Evaluating the impact of scan settings on automatic tube current modulation in CT using a novel phantom

被引:27
|
作者
Merzan, Deborah [1 ]
Nowik, Patrik [1 ]
Poludniowski, Gavin [1 ]
Bujila, Robert [1 ,2 ]
机构
[1] Karolinska Univ Hosp, Med Radiat Phys & Nucl Med, SE-17176 Stockholm, Sweden
[2] Royal Inst Technol, Dept Phys, SE-10691 Stockholm, Sweden
来源
BRITISH JOURNAL OF RADIOLOGY | 2017年 / 90卷 / 1069期
关键词
EXPOSURE CONTROL; DOSE REDUCTION; IMAGE QUALITY; SYSTEM; NOISE;
D O I
10.1259/bjr.20160308
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective: The aim of this study was to make a comprehensive evaluation of how variable scan settings can affect the performance of automatic tube current modulation (ATCM) in recent CT scanners from the four major manufacturers. Methods: A phantom was designed and manufactured for the purpose of evaluating ATCM. The phantom was scanned with four categories of systematically varied settings (scan projection radiograph, technique and reconstruction parameters and phantom miscentring). The performance of ATCM, in terms of applied tube current and noise uniformity, for the scans with varied settings was compared with a reference scan using subjective and quantitative approaches. Results: The ATCM implemented by each manufacturer is based on different principles and any affect to the performance of the ATCM, when varying scan settings, will manifest differently among the vendors. The results are summarized in four tables corresponding to the categories of varied settings. Conclusion: The developed phantom proved useful for evaluating the ATCM. It is important to understand how different implementations (vendor specific) of ATCM perform in order to make informed decisions about the selection of scan settings when designing protocols. The resulting tables can serve as a reference for understanding the different implementations of ATCM and highlight settings that should be taken into consideration when adjusting an imaging protocol. Advances in knowledge: The results from this work can serve as a reference for how changes in geometry or scan settings can affect the performance of ATCM, in terms of tube current and noise.
引用
收藏
页数:14
相关论文
共 50 条
  • [31] Comparison of angular and combined automatic tube current modulation techniques with constant tube current CT of the abdomen and pelvis
    Rizzo, S
    Kalra, M
    Schmidt, B
    Dalal, T
    Suess, C
    Flohr, T
    Blake, M
    Saini, S
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2006, 186 (03) : 673 - 679
  • [32] New technique to avoid an increase in CT radiation dose from metallic prostheses with automatic tube current modulation: Phantom and patient study
    Dalal, T
    Rizzo, SMR
    Kalra, MK
    Schmidt, B
    Suess, C
    Flohr, T
    Blake, MA
    Saini, S
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2005, 184 (04) : 51 - 51
  • [33] Pediatric cardiovascular CT angiography: Radiation dose reduction using automatic anatomic tube current modulation
    Herzog, Christopher
    Mulvihill, Denise M.
    Nguyen, Shaun A.
    Savino, Giancarlo
    Schmidt, Bernhard
    Costello, Philip
    Vogl, Thomas J.
    Schoepf, U. Joseph
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2008, 190 (05) : 1232 - 1240
  • [34] Monte Carlo Simulation and Dose Calculation of Chest CT Scan with Tube Current Modulation
    Gao, Y.
    Ding, A.
    Zhang, D.
    Liu, B.
    Caracappa, P.
    MEDICAL PHYSICS, 2013, 40 (06)
  • [35] Dose assessment in petrous bone CT: Effect of tube current modulation and scan length
    Saga, Zouhir
    Rahmouni, Abdelaali
    Talbi, Mohammed
    Anam, Choirul
    Rezzouk, Abdellah
    RADIATION PHYSICS AND CHEMISTRY, 2024, 225
  • [36] Dose conversion coefficients for CT examinations of adults with automatic tube current modulation
    Schlattl, H.
    Zankl, M.
    Becker, J.
    Hoeschen, C.
    PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (20): : 6243 - 6261
  • [37] Dose conversion coefficients for paediatric CT examinations with automatic tube current modulation
    Schlattl, H.
    Zankl, M.
    Becker, J.
    Hoeschen, C.
    PHYSICS IN MEDICINE AND BIOLOGY, 2012, 57 (20): : 6309 - 6326
  • [38] Automatic tube-current modulation in CT -: A comparison between different solutions
    Althén, JN
    RADIATION PROTECTION DOSIMETRY, 2005, 114 (1-3) : 308 - 312
  • [39] Automatic tube current modulation technique for multidetector CT: is it effective with a 64-detector CT?
    Funama Y.
    Awai K.
    Hatemura M.
    Shimamura M.
    Yanaga Y.
    Oda S.
    Yamashita Y.
    Radiological Physics and Technology, 2008, 1 (1) : 33 - 37
  • [40] Phantom study evaluating effect of lowering CT tube current for attenuation correction in quantitative SPECT
    Lawson, Susan
    Avram, Anca
    Dewaraja, Yuni
    Ackermann, Robert
    JOURNAL OF NUCLEAR MEDICINE, 2016, 57