Parylene coating for 13 MHz 1-3 composite transducer performance enhancement

被引:6
|
作者
Li, Zhangjian [1 ,2 ]
Shao, Weiwei [1 ]
Zhu, Xinle [1 ]
Lv, Jiabing [1 ]
Han, Zhile [3 ]
Cui, Yaoyao [1 ]
机构
[1] Chinese Acad Sci, Suzhou Inst Biomed Engn & Technol, Suzhou 215163, Peoples R China
[2] Univ Sci & Technol China, Hefei 230027, Peoples R China
[3] Fudan Univ, Shanghai 200433, Peoples R China
基金
中国国家自然科学基金;
关键词
Matching Layer; 1-3 Composite Material; Piezoelectric Transducer; ULTRASONIC TRANSDUCERS; SINGLE-CRYSTAL; IMPEDANCE; LAYERS; PZT;
D O I
10.1016/j.apacoust.2020.107696
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
1-3 piezoelectric composite material is widely applied in medical ultrasound transducers. The acoustic impedance mismatching still exists between composite materials and human tissue, this mismatching limits the performance of the 1-3 composite transducer. In this work, Parylene coating technique is employed to enhance the performance of a 1-3 piezoelectric composite transducer whose center frequency is 13 MHz. Theory and simulation analysis is carried out to estimation and optimize the influence of the Parylene coating layer. Pulse- echo experiment is performed to verify the hypothesis of theory and simulation. The experiment results confirm that the performances of the transducer are effectively enhanced by Parylene coating and the optimum thickness of the coating Parylene layer is about 39.5 mu m. Compared to the uncoated 1-3 composite transducer, the echo amplitude of 39.5 mu m Parylene coated transducer increases 98.8%, the value of -6 dB pulse width decreases 39.5%, the value of -20 dB pulse width decreases 22.7%, the bandwidth increases 51.9%, and the center frequency is not affected. These results prove the feasibility and effectiveness of the Parylene coating layer in transducer performance enhancement. (C) 2020 Published by Elsevier Ltd.
引用
收藏
页数:6
相关论文
共 50 条
  • [1] Fabrication and Performance of a Micro 50-MHz IVUS Transducer Based on a 1-3 Composite with Geometric Focusing
    Jian, Xiaohua
    Han, Zhile
    Liu, Pengbo
    Li, Zhangjian
    Li, Peiyang
    Shao, Wei
    Cui, Yaoyao
    2015 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2015,
  • [2] 60 MHz PMN-PT based 1-3 Composite Transducer for IVUS Imaging
    Yuan, Jian
    Rhee, Sorah
    Jiang, Xio Ning
    2008 IEEE ULTRASONICS SYMPOSIUM, VOLS 1-4 AND APPENDIX, 2008, : 682 - +
  • [3] ELECTROACOUSTIC CHARACTERISTICS OF A 1-3 COMPOSITE TRANSDUCER
    THOMPSON, W
    JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1987, 81 : S88 - S88
  • [4] Theoretical calculation model of emission performance for 1-3 piezoelectric composite transducer
    Zhong, Chao
    Wang, Likun
    Hao, Shaohua
    Sun, Ruiqing
    Qin, Lei
    FERROELECTRICS, 2021, 582 (01) : 185 - 191
  • [5] DESIGN FOR AE TRANSDUCER BY 1-3 PIEZOELECTRIC COMPOSITE
    OHARA, Y
    SHIWA, M
    YANAGIDA, H
    KISHI, T
    NIPPON SERAMIKKUSU KYOKAI GAKUJUTSU RONBUNSHI-JOURNAL OF THE CERAMIC SOCIETY OF JAPAN, 1994, 102 (04): : 368 - 373
  • [6] Simulation of 1-3 Piezoelectric Composite Cylindrical Transducer
    Li Li
    Wang Likun
    Qin Lei
    FERROELECTRICS, 2014, 471 (01) : 17 - 22
  • [7] Study of 1-3 composite transducer for ultrasonic wirebonding application
    Chong, CP
    Li, HL
    Chan, HLW
    Liu, PCK
    CERAMICS INTERNATIONAL, 2004, 30 (07) : 1141 - 1146
  • [8] Performance enhancement of 1-3 piezoelectric composite materials by alternating current polarising
    Ma, Jinpeng
    Zhu, Ke
    Huo, Da
    Shen, Bingzhong
    Liu, Yang
    Qi, Xudong
    Sun, Enwei
    Zhang, Rui
    CERAMICS INTERNATIONAL, 2021, 47 (13) : 18405 - 18410
  • [9] A 15-MHz 1-3 Piezocomposite Concave Array Transducer for Ophthalmic Imaging
    Cha, Jung Hyui
    Kang, Byungwoo
    Jang, Jihun
    Chang, Jin Ho
    IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2015, 62 (11) : 1994 - 2004
  • [10] Performance Analysis of the 1-3 Piezoelectric Composites and transducer Fabrication
    Sun Min
    Xu Dongyu
    Huang Shifeng
    FUNCTIONAL AND ELECTRONIC MATERIALS, 2011, 687 : 339 - 342