Validation of a Second-Generation Near-Infrared Spectroscopy Monitor in Children With Congenital Heart Disease

被引:17
|
作者
Nasr, Viviane G. [1 ,2 ]
Bergersen, Lisa T. [3 ,4 ]
Lin, Hung-Mo [5 ]
Benni, Paul B. [6 ]
Bernier, Rachel S. [1 ]
Anderson, Michelle E. [1 ]
Kussman, Barry D. [1 ,2 ]
机构
[1] Boston Childrens Hosp, Dept Anesthesiol Perioperat & Pain Med, 300 Longwood Ave, Boston, MA 02115 USA
[2] Harvard Med Sch, Dept Anaesthesia, Boston, MA 02115 USA
[3] Boston Childrens Hosp, Dept Cardiol, Boston, MA USA
[4] Harvard Med Sch, Dept Pediat, Boston, MA 02115 USA
[5] Icahn Sch Med, Dept Hlth Populat Sci & Policy, Mt Sinai, NY USA
[6] CASMED, Branford, CT USA
来源
ANESTHESIA AND ANALGESIA | 2019年 / 128卷 / 04期
关键词
CEREBRAL OXYGEN-SATURATION; CONCORDANCE CORRELATION-COEFFICIENT; NEURODEVELOPMENTAL OUTCOMES; VENOUS CONTRIBUTIONS; TISSUE OXYGENATION; HEALTHY-VOLUNTEERS; CARDIAC-SURGERY; ECMO PATIENTS; OXIMETER; ARTERIAL;
D O I
10.1213/ANE.0000000000002796
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
BACKGROUND: Cerebral oximetry using near-infrared spectroscopy is a noninvasive optical technology to detect cerebral hypoxia-ischemia and develop interventions to prevent and ameliorate hypoxic brain injury. Cerebral oximeters are calibrated and validated by comparison of the near-infrared spectroscopy-measured cerebral O-2 saturation (SctO(2)) to a "field" or reference O-2 saturation (REF CX) calculated as a weighted average from arterial and jugular bulb oxygen saturations. In this study, we calibrated and validated the second-generation, 5 wavelength, FORE-SIGHT Elite with the medium sensor (source-detector separation 12 and 40 mm) for measurement of SctO(2) in children with congenital heart disease. METHODS: After institutional review board approval and written informed consent, 63 children older than 1 month and >= 2.5 kg scheduled for cardiac catheterization were enrolled. Self-adhesive FORE-SIGHT Elite medium sensors were placed on the right and left sides of the forehead. Blood samples for calculation of REF CX were drawn simultaneously from the aorta or femoral artery and the jugular bulb before (T1) and shortly after (T2) baseline hemodynamic measurements. FORE-SIGHT Elite SctO(2) measurements were compared to the REF CX (REF CX = [0.3 SaO(2)] + [0.7 SjbO(2)]) using Deming regression, least squares linear regression, and Bland-Altman analysis. RESULTS: Sixty-one subjects (4.5 [standard deviation 4.4] years of age; 17 [standard deviation 13] kg, male 56%) completed the study protocol. Arterial oxygen saturation ranged from 64.7% to 99.1% (median 96.0%), jugular bulb venous oxygen saturation from 34.1% to 88.1% (median 68.2%), the REF CX from 43.8% to 91.4% (median 76.9%), and the SctO(2) from 47.8% to 90.8% (median 76.3%). There was a high degree of correlation in SctO(2) between the right and left sensors at a given time point (within subject between sensor correlation r = 0.91 and 95% confidence interval [CI], 0.85-0.94) or between T1 and T2 for the right and left sensors (replicates, within subject between time point correlation r = 0.95 and 95% CI, 0.92-0.96). By Deming regression, the estimated slope was 0.966 (95% CI, 0.786-1.147; P = .706 for testing against null hypothesis of slope = 1) with a y intercept of 2.776 (95% CI, -11.102 to 16.654; P = .689). The concordance correlation coefficient was 0.873 (95% CI, 0.798-0.922). Bland-Altman analysis for agreement between SctO(2) and REF CX that accounted for repeated measures (both in times and sensors) found a bias of -0.30% (95% limits of agreement: -10.56% to 9.95%). CONCLUSIONS: This study calibrated and validated the FORE-SIGHT Elite tissue oximeter to accurately measure SctO(2) in pediatric patients with the medium sensor.
