Development of a preclinical CCD- based temperature modulated fluorescence tomography platform

被引:4
|
作者
Nouizi, Farouk [1 ,2 ]
Erkol, Hakan [3 ]
Nikkhah, Deniz [1 ]
Kwong, Tiffany C. [1 ]
Gulsen, Gultekin [1 ,2 ]
机构
[1] Univ Calif Irvine, Dept Radiol Sci, Tu & Yuen Ctr Funct Oncoimaging, Irvine, CA 92697 USA
[2] Univ Calif Irvine, Chao Family Comprehens Canc Ctr, Irvine, CA 92697 USA
[3] Bogazici Univ, Dept Phys, TR-34342 Istanbul, Turkey
基金
美国国家卫生研究院;
关键词
DIFFUSE OPTICAL TOMOGRAPHY; QUANTITATIVE ACCURACY; RESOLUTION; INFORMATION; ALGORITHM;
D O I
10.1364/BOE.470723
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
In preclinical research, fluorescence molecular tomography (FMT) is the most sensitive imaging modality to interrogate whole-body and provide 3D distribution of fluorescent contract agents. Despite its superior sensitivity, its mediocre spatial-resolution has been the main barrier to its clinical translation. This limitation is mainly due to the high scattering of optical photons in biological tissue together with the limited boundary measurements that lead to an undetermined and ill-posed inverse problem. To overcome the limitations of FMT, we previously introduced a novel method termed, Temperature Modulated Fluorescence Tomography (TMFT). TMFT utilizes thermos-sensitive fluorescent agents (ThermoDots) as a key component and localizes them with high-intensity focused ultrasound (HIFU). Scanning the focused HIFU beam having a diameter phi= 1.3 mm across the tissue while monitoring the variation in the measured fluorescence signals reveals the position of the ThermoDots with high spatial accuracy. We have formerly built a prototype TMFT system that uses optical fibers for detection. In this paper, we present an upgraded version using a CCD camera-based detection that enables non-contact imaging. In this version, the animal under investigation is placed on an ultrasound transparent membrane, which eliminates the need for its immersion in optical matching fluids that were required by the fiber-based system. This CCD-based system will pave the way for convenient and wide-spread use of TMFT in preclinical research. Its performance validation on phantom studies demonstrates that high spatial-resolution (similar to 1.3 mm) and quantitative accuracy in recovered fluorophore concentration (<3% error) can be achieved.
引用
收藏
页码:5740 / 5752
页数:13
相关论文
共 50 条
  • [1] CCD-Based Temperature Modulated Fluorescence Tomography
    Mehrabi, Mehrnaz
    Nouizi, Farouk
    Algarawi, Maha
    Kwong, Tiffany C.
    Erkol, Hakan
    Sampathkumaran, Uma
    Gulsen, Gultekin
    [J]. OPTICAL TOMOGRAPHY AND SPECTROSCOPY OF TISSUE XIII, 2019, 10874
  • [2] Fast ICCD-based temperature modulated fluorescence tomography
    Nouizi, Farouk
    Kwong, Tiffany C.
    Turong, Bryan
    Nikkhah, Deniz
    Sampathkumaran, Uma
    Gulsen, Gultekin
    [J]. APPLIED OPTICS, 2023, 62 (28) : 7420 - 7430
  • [3] Temperature-modulated fluorescence tomography based on both concentration and lifetime contrast
    Lin, Yuting
    Kwong, Tiffany C.
    Bolisay, Linden
    Gulsen, Gultekin
    [J]. JOURNAL OF BIOMEDICAL OPTICS, 2012, 17 (05)
  • [4] Development of a theranostic preclinical fluorescence molecular tomography/cone beam CT-guided irradiator platform
    Nouizi, Farouk
    Brooks, Jamison
    Zuro, Darren M.
    Hui, Susanta K.
    Gulsen, Gultekin
    [J]. BIOMEDICAL OPTICS EXPRESS, 2022, 13 (11) : 6100 - 6112
  • [5] Temperature-modulated fluorescence tomography in a turbid media
    Lin, Yuting
    Bolisay, Linden
    Ghijsen, Michael
    Kwong, Tiffany C.
    Gulsen, Gultekin
    [J]. APPLIED PHYSICS LETTERS, 2012, 100 (07)
  • [6] Validation of Temperature-Modulated Fluorescence Tomography in vivo
    Kwong, Tiffany C.
    Nouizi, Farouk
    Lin, Yuting
    Rajyaguru, Rushi
    Trinh Nguyen
    Alptekin, Lara
    Sampathkumaran, Uma
    Zhu, Yue
    Ahmed, Shaaz
    Gulsen, Gultekin
    [J]. MULTIMODAL BIOMEDICAL IMAGING IX, 2014, 8937
  • [7] Temperature-Modulated Fluorescence Tomography: Modulating Tissue Temperature Using HIFU for High-Resolution in vivo Fluorescence Tomography
    Kwong, Tiffany C.
    Nouizi, Farouk
    Lin, Yuting
    Sampathkumaran, Uma
    Ahmed, Shaaz
    Gulsen, Gultekin
    [J]. MULTIMODAL BIOMEDICAL IMAGING VIII, 2013, 8574
  • [8] Simulation-based evaluation of the resolution and quantitative accuracy of temperature-modulated fluorescence tomography
    Lin, Yuting
    Nouizi, Farouk
    Kwong, Tiffany C.
    Gulsen, Gultekin
    [J]. APPLIED OPTICS, 2015, 54 (25) : 7612 - 7621
  • [9] Experimental evaluation of the resolution and quantitative accuracy of temperature-modulated fluorescence tomography
    Kwong, Tiffany C.
    Nouizi, Farouk
    Lin, Yuting
    Cho, Jaedu
    Zhu, Yue
    Sampathkumaran, Uma
    Gulsen, Gultekin
    [J]. APPLIED OPTICS, 2017, 56 (03) : 521 - 529
  • [10] Development of a.CCD-based optical fluorescence imaging system
    Li, YH
    Tang, D
    Chen, JY
    Wong, J
    Liu, H
    [J]. COMPLEX DYNAMICS, FLUCTUATIONS, CHAOS, AND FRACTALS IN BIOMEDICAL PHOTONICS, 2004, 5 (19): : 1 - 4