Calibration-based thermal error model for articulated arm coordinate measuring machines

被引:70
|
作者
Santolaria, Jorge [1 ]
Yaguee, Jose-Antonio [1 ]
Jimenez, Roberto [1 ]
Aguilar, Juan-Jose [1 ]
机构
[1] Univ Zaragoza, Dept Design & Mfg Engn, Zaragoza 50018, Spain
关键词
Coordinate measurement arms; Thermal error model; Parameter identification; Data capture; Error correction; COMPENSATION; ACCURACY; CMM;
D O I
10.1016/j.precisioneng.2009.01.002
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
The influence of thermal changes in the accuracy of articulated arm coordinate measuring machines (AACMMs) is one of their main error sources. Usually, the effects of this problem are minimized by using low thermal-expansion materials in the arm design or by implementing an empirical error correction model based on the outputs of several temperature sensors placed inside the arm. These models, inherited from temperature correction models for robot arms, are based on the empirical arm positioning error characterization of several temperatures inside its measuring range and by posterior corrections implemented using an average error approximation. Considering this approximation as a correction valid for all the temperature range violates the AACMM calibration conditions, which are defined by a kinematic parameter identification procedure at 20 degrees C. This implies that, by applying these models, the AACMM will work outside of calibration conditions and will not meet the nominal accuracy values obtained from the identification procedure. In this work a new empirical correction model for thermal errors is presented. This model keeps the calibration conditions established in the parameters identification procedure unaltered. This fact makes it possible to apply other correction models for geometric and non-geometric errors obtained at the calibration temperature after having applied the temperature correction. The algorithm and objective functions used, which are based on a new approach including terms regarding measurement accuracy and repeatability and using the measurements of a ball-beam gauge artifact are shown. The empirical correction model and the arm accuracy and repeatability results before and after correction are also explained, showing an important accuracy improvement in the arm performance for typical arm operation at temperatures different from 20 degrees C. (C) 2009 Elsevier Inc. All rights reserved.
引用
收藏
页码:476 / 485
页数:10
相关论文
共 50 条
  • [21] Probe Parameters Calibration for Articulated Arm Coordinate Measuring Machine
    Cheng Wentao
    Fei Yetai
    Yu Liandong
    Yang Ruichang
    6TH INTERNATIONAL SYMPOSIUM ON PRECISION ENGINEERING MEASUREMENTS AND INSTRUMENTATION, 2010, 7544
  • [22] Parameter calibration method of articulated arm coordinate measuring machine
    Wang X.
    Wang H.
    Lu Y.
    Zhang P.
    2016, Chinese Society of Agricultural Machinery (47): : 408 - 412
  • [23] Articulated Arm Length Calibration for Cantilever Coordinate Measuring Machine
    Ran, Honglei
    Liu, Shugui
    Zhang, Haitao
    MEASUREMENT TECHNOLOGY AND ENGINEERING RESEARCHES IN INDUSTRY, PTS 1-3, 2013, 333-335 : 77 - 80
  • [24] Probe Error Analysis of Articulated Arm Coordinate Measuring Machine
    Cai Zeliang
    Luo Zai
    Liu Hui
    TENTH INTERNATIONAL SYMPOSIUM ON PRECISION ENGINEERING MEASUREMENTS AND INSTRUMENTATION, 2019, 11053
  • [25] Screw theory-based mathematical model and parameter calibration for articulated arm coordinate measuring machine
    Huang, Kui
    Mo, Jianhua
    Fu, Zemin
    Shi, Yusheng
    Huazhong Keji Daxue Xuebao (Ziran Kexue Ban)/Journal of Huazhong University of Science and Technology (Natural Science Edition), 2010, 38 (04): : 94 - 97
  • [26] Kinematic calibration for industrial robots using articulated arm coordinate machines
    Gao, Guanbin
    Zhang, Hongwei
    San, Hongjun
    Sun, Guoqing
    Wu, Xing
    Wang, Wen
    INTERNATIONAL JOURNAL OF MODELLING IDENTIFICATION AND CONTROL, 2019, 31 (01) : 16 - 26
  • [27] Evaluation of Techniques to Capture Nominal Data for Verification and Kinematic Calibration of Articulated Arm Coordinate Measuring Machines
    Santolaria, Jorge
    Brau, Agustin
    Javier Brosed, Francisco
    Jose Aguilar, Juan
    THIRD MANUFACTURING ENGINEERING SOCIETY INTERNATIONAL CONFERENCE: MESIC-09, 2009, 1181 : 202 - 212
  • [28] Kinematic modeling and simulation of articulated arm coordinate measuring machines based on MCPC method
    Lu, Jian
    Gao, Guanbin
    SIXTH INTERNATIONAL SYMPOSIUM ON PRECISION MECHANICAL MEASUREMENTS, 2013, 8916
  • [29] Modeling and Error Compensation of Robotic Articulated Arm Coordinate Measuring Machines Using BP Neural Network
    Gao, Guanbin
    Zhang, Hongwei
    San, Hongjun
    Wu, Xing
    Wang, Wen
    COMPLEXITY, 2017,
  • [30] Study on the structrual parameter coupling of articulated arm coordinate measuring machines
    Gao, G. (gbgao@163.com), 1600, Universitas Ahmad Dahlan (11):