Relationship between size-specific dose estimates and image quality in computed tomography depending on patient size

被引:16
|
作者
Kawashima, Hiroki [1 ]
Ichikawa, Katsuhiro [1 ]
Hanaoka, Shinsuke [2 ]
Matsubara, Kosuke [1 ]
Takata, Tadanori [2 ]
机构
[1] Kanazawa Univ, Inst Med Pharmaceut & Hlth Sci, Fac Hlth Sci, Kanazawa, Ishikawa, Japan
[2] Kanazawa Univ Hosp, Radiol Div, Kanazawa, Ishikawa, Japan
来源
基金
日本学术振兴会;
关键词
auto exposure control; computed tomography; contrast-to-noise ratio; image quality; patient size; size-specific dose estimates; AUTOMATIC EXPOSURE CONTROL; TUBE CURRENT MODULATION; PEDIATRIC CT; PHANTOM; OPTIMIZATION; MDCT; VOLTAGE; SYSTEMS; LIVER;
D O I
10.1002/acm2.12340
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
This study investigates the relationship between contrast-to-noise ratio (CNR) and size-specific dose estimate (SSDE) in computed tomography (CT) depending on patient size. In addition, the relationship to the auto exposure control (AEC) techniques is examined. A tissue-equivalent material having human-liver energy dependence is developed and used to evaluate these relationships. Three exposure dose levels (constant CT dose index, constant SSDE, and with AEC) are tested using four different phantom sizes (diameter: 15, 20, 25 and 30cm) in two different CT scanners (SOMATOM Definition Flash, Siemens, and LightSpeed VCT, GE). The contrast-to-noise ratios (CNRs) are measured using the developed phantom. It is found that the CNR increases with decreasing phantom size at constant SSDE, although the increase ratio is smaller than that of the constant CT dose index. This result indicates that the image characteristics differ even when the patient dose received from the CT examination is equivalent for each patient size. In the case of AEC use, the CNR results of the Siemens scanner exhibit a similar trend to those obtained for constant SSDE, for each phantom size. This suggests that the AEC technique that maintains a constant image quality (CARE Dose 4D) for each patient size corresponds well to the image quality obtained for constant SSDE. These findings facilitate further understanding of the relationship between image quality and exposure CT dose depending on patient size.
引用
收藏
页码:246 / 251
页数:6
相关论文
共 50 条
  • [41] Size-specific dose estimates for various weighting factors of CTDI equation
    Tomonobu Haba
    Masanao Kobayashi
    Shuji Koyama
    Physical and Engineering Sciences in Medicine, 2020, 43 : 155 - 162
  • [42] Size-specific dose estimates for various weighting factors of CTDI equation
    Haba, Tomonobu
    Kobayashi, Masanao
    Koyama, Shuji
    PHYSICAL AND ENGINEERING SCIENCES IN MEDICINE, 2020, 43 (01) : 155 - 162
  • [43] Variations of Size-Specific Dose Estimates (SSDE) Factors with Scan Mode
    Abuhaimed, A.
    Martin, C.
    MEDICAL PHYSICS, 2019, 46 (06) : E441 - E442
  • [44] The impact of patient table on size-specific dose estimate (SSDE)
    Anam, Choirul
    Haryanto, Freddy
    Widita, Rena
    Arif, Idam
    Dougherty, Geoff
    McLean, Donald
    AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2017, 40 (01) : 153 - 158
  • [45] The impact of patient table on size-specific dose estimate (SSDE)
    Choirul Anam
    Freddy Haryanto
    Rena Widita
    Idam Arif
    Geoff Dougherty
    Donald McLean
    Australasian Physical & Engineering Sciences in Medicine, 2017, 40 : 153 - 158
  • [46] Establishment of baseline size-specific dose estimate (SSDE) for paediatric head computed tomography (CT) examinations
    Hamza Sulemana
    Abdul Nashirudeen Mumuni
    Izre-deen Osman Mohammed Abubakari
    Egyptian Journal of Radiology and Nuclear Medicine, 55 (1):
  • [47] Variation of Size-Specific Dose Estimates Across Patient Sizes Under the Conditions of Automatic Exposure Control
    McMillan, K.
    Huang, A.
    Leng, S.
    McCollough, C.
    MEDICAL PHYSICS, 2016, 43 (06) : 3861 - 3861
  • [48] Can Patient's Body Weight Represent Body Diameter for Pediatric Size-Specific Dose Estimate in Thoracic and Abdominal Computed Tomography?
    Kritsaneepaiboon, Supika
    Eng-chuan, Suwadee
    Yoykaew, Saowapark
    JOURNAL OF CLINICAL IMAGING SCIENCE, 2019, 9 : 1 - 8
  • [49] Evaluation of Size-Specific Dose Estimates for Optimizing Pediatric Chest CT Protocol
    Zarei F.
    Nasiri M.
    Etemadi Z.
    Haghighi R.R.
    Chatterjee S.
    Khaneghah P.A.
    Chatterjee V.V.
    Iranian Journal of Medical Physics, 2022, 19 (05): : 315 - 321
  • [50] Size-Specific Dose Estimates (SSDE) Modified for Lung Inhomogeneities: A Practical Approach
    Mihailidis, D.
    Tomara, P.
    Kagadis, G.
    MEDICAL PHYSICS, 2013, 40 (06)