Relationship between size-specific dose estimates and image quality in computed tomography depending on patient size

被引:16
|
作者
Kawashima, Hiroki [1 ]
Ichikawa, Katsuhiro [1 ]
Hanaoka, Shinsuke [2 ]
Matsubara, Kosuke [1 ]
Takata, Tadanori [2 ]
机构
[1] Kanazawa Univ, Inst Med Pharmaceut & Hlth Sci, Fac Hlth Sci, Kanazawa, Ishikawa, Japan
[2] Kanazawa Univ Hosp, Radiol Div, Kanazawa, Ishikawa, Japan
来源
基金
日本学术振兴会;
关键词
auto exposure control; computed tomography; contrast-to-noise ratio; image quality; patient size; size-specific dose estimates; AUTOMATIC EXPOSURE CONTROL; TUBE CURRENT MODULATION; PEDIATRIC CT; PHANTOM; OPTIMIZATION; MDCT; VOLTAGE; SYSTEMS; LIVER;
D O I
10.1002/acm2.12340
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
This study investigates the relationship between contrast-to-noise ratio (CNR) and size-specific dose estimate (SSDE) in computed tomography (CT) depending on patient size. In addition, the relationship to the auto exposure control (AEC) techniques is examined. A tissue-equivalent material having human-liver energy dependence is developed and used to evaluate these relationships. Three exposure dose levels (constant CT dose index, constant SSDE, and with AEC) are tested using four different phantom sizes (diameter: 15, 20, 25 and 30cm) in two different CT scanners (SOMATOM Definition Flash, Siemens, and LightSpeed VCT, GE). The contrast-to-noise ratios (CNRs) are measured using the developed phantom. It is found that the CNR increases with decreasing phantom size at constant SSDE, although the increase ratio is smaller than that of the constant CT dose index. This result indicates that the image characteristics differ even when the patient dose received from the CT examination is equivalent for each patient size. In the case of AEC use, the CNR results of the Siemens scanner exhibit a similar trend to those obtained for constant SSDE, for each phantom size. This suggests that the AEC technique that maintains a constant image quality (CARE Dose 4D) for each patient size corresponds well to the image quality obtained for constant SSDE. These findings facilitate further understanding of the relationship between image quality and exposure CT dose depending on patient size.
引用
收藏
页码:246 / 251
页数:6
相关论文
共 50 条
  • [21] Patient size matters: Effect of tube current modulation on size-specific dose estimates (SSDE) and image quality in low-dose lung cancer screening CT
    Barreto, Izabella
    Verma, Nupur
    Quails, Nathan
    Olguin, Catherine
    Correa, Nathalie
    Mohammed, Tan-Lucien
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2020, 21 (04): : 87 - 94
  • [22] Effects of Metal On CT Size-Specific Dose Estimates
    Alsanea, M.
    Krugh, K.
    Layman, R.
    MEDICAL PHYSICS, 2016, 43 (06) : 3720 - 3720
  • [23] Novel method for calculating the effective dose using size-specific dose estimates conversion factors in abdomen–pelvis computed tomography
    Kentaro Funashima
    Shigeru Abiko
    Kazuhiro Sato
    Radiological Physics and Technology, 2023, 16 (4) : 506 - 515
  • [24] Determination of radiation dose during computed tomography examinations in southern Saudi Arabian hospitals using size-specific dose estimates
    Saeed, Mohammed K.
    Asiri, Abdullah Ali M.
    MAEJO INTERNATIONAL JOURNAL OF SCIENCE AND TECHNOLOGY, 2021, 15 (02) : 147 - 158
  • [25] Size-Specific Dose Estimates Based on Water-Equivalent Diameter and Effective Diameter in Computed Tomography Coronary Angiography
    Xu, Jian
    Wang, Xiangquan
    Xiao, Huawei
    Xu, Jianguo
    MEDICAL SCIENCE MONITOR, 2019, 25 : 9299 - 9305
  • [26] Factors influencing size-specific dose estimates of selected computed tomography protocols at two clinical practices in South Africa
    Sebelego, Ida-Keshia
    Acho, Sussan
    van der Merwe, Belinda
    Rae, William I. D.
    RADIATION PROTECTION DOSIMETRY, 2023, 199 (07) : 588 - 602
  • [27] Institutional computed tomography diagnostic reference levels based on water-equivalent diameter and size-specific dose estimates
    Boos, Johannes
    Thomas, Christoph
    Appel, Elisabeth
    Klosterkemper, Yan
    Schleich, Christoph
    Aissa, Joel
    Bethge, Oliver Th
    Antoch, Gerald
    Kroepil, Patric
    JOURNAL OF RADIOLOGICAL PROTECTION, 2018, 38 (02) : 536 - 548
  • [28] ACCURACY OF SIZE-SPECIFIC DOSE ESTIMATE CALCULATION FROM CENTER SLICE IN COMPUTED TOMOGRAPHY
    Boos, Johannes
    Kroepil, Patric
    Bethge, Oliver Th.
    Aissa, Joel
    Schleich, Christoph
    Sawicki, Lino Morris
    Heinzler, Niklas
    Antoch, Gerald
    Thomas, Christoph
    RADIATION PROTECTION DOSIMETRY, 2018, 178 (01) : 8 - 19
  • [29] Size-Specific Dose Estimates for Evaluation of Individual Patient Dose in CT Protocol for Renal Colic
    Waszczuk, Lukasz A.
    Guzinski, Maciej
    Czarnecka, Anna
    Sasiadek, Marek J.
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2015, 205 (01) : 100 - 105
  • [30] Novel method for calculating the effective dose using size-specific dose estimates conversion factors in abdomen-pelvis computed tomography
    Funashima, Kentaro
    Abiko, Shigeru
    Sato, Kazuhiro
    RADIOLOGICAL PHYSICS AND TECHNOLOGY, 2023, 16 (04) : 506 - 515