Radiochromic film dosimetry of a high dose rate beta source for intravascular brachytherapy

被引:27
|
作者
Duggan, DM [1 ]
Coffey, CW
Lobdell, JL
Schell, MC
机构
[1] Vanderbilt Univ, Med Ctr, Ctr Radiat Oncol, Nashville, TN 37232 USA
[2] Novoste Corp, Norcross, GA 30093 USA
[3] Univ Rochester, Med Ctr, Dept Radiat Oncol, Rochester, NY 14642 USA
关键词
intravascular brachytherapy; Sr-90; radiochromic film; brachytherapy dosimetry; quality assurance;
D O I
10.1118/1.598814
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Good clinical physics practice requires that dose rates of brachytherapy sources be checked by the institution using them, as recommended by American Association of Physicists in Medicine Task Group 56 and The American College of Radiology. For intravascular brachytherapy with catheter-based systems, AAPM Task Group 60 recommends that the dose rate be measured at a reference point located at a radial distance of 2 mm from the center of the catheter axis. AAPM Task Group 60 also recommends that the dose rate along the catheter axis at a radial distance of 2 mm should be uniform to within +/- 10% in the center two-thirds of the treated length, and the relative dose rate in the plane perpendicular to the catheter axis through the center of the source should be measured at distances from 0.5 mm to R-90 (the distance from a point source within which 90% of the energy is deposited) at intervals of 0.5 mm. Radiochromic him dosimetry has been used to measure the dose distribution in a plane parallel to and at a radial distance of 2 mm from the axis of a novel, catheter-based, beta source for intravascular brachytherapy. The dose rate was averaged along a line parallel to the catheter axis at a radial distance of 2 mm, in the centered 24.5 mm of the treated length. This average dose rate agreed with the dose rate measured with a well ionization chamber by the replacement method using source trains calibrated with an extrapolation chamber at the National institute of Standards and Technology. Ail of the dose rates in the centered 24.5 mm of a line parallel to the axis at a distance of 2 mm were within +/- 10% of the average. (C) 1999 American Association of Physicists in Medicine. [S0094-2405(99)01711-3].
引用
收藏
页码:2461 / 2464
页数:4
相关论文
共 50 条
  • [31] Dosimetry of high dose rate (HDR) brachytherapy is highly reliable
    Ingram, G
    Kelly, D
    Flynn, J
    Brookover, W
    Kramer, B
    RADIOTHERAPY AND ONCOLOGY, 2004, 71 : S107 - S108
  • [32] Commissioning of applicator-guided stereotactic body radiation therapy boost with high-dose-rate brachytherapy for advanced cervical cancer using radiochromic film dosimetry
    Aldelaijan, Saad
    Wadi-Ramahi, Shada
    Nobah, Ahmad
    Moftah, Belal
    Devic, Slobodan
    Jastaniyah, Noha
    BRACHYTHERAPY, 2017, 16 (04) : 893 - 902
  • [33] A Novel Source Tracking Method for Accurate and Comprehensive HDR Brachytherapy Quality Assurance Using Radiochromic Film Dosimetry
    Aldelaijan, S.
    Papaconstadopoulos, P.
    Bekerat, H.
    Schneider, J.
    Seuntjens, J.
    Buzurovic, I.
    Devic, S.
    MEDICAL PHYSICS, 2018, 45 (06) : E572 - E572
  • [34] Relative dosimetry analysis of a high dose-rate 103Pd ocular brachytherapy source
    Millage, K
    Napolitano, M
    Hearn, R
    Rodgers, J
    MEDICAL PHYSICS, 2004, 31 (06) : 1914 - 1914
  • [35] Dosimetry Parameters of the IRH10 192Ir High Dose Rate Brachytherapy Source
    Yi, Chul-Young
    Chun, Kook Jin
    Hah, Suck-Ho
    Kim, Hyun-Moon
    Lim, Chunil
    JOURNAL OF RADIATION RESEARCH, 2010, 51 (04) : 485 - 492
  • [36] Erratum to: Monte Carlo dosimetry of the IRAsource high dose rate 192Ir brachytherapy source
    Akbar Sarabiasl
    Navid Ayoobian
    Hossein Poorbaygi
    Iraj Jabbari
    Mohammad Reza Javanshir
    Australasian Physical & Engineering Sciences in Medicine, 2016, 39 : 591 - 591
  • [37] Linearization of Dose Response Curve for the Radiochromic Film Dosimetry System
    Lewis, D.
    Devic, S.
    Tomic, N.
    Aldelaijan, S.
    DeBlois, F.
    Chan, M.
    Seuntjens, J.
    MEDICAL PHYSICS, 2012, 39 (06) : 4009 - 4009
  • [38] Monte Carlo and TLD dosimetry of an 192Ir high dose-rate brachytherapy source
    Karaiskos, P
    Angelopoulos, A
    Sakelliou, L
    Sandilos, P
    Antypas, C
    Vlachos, L
    Koutsouveli, E
    MEDICAL PHYSICS, 1998, 25 (10) : 1975 - 1984
  • [39] Comparative dosimetry of GammaMed Plus high-dose rate Ir-192 brachytherapy source
    Patel, N. P.
    Majumdar, B.
    Vijayan, V.
    JOURNAL OF MEDICAL PHYSICS, 2010, 35 (03) : 137 - 143
  • [40] Monte Carlo dosimetry of a tandem positioned beta-emitting intravascular brachytherapy source train
    Wallace, SA
    Schumer, W
    Horrigan, M
    MEDICAL PHYSICS, 2002, 29 (04) : 544 - 549