A narrative review on the use of near-infrared spectroscopy to monitor bladder volume and in vitro validation approaches

被引:0
|
作者
Tu, Kelsey J. [1 ]
Lam, Jesse H. [2 ]
Kim, Sehwan [1 ,3 ,4 ]
机构
[1] Dankook Univ, Dept Biomed Engn, Cheonan, Chungnam, South Korea
[2] Dankook Univ, Beckman Laser Inst Korea, Sch Med, Cheonan, Chungnam, South Korea
[3] Medithings Co Ltd, Seoul, South Korea
[4] Dankook Univ, Dept Biomed Engn, 119,Dandae ro, Cheonan 31116, Chungnam, South Korea
基金
新加坡国家研究基金会;
关键词
Non-invasive bladder monitoring; bladder phantom; neurogenic bladder (NB); PHANTOM; BRAIN;
D O I
10.21037/tau-23-297
中图分类号
R69 [泌尿科学(泌尿生殖系疾病)];
学科分类号
摘要
Background and Objective: Current guidelines for patients with neurogenic bladder (NB) dysfunction suggest the use of self-intermittent catheterization with adherence to catheterization timers. Due to biorhythmic variability, unpredictable voiding times may occur. As a result, many patients abstain from extended social or work activities, turn to more secluded lifestyles, and generally experience a decrease in quality of life. Being able to know when to void is essential for patients with NB dysfunction. To solve the problem of variable void timings, wearable devices using near-infrared spectroscopy (NIRS) have been emergent and showed potential for bladder volume monitoring. Therefore, in this review, we provided a comprehensive overview of research which implemented NIRS for bladder volume monitoring, and discussed how the researchers validated their device by in-vitro methods and suggested a potential validation approach. Methods: A literature search using PubMed and Google Scholar was conducted on February 2023. Publications involving bladder volume monitoring incorporating NIRS technique with in-vitro validation was considered for review. Key Content and Findings: Due to the novelty of NIRS being applied to bladder monitoring, there are a few possibilities to effectively validate this technique through in-vitro methods. Ballistics gel, which has been proven to be a versatile material for applications involving ultrasound, could be a suitable material when constructing a bladder phantom for in-vitro validation of NIRS technology. Conclusions: By outlining a more standardized in-vitro validation model based on ballistics gel, we hope to facilitate development in this field towards a more accurate and robust NIRS-based bladder monitoring device.
引用
收藏
页码:1732 / 1739
页数:8
相关论文
共 50 条
  • [1] Near-Infrared Spectroscopy to Monitor Nutritional Status of Neonates: A Review
    Miller, Alexander P.
    Mustafa, Fatin Hamimi
    Jones, Peter W.
    Jeffery, Heather E.
    Carberry, Angela E.
    McEwan, Alistair L.
    [J]. IEEE REVIEWS IN BIOMEDICAL ENGINEERING, 2020, 13 : 280 - 291
  • [2] In vivo validation of cerebral near-infrared spectroscopy: a review
    la Cour, Amalie
    Greisen, Gorm
    Hyttel-Sorensen, Simon
    [J]. NEUROPHOTONICS, 2018, 5 (04)
  • [3] VALIDATION OF NEAR-INFRARED SPECTROSCOPY IN MAN
    MANCINI, DM
    FARRELL, L
    WILSON, JR
    [J]. CIRCULATION, 1992, 86 (04) : 401 - 401
  • [4] VALIDATION OF NEAR-INFRARED SPECTROSCOPY IN HUMANS
    MANCINI, DM
    BOLINGER, L
    LI, H
    KENDRICK, K
    CHANCE, B
    WILSON, JR
    [J]. JOURNAL OF APPLIED PHYSIOLOGY, 1994, 77 (06) : 2740 - 2747
  • [5] Advancing Patient Care: Innovative Use of Near-Infrared Spectroscopy for Monitoring Urine Volume in Neurogenic Bladder
    Kang, Byeong-Il
    Kim, Aram
    Kim, Sehwan
    [J]. INTERNATIONAL NEUROUROLOGY JOURNAL, 2023, 27 : S27 - S33
  • [6] Use of multivariate curve resolution to monitor an esterification reaction by near-infrared spectroscopy
    Blanco, M
    Castillo, M
    Beneyto, R
    Porcel, M
    [J]. SPECTROSCOPY LETTERS, 2005, 38 (06) : 825 - 837
  • [7] Use of near-infrared spectroscopy as a physiologic monitor for intra-abdorninal hypertension
    Widder, Sandy
    Ranson, Michael K.
    Zygun, David
    Knox, Linda
    Laupland, Kevin B.
    Laird, Philip
    Ball, Chad G.
    Kirkpatrick, Andrew W.
    [J]. JOURNAL OF TRAUMA-INJURY INFECTION AND CRITICAL CARE, 2008, 64 (05): : 1165 - 1168
  • [8] Near-infrared spectroscopy as a hemodynamic monitor in critical illness
    Ghanayem, Nancy S.
    Wernovsky, Gil
    Hoffman, George M.
    [J]. PEDIATRIC CRITICAL CARE MEDICINE, 2011, 12 (04) : S27 - S32
  • [9] Near-infrared spectroscopy as a real-time monitor
    Widman, Ronald A.
    [J]. SURGICAL NEUROLOGY, 2007, 68 (03): : 296 - 296
  • [10] Diagnosis of Bladder Outlet Obstruction in Men Using a Near-infrared Spectroscopy Instrument as the Noninvasive Monitor for Bladder Function
    Zhang, Peng
    Yang, Yong
    Wu, Zhi-jin
    Zhang, Chao-hua
    Zhang, Xiao-dong
    [J]. UROLOGY, 2013, 82 (05) : 1098 - 1102