引用
收藏
页码:661 / 668
页数:8
相关论文
共 50 条
  • [21] Near-infrared spectroscopy as a real-time monitor
    Widman, Ronald A.
    [J]. SURGICAL NEUROLOGY, 2007, 68 (03): : 296 - 296
  • [22] Cerebral near-infrared spectroscopy in children undergoing heart surgery - Reply
    duPlessis, A
    Newburger, J
    Jonas, R
    Hickey, P
    Volpe, J
    [J]. ANNALS OF NEUROLOGY, 1996, 40 (05) : 818 - 819
  • [23] A narrative review on the use of near-infrared spectroscopy to monitor bladder volume and in vitro validation approaches
    Tu, Kelsey J.
    Lam, Jesse H.
    Kim, Sehwan
    [J]. TRANSLATIONAL ANDROLOGY AND UROLOGY, 2023, 12 (11) : 1732 - 1739
  • [24] Intraoperative near-infrared imaging of parathyroid glands: A comparison of first- and second-generation technologies
    Akbulut, Serkan
    Erten, Ozgun
    Gokceimam, Mehmet
    Kim, Yoo S.
    Krishnamurthy, Vikram
    Heiden, Katherine
    Jin, Judy
    Siperstein, Allan
    Berber, Eren
    [J]. JOURNAL OF SURGICAL ONCOLOGY, 2021, 123 (04) : 866 - 871
  • [25] Towards the next generation of near-infrared spectroscopy
    Hoshi, Yoko
    [J]. PHILOSOPHICAL TRANSACTIONS OF THE ROYAL SOCIETY A-MATHEMATICAL PHYSICAL AND ENGINEERING SCIENCES, 2011, 369 (1955): : 4425 - 4439
  • [26] Ductal closure and near-infrared spectroscopy for regional oxygenation monitoring in ductus-dependent congenital heart disease
    Seguela, P-E
    Guillet, E.
    Thambo, J-B.
    Mauriat, P.
    [J]. ARCHIVES DE PEDIATRIE, 2015, 22 (08): : 857 - 860
  • [27] Four-side near-infrared spectroscopy measured in a paediatric population during surgery for congenital heart disease
    Amigoni, Angela
    Mozzo, Elena
    Brugnaro, Luca
    Tiberio, Ivo
    Pittarello, Demetrio
    Stellin, Giovanni
    Bonato, Raffaele
    [J]. INTERACTIVE CARDIOVASCULAR AND THORACIC SURGERY, 2011, 12 (05) : 707 - 712
  • [28] Correlation of abdominal site near-infrared spectroscopy with gastric tonometry in infants following surgery for congenital heart disease
    Kaufman, Jon
    Almodovar, Melvin C.
    Zuk, Jeannie
    Friesen, Robert H.
    [J]. PEDIATRIC CRITICAL CARE MEDICINE, 2008, 9 (01) : 62 - 68
  • [29] Perioperative Cerebral Near-Infrared Spectroscopy Variability and Neurodevelopmental Outcomes in Congenital Heart Patients
    Ghimire, Laxmi V.
    [J]. PEDIATRIC CRITICAL CARE MEDICINE, 2017, 18 (05) : 499 - 500
  • [30] Near-infrared spectroscopy as a green technology to monitor coffee roasting
    Wojcicki, Krzysztof
    [J]. FOODS AND RAW MATERIALS, 2022, 10 (02) : 295 - 303