Comparative dosimetry of GammaMed Plus high-dose rate Ir-192 brachytherapy source

被引:6
|
作者
Patel, N. P. [1 ,2 ]
Majumdar, B. [1 ]
Vijayan, V. [3 ]
机构
[1] Govt Coll Sci, Dept Phys, Raipur, Madhya Pradesh, India
[2] AH Reg Canc Ctr, Dept Med Phys, Cuttack, Orissa, India
[3] Inst Phys, Bhubaneswar, Orissa, India
关键词
0.1-cc chamber; AAPM TG-43; dosimetry; EGSnrc simulation code; high dose rate;
D O I
10.4103/0971-6203.66761
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The comparative dosimetry of GammaMed (GM) Plus high-dose rate brachytherapy source was performed by an experiment using 0.1-cc thimble ionization chamber and simulation-based study using EGSnrc code. In-water dose measurements were performed with 0.1-cc chamber to derive the radial dose function (r = 0.8 to 20.0 cm) and anisotropy function (r = 5.0 cm with polar angle from 10 degrees to 170 degrees). The nonuniformity correction factor for 0.1-cc chamber was applied for in-water measurements at shorter distances from the source. The EGSnrc code was used to derive the dose rate constant (L), radial dose function g L (r) and anisotropy function F(r, q) of GM Plus source. The dosimetric data derived using EGSnrc code in our study were in very good agreement relative to published data for GM Plus source. The radial dose function up to 12 cm derived from measured dose using 0.1-cc chamber was in agreement within 3% of data derived by the simulation study.
引用
收藏
页码:137 / 143
页数:7
相关论文
共 50 条
  • [1] Monte Carlo characterization of the GammaMed HDR Plus Ir-192 brachytherapy source
    Reyes, E.
    Sosa, M.
    Gil-Villegas, A.
    Monzon, E.
    [J]. BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2016, 2 (01):
  • [2] Dosimetry of indigenously developed Ir-192 high-dose rate brachytherapy source: An EGSnrc Monte Carlo study
    Sahoo, Sridhar
    Selvam, T. Palani
    Sharma, S. D.
    Das, Trupti
    Dey, A. C.
    Patil, B. N.
    Sastri, K. V. S.
    [J]. JOURNAL OF MEDICAL PHYSICS, 2016, 41 (02) : 115 - 122
  • [3] Transit dose of an Ir-192 high dose rate brachytherapy stepping source
    Wong, TPY
    Fernando, W
    Johnston, PN
    Bubb, IF
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2001, 46 (02): : 323 - 331
  • [4] TLD, diode and monte carlo dosimetry of an Ir-192 source for high dose-rate brachytherapy
    Kirov, AS
    Williamson, JF
    Meigooni, AS
    Zhu, Y
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1995, 40 (12): : 2015 - 2036
  • [5] Monte Carlo study of TG-43 dosimetry parameters of GammaMed Plus high dose rate 192Ir brachytherapy source using TOPAS
    Wu, Jianan
    Xie, Yaoqin
    Ding, Zhen
    Li, Feipeng
    Wang, Luhua
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2021, 22 (06): : 146 - 153
  • [6] HIGH-DOSE RATE (HDR) IR-192 BRACHYTHERAPY APPLICATION OF BASAL CELL CARCINOMA
    Basaric, Borko
    Petrovic, Borislava
    Teodorovic, Milan
    Baucal, Milutin
    Rutonjski, Laza
    Cudic, Ozren
    Djuran, Branislav
    Mitric-Askovic, Milana
    [J]. RAD 2015: THE THIRD INTERNATIONAL CONFERENCE ON RADIATION AND APPLICATIONS IN VARIOUS FIELDS OF RESEARCH, 2015, : 65 - 68
  • [7] EVALUATION OF THE BUILDUP EFFECT OF AN IR-192 HIGH DOSE-RATE BRACHYTHERAPY SOURCE
    PARK, HC
    ALMOND, PR
    [J]. MEDICAL PHYSICS, 1992, 19 (05) : 1293 - 1297
  • [8] Impact of different Ir-192 source models on dose calculations in high-dose-rate brachytherapy
    Otani, Yuki
    Shimamoto, Shigetoshi
    Sumida, Iori
    Takahashi, Yutaka
    Tani, Shoji
    Oshima, Tetsuya
    Onosaka, Satoshi
    Isohashi, Fumiaki
    Tamari, Keisuke
    Ogawa, Kazuhiko
    [J]. PHYSICS & IMAGING IN RADIATION ONCOLOGY, 2018, 7 : 23 - 26
  • [9] Calibration of Ir-192 high dose rate brachytherapy source using different calibration procedures
    Bondel, Shwetha
    Ravikumar, Manickham
    Supe, Sanjay Sudhakar
    Reddy, Buchuppudi Rekha
    [J]. REPORTS OF PRACTICAL ONCOLOGY AND RADIOTHERAPY, 2014, 19 (03) : 151 - 156
  • [10] In-Phantom Calibration of High Dose Rate Ir-192 Brachytherapy Source - A Feasibility Study
    Mandiram, Kuppuswamy
    Senthil, G.
    Sundaram, Visalatchi
    Muthulingam, S.
    Karthik, S.
    [J]. MEDICAL PHYSICS, 2009, 36 (06)