CCRI supplementary comparison of standards for absorbed dose to water in 60Co gamma radiation at radiation processing dose levels

被引:4
|
作者
Burns, D. T.
Allisy-Roberts, P. J.
Desrosiers, M. F.
Nagy, V. Yu.
Sharpe, P. H. G.
Laitano, R. F.
Mehta, K.
Schneider, M. K. H.
Zhang, Y. L.
机构
[1] Bur Int Poids & Mesures, F-92312 Sevres, France
[2] Natl Inst Stand & Technol, Gaithersburg, MD 20899 USA
[3] Natl Phys Lab, Teddington TW11 0LW, Middx, England
[4] Ist Nazl Metrol Radiaz Ionizzanti, Rome, Italy
[5] IAEA, A-1400 Vienna, Austria
[6] Phys Tech Bundesanstalt, D-3300 Braunschweig, Germany
[7] Natl Inst Metrol, Beijing, Peoples R China
关键词
D O I
10.1016/j.radphyschem.2006.02.003
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
Six national standards for absorbed dose to water in Co-60 gamma. radiation at the dose levels used in radiation processing have been compared over the range from 5 to 30 kGy using the alanine dosimeters of the NIST and the NPL as the transfer dosimeters. The standards are in agreement at the level of around 0.5%, which is significantly smaller than the stated standard uncertainties. (c) 2006 Elsevier Ltd. All rights reserved.
引用
收藏
页码:1087 / 1092
页数:6
相关论文
共 50 条
  • [1] Supplementary comparison CCRI(I)-S3 of standards for absorbed dose to water in 60Co gamma radiation at radiation processing dose levels
    McEwen, M.R.
    Sharpe, P.H.G.
    Pazos, I.M.
    Miller, A.
    Pawlak, E.
    Ninlaphruk, S.
    Zhang, Y.
    Kessler, C.
    [J]. Metrologia, 2022, 59 (1 A)
  • [2] Supplementary comparison CCRI(I)-S2 of standards for absorbed dose to water in 60Co gamma radiation at radiation processing dose levels
    Burns, D. T.
    Allisy-Roberts, P. J.
    Desrosiers, M. F.
    Sharpe, P. H. G.
    Pimpinella, M.
    Lourenco, V.
    Zhang, Y. L.
    Miller, A.
    Generalova, V.
    Sochor, V.
    [J]. METROLOGIA, 2011, 48
  • [3] Supplementary comparison APMP.RI(I)-S1 of standards for absorbed dose to water in 60Co gamma radiation at radiation processing dose levels
    Mungpayaban, H.
    Ninlaphruk, S.
    Chen, C. L.
    Yuan, M. C.
    Zhang, Y. L.
    Yuandi, Y.
    Hassan, G. M.
    Rabie, N.
    Kadni, T. B.
    Butler, D.
    [J]. METROLOGIA, 2021, 58 (1A)
  • [4] Comparison of the standards for absorbed dose to water of the NRC and the BIPM for 60Co gamma radiation
    Kessler, C.
    Burns, D. T.
    Allisy-Roberts, P. J.
    McCaffrey, J. P.
    McEwen, M. R.
    Ross, C. K.
    [J]. METROLOGIA, 2010, 47
  • [5] Comparison of the standards for absorbed dose to water of the ARPANSA and the BIPM for 60Co gamma radiation
    Kessler, C.
    Burns, D. T.
    Allisy-Roberts, P. J.
    Butler, D.
    Lye, J.
    Webb, D.
    [J]. METROLOGIA, 2012, 49
  • [6] Comparison of the standards for absorbed dose to water of the BEV, Austria, and the BIPM for 60Co gamma radiation
    Kessler, C.
    Allisy-Roberts, P. J.
    Steurer, A.
    Tiefenboeck, W.
    Gabris, F.
    [J]. METROLOGIA, 2010, 47
  • [7] The PTB water calorimeter for the absolute determination of absorbed dose to water in 60Co radiation
    Krauss, A
    [J]. METROLOGIA, 2006, 43 (03) : 259 - 272
  • [8] Comparison of the standards for absorbed dose to water of the VNIIFTRI, Russia and the BIPM in 60Co γ rays
    Allisy-Roberts, P. J.
    Kessler, C.
    Burns, D. T.
    Berlyand, V.
    Berlyand, A.
    [J]. METROLOGIA, 2010, 47
  • [9] Comparison of the standards for absorbed dose to water of the NMIJ and the BIPM for 60Co γ-ray beams
    Kessler, C.
    Allisy-Roberts, P. J.
    Morishita, Y.
    Kato, M.
    Takata, N.
    Kurosawa, T.
    Tanaka, T.
    Saito, N.
    [J]. METROLOGIA, 2011, 48
  • [10] Comparison of the standards for absorbed dose to water of the VSL, The Netherlands, and the BIPM for 60Co γ rays
    Kessler, C.
    Burns, D.
    Jansen, B. J.
    de Pooter, J. A.
    de Prez, L. A.
    [J]. METROLOGIA, 2018, 55