Validation of a new noninvasive method to measure blood pressure and assess baroreflex sensitivity in preterm infants during sleep

被引:30
|
作者
Yiallourou, Stephanie R. [1 ]
Walker, Adrian M. [1 ]
Horne, Rosemary S. C. [1 ]
机构
[1] Monash Univ, Monash Inst Med Res, Ritchie Ctr Baby Hlth Res, Melbourne, Vic 3004, Australia
关键词
blood pressure; baroreflex; sleep; cardiovascular control;
D O I
10.1093/sleep/29.8.1083
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Study Objectives: Accuracy and precision of a noninvasive device for continuously measuring blood pressure (BP) (Finometer (TM), FMS, The Netherlands) during sleep was assessed in preterm infants. Design: Absolute BP beat-to-beat values, interbeat changes, measurement precision, and baroreflex sensitivity were compared with BP measurements from intraarterial catheters. Participants: Ten preterm infants (gestational age 27-36 weeks; birth weight 964-2620 gm) were studied in the neonatal intensive care unit. Measurements and Results: The 2 modes of BP measurement were compared in 2-minute epochs (n = 10-12/infant). Mean arterial pressure, systolic arterial pressure, and diastolic arterial pressure were analyzed beat to beat, and baroreflex sensitivity was assessed using spontaneous sequence analysis. Mean differences for absolute BP (mm Hg) were as follows: mean arterial pressure, 3 (limits of agreement, -1 to 8); systolic arterial pressure, -4 (-8 to 1); and diastolic arterial pressure, 7 (4 to 10). Mean differences and limits of agreement for interbeat changes were essentially 0 for mean arterial pressure, systolic arterial pressure, and diastolic arterial pressure. Precision ( 95% confidence intervals, mm Hg) for the Finometer (TM) were mean arterial pressure +/- 7, systolic arterial pressure 8, and diastolic arterial pressure +/- 6. Precision was greater for the arterial catheter (mean arterial pressure 3, systolic arterial pressure 4, and diastolic arterial pressure 4). Baroreflex sensitivity calculated from the Finometer (TM) BP was (mean +/- SEM, ms/mm Hg) 1.74 +/- 0.23 and, from the catheter system, BP was 1.56 +/- 0.21 (p value NS). Conclusions: The Finometer (TM) provides accurate measurements of beat-to-beat BP and baroreflex sensitivity. The ability to continuously measure BP and baroreflex sensitivity during sleep in infants may provide vital clues into pathologic conditions associated with impaired autonomic control during sleep.
引用
收藏
页码:1083 / 1088
页数:6
相关论文
共 50 条
  • [21] NONINVASIVE BLOOD-PRESSURE MEASUREMENT IN NEONATES AND INFANTS WITH THE OSCILLOMETRIC METHOD
    PAPADOPOULOS, G
    ANAESTHESIST, 1994, 43 (07): : 441 - 446
  • [22] Wavelet Analysis of Cardiac Baroreflex Sensitivity during Stepwise Blood Pressure Changes
    Tarumi, Takashi
    Tian, Fenghua
    Xing, Chang-Yang
    Turner, Marcel
    Repshas, Justin
    Zhang, Rong
    FASEB JOURNAL, 2017, 31
  • [23] Blood pressure normative values in preterm infants during postnatal transition
    Elsayed, Yasser
    Ahmed, Faraz
    PEDIATRIC RESEARCH, 2024, 95 (03) : 698 - 704
  • [24] Blood pressure normative values in preterm infants during postnatal transition
    Yasser Elsayed
    Faraz Ahmed
    Pediatric Research, 2024, 95 : 698 - 704
  • [25] Noninvasive method to validate the variability of blood pressure during arrhythmias
    Tomonori Watanabe
    Satoshi Hoshide
    Kazuomi Kario
    Hypertension Research, 2022, 45 : 530 - 532
  • [26] Noninvasive method to validate the variability of blood pressure during arrhythmias
    Watanabe, Tomonori
    Hoshide, Satoshi
    Kario, Kazuomi
    HYPERTENSION RESEARCH, 2022, 45 (03) : 530 - 532
  • [27] Quantitative validation of the suprasternal pressure signal to assess respiratory effort during sleep
    Cerina, Luca
    Papini, Gabriele B.
    Fonseca, Pedro
    Overeem, Sebastiaan
    van Dijk, Johannes P.
    van Meulen, Fokke
    Vullings, Rik
    PHYSIOLOGICAL MEASUREMENT, 2024, 45 (05)
  • [28] Baroreflex Sensitivity and Blood Pressure in Premature Infants-Dependence on Gestational Age, Postnatal Age and Sex
    Javorka, Kamil
    Haskova, Katarina
    Czippelova, Barbora
    Zibolen, Mirko
    Javorka, Michal
    PHYSIOLOGICAL RESEARCH, 2021, 70 : S349 - S356
  • [29] Baroreflex sensitivity, heart rate and blood pressure during mental load and emotional activation
    Honzikova, N
    Rybkova, I
    Honzik, JM
    HOMEOSTASIS IN HEALTH AND DISEASE, 1996, 37 (1-2): : 72 - 76
  • [30] Blood pressure variability during sleep in infants: effect sleep state and sleeping position
    Yiallourou, S. R.
    Walker, A. M.
    Horne, R. S.