Characterization of HPGe gamma spectrometers by geant4 Monte Carlo simulations

被引:16
|
作者
Szentmiklosi, Laszlo [1 ]
Belgya, Tamas [1 ]
Maroti, Boglarka [1 ]
Kis, Zoltan [1 ]
机构
[1] Hungarian Acad Sci, Energy Res Ctr, Nucl Anal & Radiog Dept, H-1525 Budapest, Hungary
关键词
HPGe detector; Monte Carlo simulations; Geant4; Detector response function; Efficiency; Peak-to-total ratio; PRECISE EFFICIENCY CALIBRATION; DETECTOR; ENERGY;
D O I
10.1007/s10967-014-2976-6
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Two coaxial and a low-energy HPGe detector were characterized with Monte Carlo simulations, using the geant4 toolkit. The geometry of the detectors, including the dimensions of the crystal and the internal structural parts, were initially taken from the factory specifications and from X-ray radiographies, and were later fine-tuned. The detector response functions, with special emphasis on the absolute full-energy peak efficiencies and peak-to-total ratios, were calculated and compared to experimental data taken at different measurement geometries. Between 150 keV and 11 MeV an agreement within 1-2 standard deviation has been achieved, whereas systematic deviations were experienced at lower energies.
引用
收藏
页码:553 / 558
页数:6
相关论文
共 50 条
  • [11] A framework for Monte Carlo simulation calculations in GEANT4
    Walker, B.
    Figgins, J.
    Comfort, J. R.
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2006, 568 (02): : 889 - 895
  • [12] Monte Carlo studies of the Tunisian gamma irradiation facility using GEANT4 code
    Kadri, O
    Gharbi, F
    Farah, K
    Mannai, K
    Trabelsi, A
    [J]. APPLIED RADIATION AND ISOTOPES, 2006, 64 (02) : 170 - 177
  • [13] Environmental radionuclides as contaminants of HPGe gamma-ray spectrometers: Monte Carlo simulations for Modane underground laboratory
    Breier, R.
    Brudanin, V. B.
    Loaiza, P.
    Piquemal, F.
    Povinec, P. P.
    Rukhadze, E.
    Rukhadze, N.
    Stekl, I.
    [J]. JOURNAL OF ENVIRONMENTAL RADIOACTIVITY, 2018, 190 : 134 - 140
  • [14] PRELIMINARY STUDY OF PROTON RADIOGRAPHY IMAGING QUALITIES USING GEANT4 MONTE CARLO SIMULATIONS
    Depauw, Nicolas
    Danto, Sylvain
    Bednarz, Bryan
    Paganetti, Harald
    Fink, Yoel
    Seco, Joao
    [J]. NUCLEAR TECHNOLOGY, 2011, 175 (01) : 6 - 10
  • [15] Comparison of Monte Carlo simulations based on Geant4 and Fluka to the TileCal test beam data
    Cascella, M.
    Gallas, M.
    Pokorski, W.
    Ribon, A.
    [J]. NUOVO CIMENTO C-COLLOQUIA AND COMMUNICATIONS IN PHYSICS, 2009, 32 (3-4): : 425 - 428
  • [16] Comparison of scattering experiments using synchrotron radiation with Monte Carlo simulations using Geant4
    Gerlach, M.
    Krumrey, M.
    Cibik, L.
    Mueller, P.
    Ulm, G.
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2009, 608 (02): : 339 - 343
  • [17] An Optimized Design of Single-Channel Beta-Gamma Coincidence Phoswich Detector by Geant4 Monte Carlo Simulations
    Zhang, Weihua
    Mekarski, Pawel
    Bean, Marc
    Yi, Jing
    Ungar, Kurt
    [J]. SCIENCE AND TECHNOLOGY OF NUCLEAR INSTALLATIONS, 2011, 2011
  • [18] Implementation of the reverse/adjoint Monte Carlo method into Geant4
    Desorgher, L.
    Lei, F.
    Santin, G.
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2010, 621 (1-3): : 247 - 257
  • [19] A Geant4 Monte Carlo method for synthesizing radioxenon beta-gamma coincidence spectra
    Zhang, Weihua
    Mekarski, Pawel
    Lam, Jonathan
    Ungar, Kurt
    [J]. JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY, 2010, 286 (01) : 235 - 240
  • [20] Monte Carlo improvement of dose uniformity in gamma irradiation processing using the GEANT4 code
    Kadri, O
    Gharbi, F
    Farah, K
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2005, 239 (04): : 391 - 398