Setup accuracy and dose attenuation of a wooden immobilization system for lung stereotactic body radiotherapy

被引:1
|
作者
Nitta, Yuya [1 ]
Ueda, Yoshihiro [1 ]
Murata, Seiya [1 ]
Isono, Masaru [1 ]
Ohira, Shingo [1 ]
Masaoka, Akira [1 ]
Inui, Shoki [1 ,2 ]
Washio, Hayate [3 ,4 ]
Yoshinaka, Reimi [1 ]
Sagawa, Tomohiro [1 ]
Miyazaki, Masayoshi [1 ]
Teshima, Teruki [5 ]
机构
[1] Osaka Int Canc Inst, Dept Radiat Oncol, Osaka, Japan
[2] Osaka Univ, Grad Sch, Dept Med Phys & Engn, Osaka, Japan
[3] Osaka Int Canc Inst, Dept Radiol, Osaka, Japan
[4] Kumamoto Univ, Grad Sch Hlth Sci, Kumamoto, Japan
[5] Osaka Heavy Ion Therapy Ctr, Osaka, Japan
关键词
lung SBRT; wooden baseplate; setup verification; IMAGE-GUIDED RADIOTHERAPY; COMPUTED-TOMOGRAPHY; RADIATION-THERAPY; CANCER; LOCALIZATION; DEVICES; ERRORS;
D O I
10.5603/RPOR.a2022.0089
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background: We evaluated the setup error and dose absorption of an immobilization system with a shell and wooden baseplate (SW) for lung stereotactic body radiotherapy (SBRT). Materials and methods: Setup errors in 109 patients immobilized with an SW or BodyFix system (BF) were compared. Dose attenuation rates of materials for baseplates were measured with an ion-chamber. Ionization measurements were performed from 90 degrees to 180 degrees gantry angle in 10 degrees increments, with the ball water equivalent phantom placed at the center of the wood and carbon baseplates whose effects on dose distribution were compared using an electron portal imaging device. Results: The ratio for the anterior-posterior, cranial-caudal, and right-left of the cases within 3-mm registered shifts in interfractional setup error were 90.9%, 89.2%, and 97.4% for the SW, and 93.2%, 91.6%, and 98.0% for the BF, respectively. For intrafractional setup error, 98.3%, 97.4%, and 99.1% for the SW and 96.6%, 95.8%, and 98.7% for the BF were within 3-mm registered shifts, respectively. In the center position, the average (minimum/maximum) dose attenuation rates from 90 degrees to 180 degrees for the wooden and carbon baseplates were 0.5 (0.1/2.8)% and 1.0 (-0.1/10.1)% with 6 MV, respectively. The gamma passing rates of 2%/2 mm for the wooden and carbon baseplates were 99.7% and 98.3% (p < 0.01). Conclusions: The immobilization system with an SW is effective for lung SBRT since it is comparable to the BF in setup accuracy. Moreover, the wooden baseplate had lower radiation attenuation rates and affected the dose distribution less than the carbon baseplate.
引用
收藏
页码:809 / 820
页数:12
相关论文
共 50 条
  • [41] Retrospective Correlation of Dose Accuracy with Regions of Local Failure for Early Stage Lung Cancer Patients Treated with Stereotactic Body Radiotherapy
    Devpura, S.
    Li, H.
    Liu, C.
    Fraser, C.
    Ajlouni, M.
    Movsas, B.
    Chetty, I.
    MEDICAL PHYSICS, 2016, 43 (06) : 3333 - 3333
  • [42] Dose to lung tumours treated with stereotactic radiotherapy
    Christensen, JJ
    Hansen, AT
    RADIOTHERAPY AND ONCOLOGY, 2004, 73 : S435 - S435
  • [43] Harmonization of dose prescription for lung stereotactic radiotherapy
    Beldjoudi, Guillaume
    Bosson, Fanny
    Bernard, Vivien
    Puel, Lise-Marie
    Martel-Lafay, Isabelle
    Ayadi, Myriam
    Tanguy, Ronan
    PHYSICS & IMAGING IN RADIATION ONCOLOGY, 2022, 24 : 65 - 70
  • [44] Optimizing dose prescription in stereotactic body radiotherapy for lung tumours using Monte Carlo dose calculation
    Widder, Joachim
    Hollander, Miranda
    Ubbels, Jan F.
    Bolt, Rene A.
    Langendijk, Johannes A.
    RADIOTHERAPY AND ONCOLOGY, 2010, 94 (01) : 42 - 46
  • [45] Stereotactic body radiotherapy for oligometastatic lung tumors
    Norihisa, Yoshiki
    Nagata, Yasushi
    Takayama, Kenji
    Matsuo, Yukinori
    Sakamoto, Takashi
    Sakamoto, Masato
    Mizowaki, Takashi
    Yano, Shinsuke
    Hiraoka, Masahiro
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2008, 72 (02): : 398 - 403
  • [46] Editorial: Stereotactic body radiotherapy for lung cancer
    Liu, An
    Kroeze, Stephanie G. C.
    FRONTIERS IN ONCOLOGY, 2023, 13
  • [47] Biologically Effective Dose Was Associated with Overall Survival in Stereotactic Body Radiotherapy for Lung Tumors
    Tanaka, O.
    Funaguchi, N.
    Toyoshi, S.
    Taniguchi, T.
    Ono, K.
    Kunishima, Y.
    Masui, Y.
    Matsuo, M.
    JOURNAL OF THORACIC ONCOLOGY, 2019, 14 (10) : S1092 - S1092
  • [48] The characteristics of dose at mass interface on lung cancer Stereotactic Body Radiotherapy (SBRT) simulation
    Wulansari, I. H.
    Wibowo, W. E.
    Pawiro, S. A.
    10TH INTERNATIONAL SEMINAR ON MEDICAL PHYSICS & 1ST AMDI INTERNATIONAL ONCOLOGY SYMPOSIUM, 2017, 851
  • [49] STEREOTACTIC BODY RADIOTHERAPY FOR OLIGOMETASTATIC LUNG TUMORS
    Beltramo, G.
    Bianchi, L. C.
    Bergantin, A.
    Martinotti, A. S.
    Bonfanti, P.
    Gerosa, C.
    Varoli, F.
    Locatelli, M. C.
    Mariani, P.
    LUNG CANCER, 2011, 71 : S45 - S45
  • [50] STEREOTACTIC BODY RADIOTHERAPY FOR OLIGOMETASTATIC DISEASE TO THE LUNG
    Ashman, Jonathan
    Goyal, Uma
    Vora, Sujay
    Clouser, Edward
    Jaroszewski, Dawn
    Dueck, Amylou
    Paripati, Harshita
    Wesselius, Lewis
    Rule, William
    Ross, Helen J.
    Schild, Steve
    JOURNAL OF THORACIC ONCOLOGY, 2013, 8 : S1138 - S1138