Comparison of Skull Radiograph and Computed Tomography Measurements of Cochlear Implant Insertion Angles

被引:10
|
作者
Gallant, Sara [1 ]
Friedmann, David R. [1 ]
Hagiwara, Mari [2 ]
Roland, J. Thomas, Jr. [1 ]
Svirsky, Mario A. [1 ]
Jethanamest, Daniel [1 ]
机构
[1] NYU, Sch Med, Dept Otolaryngol, New York, NY 10003 USA
[2] NYU, Sch Med, Dept Radiol, New York, NY 10003 USA
关键词
Angular depth of insertion; Cochlear implant; Cochlear implant imaging; POSITION;
D O I
10.1097/MAO.0000000000002121
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Background: Measurement of the angular depth of insertion (aDOI) of cochlear implant electrode arrays has numerous clinical and research applications. Plain-film radiographs are easily obtained intraoperatively and have been described as a means to calculate aDOI. CT imaging with 3D reformatting can also be used for this measurement, but is less conveniently obtained and requires higher radiation doses, a particular concern in pediatrics. The extent to which plain-film and 3D CT image-based measurements are representative of the true position of the electrode within the cochlea is unknown. Methods: Cochlear implantation was performed on 10 cadaveric temporal bones. Five bones were implanted with perimodiolar electrodes (Contour Advance TM, Cochlear, Sydney, Australia) and five were implanted with lateral wall electrodes (Slim Straight, Cochlear). The insertion depths of the electrodes were varied. Each bone was imaged with a radiograph and CT. aDOI was measured for each bone in each imaging modality by a neurotologist and a neuroradiologist. To obtain a "gold standard" estimate of aDOI, the implanted temporal bones were embedded in an epoxy resin and methodically sectioned at 100 mm intervals; histologic images were captured at each interval. A 3D stack of the images was compounded, and a MATLAB script used to calculate aDOI of the most apical electrode. Measurements in the three modalities (radiograph, CT, and histology) were then compared. Results: The average aDOI across all bones was similar for all modalities: 4238 for radiographs, 4258 for CT scans, and 4278 for histology, indicating that neither imaging modality resulted in large systematic errors. Using the histology-measured angles as a reference, the average error for CT-based measures (regardless of whether the error was in the positive or negative direction) was 128, and that for radiograph-based measures was 158. This small difference (12 vs 158 error) was not statistically significant. Conclusion: Based on this cadaveric temporal bone model, both radiographs and CTs can provide reasonably accurate aDOI measurements. In this small sample, and as expected, the CT-based estimates were more accurate than the radiograph-based measurements. However, the difference was small and not statistically significant. Thus, the use of plain radiographs to calculate aDOI seems judicious whenever it is desired to prevent unnecessary radiation exposure and expense.
引用
收藏
页码:E298 / E303
页数:6
相关论文
共 50 条
  • [41] Measurement of Cochlear Implant Electrode Position From Intraoperative Post-insertion Skull Radiographs: A Validation Study
    Svrakic, Maja
    Friedmann, David R.
    Berman, Phillip M.
    Davis, Adam J.
    Roland, J. Thomas, Jr.
    Svirsky, Mario A.
    OTOLOGY & NEUROTOLOGY, 2015, 36 (09) : 1486 - 1491
  • [42] A comparison of the diagnostic ability between Waters' radiograph and computed tomography in the diagnosis of midfacial fracture
    Jeon, IS
    Choi, SC
    Park, TW
    You, DS
    JOURNAL OF DENTAL RESEARCH, 1997, 76 (05) : 1190 - 1190
  • [43] A comparison of aortic root measurements by echocardiography and computed tomography
    Plonek, Tomasz
    Berezowski, Mikolaj
    Bochenek, Maciej
    Filip, Grzegorz
    Rylski, Bartosz
    Golesworthy, Tal
    Jasinski, Marek
    JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2019, 157 (02): : 479 - 486
  • [44] Detection of intracochlear damage during cochlear implant electrode insertion using extracochlear measurements in the gerbil
    Ahmad, Faisal I.
    Choudhury, Baishakhi
    De Mason, Christine E.
    Adunka, Oliver F.
    Finley, Charles C.
    Fitzpatrick, Douglas C.
    LARYNGOSCOPE, 2012, 122 (03): : 636 - 644
  • [45] Comparison of bone thickness in infrazygomatic crest area at various miniscrew insertion angles in Dravidian population - A cone beam computed tomography study
    Murugesan, Arathi
    Sivakumar, Arvind
    INTERNATIONAL ORTHODONTICS, 2020, 18 (01) : 105 - 114
  • [46] Comparison of radiation dose for implant imaging using conventional spiral tomography, computed tomography, and cone-beam computed tomography
    Chau, Anson C. M.
    Fung, Karl
    ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY AND ENDODONTOLOGY, 2009, 107 (04): : 559 - 565
  • [47] Cochlear implant positioning: development and validation of an automatic method using computed tomography image analysis
    Kemper, Erik H. M.
    Markodimitraki, Laura M.
    Magre, Joell
    Simons, Dominique C.
    Thomeer, Hans G. X. M.
    FRONTIERS IN SURGERY, 2024, 11
  • [48] Utility of Preoperative Computed Tomography and Magnetic Resonance Imaging in Adult and Pediatric Cochlear Implant Candidates
    Tamplen, Matthew
    Schwalje, Adam
    Lustig, Lawrence
    Alemi, Aurash S.
    Miller, Mia E.
    LARYNGOSCOPE, 2016, 126 (06): : 1440 - 1445
  • [49] Intraoperative Evaluation of Cochlear Implant Electrodes Using Mobile Cone-Beam Computed Tomography
    Yamamoto, Norio
    Okano, Takayuki
    Yamazaki, Hiroshi
    Hiraumi, Harukazu
    Sakamoto, Tatsunori
    Ito, Juichi
    Omori, Koichi
    OTOLOGY & NEUROTOLOGY, 2019, 40 (02) : 177 - 183
  • [50] Merged Volume Rendered Flat-panel Computed Tomography for Postoperative Cochlear Implant Assessment
    Eisenhut, Felix
    Lang, Stefan
    Taha, Lava
    Doerfler, Arnd
    Iro, Heinrich
    Hornung, Joachim
    CLINICAL NEURORADIOLOGY, 2020, 30 (04) : 721 - 728