60 MHz PMN-PT based 1-3 Composite Transducer for IVUS Imaging

被引:42
|
作者
Yuan, Jian [1 ]
Rhee, Sorah [1 ]
Jiang, Xio Ning [2 ]
机构
[1] Boston Sci Imaging, Fremont, CA USA
[2] TRS Technol Inc, State Coll, PA USA
关键词
high frequency transducer; piezoelectric composite; microfabrication; single element; IVUS; PMN-PT;
D O I
10.1109/ULTSYM.2008.0164
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
High frequency transducers are desired for usage in catheter-based IVUS (intravascular ultrasound) imaging, which is the cross-sectional visualization of the interior of blood vessels in the human body. High frequency intravascular imaging benefits from the use of high sensitivity and broad bandwidths as they allow for high resolution imaging at increased depth of field. The 40 MHz piezo-composite micromachined ultrasound transducer (PC-MUT) technique developed by the authors in the past years demonstrated that the limitations of traditional dice-and-fill methodology for the fabrication of 1-3 composites could be overcome by the application of deep reactive ion etching (DRIE) of PMN-PT single crystals [1,2,3]. In this work, the fabrication and testing of higher frequency transducers and prototype catheters is presented. 1-3 PC-MUT composites were characterized and transducer devices with a target frequency of 60 MHz were prototyped and incorporated into functional catheters for imaging using a porcine animal model.
引用
收藏
页码:682 / +
页数:2
相关论文
共 50 条
  • [41] Homogenization of PMN-PT/epoxy 1-3 piezocomposites by resonator measurements and finite element analysis
    Kim, Jinwook
    Roh, Yongrae
    SENSORS AND ACTUATORS A-PHYSICAL, 2014, 206 : 97 - 106
  • [42] 3D printing PMN-PT textured ceramics for transducer applications
    Zheng, Kun
    Quan, Yi
    Ding, Dafei
    Zhuang, Jian
    Fei, Chunlong
    Zhao, Jinyan
    Wang, Lingyan
    Zhao, Tianlong
    Wang, Zhe
    Zhao, Yifan
    Wang, Chenying
    Jiang, Zhuangde
    Wu, Shanghua
    Ren, Wei
    CERAMICS INTERNATIONAL, 2024, 50 (23) : 51870 - 51876
  • [43] Parylene coating for 13 MHz 1-3 composite transducer performance enhancement
    Li, Zhangjian
    Shao, Weiwei
    Zhu, Xinle
    Lv, Jiabing
    Han, Zhile
    Cui, Yaoyao
    APPLIED ACOUSTICS, 2021, 174
  • [44] Equivalent properties of 1-3 piezocomposites made of PMN-PT single crystals for underwater SONAR transducers
    Kim, Jinwook
    Roh, Yongrae
    BEHAVIOR AND MECHANICS OF MULTIFUNCTIONAL MATERIALS AND COMPOSITES 2011, 2011, 7978
  • [45] PIN-PMN-PT Single-Crystal-Based 1-3 Piezoelectric Composites for Ultrasonic Transducer Applications
    Li, Lili
    Xu, Zhuo
    Xia, Song
    Li, Zhengrong
    Ji, Xuanrong
    Long, Shaojun
    JOURNAL OF ELECTRONIC MATERIALS, 2013, 42 (08) : 2564 - 2569
  • [46] Hybrid polymer matrix Terfenol-D composite/PMN-PT transducer in mechanical series configuration
    Mortensen, AP
    Dapino, MJ
    SMART STRUCTURES AND MATERIALS 2005: ACTIVE MATERIALS: BEHAVIOR AND MECHANICS, 2005, 5761 : 136 - 147
  • [47] A 15-MHz 1-3 Piezocomposite Concave Array Transducer for Ophthalmic Imaging
    Cha, Jung Hyui
    Kang, Byungwoo
    Jang, Jihun
    Chang, Jin Ho
    IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2015, 62 (11) : 1994 - 2004
  • [48] Single crystal PMN-0.33PT/epoxy 1-3 composites for ultrasonic transducer applications
    Cheng, KC
    Chan, HLW
    Choy, CL
    Yin, QR
    Luo, HS
    Yin, ZW
    IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2003, 50 (09) : 1177 - 1183
  • [49] Fabrication of flexible piezoelectric PMN-PT based composite films for energy harvesting
    Das, Satyabati
    Biswal, Asutya Kumar
    Roy, Amritendu
    NATIONAL CONFERENCE ON PROCESSING AND CHARACTERIZATION OF MATERIALS (NCPCM), 2017, 178
  • [50] Micromachined PMN-PT Single Crystal Composite Transducers-15-75 MHz PC-MUT
    Jiang, X.
    Snook, K.
    Cheng, A.
    Hackenberger, W. S.
    Geng, X.
    2008 IEEE ULTRASONICS SYMPOSIUM, VOLS 1-4 AND APPENDIX, 2008, : 164 - +