampl, R
    Al-Maharik, N
    Mäkelä, T
    Wähälä, K
    Mikola, H
    Adlercreutz, H
    [J]. JOURNAL OF STEROID BIOCHEMISTRY AND MOLECULAR BIOLOGY, 2000, 72 (05): : 273 - 282
  • [10] TIME-RESOLVED FLUOROIMMUNOASSAY OF AFLATOXINS
    DEGAN, P
    MONTAGNOLI, G
    WILD, CP
    [J]. CLINICAL CHEMISTRY, 1989, 35 (12) : 2308 - 2310