Relationship between size-specific dose estimates and image quality in computed tomography depending on patient size

被引:16
|
作者
Kawashima, Hiroki [1 ]
Ichikawa, Katsuhiro [1 ]
Hanaoka, Shinsuke [2 ]
Matsubara, Kosuke [1 ]
Takata, Tadanori [2 ]
机构
[1] Kanazawa Univ, Inst Med Pharmaceut & Hlth Sci, Fac Hlth Sci, Kanazawa, Ishikawa, Japan
[2] Kanazawa Univ Hosp, Radiol Div, Kanazawa, Ishikawa, Japan
来源
基金
日本学术振兴会;
关键词
auto exposure control; computed tomography; contrast-to-noise ratio; image quality; patient size; size-specific dose estimates; AUTOMATIC EXPOSURE CONTROL; TUBE CURRENT MODULATION; PEDIATRIC CT; PHANTOM; OPTIMIZATION; MDCT; VOLTAGE; SYSTEMS; LIVER;
D O I
10.1002/acm2.12340
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
This study investigates the relationship between contrast-to-noise ratio (CNR) and size-specific dose estimate (SSDE) in computed tomography (CT) depending on patient size. In addition, the relationship to the auto exposure control (AEC) techniques is examined. A tissue-equivalent material having human-liver energy dependence is developed and used to evaluate these relationships. Three exposure dose levels (constant CT dose index, constant SSDE, and with AEC) are tested using four different phantom sizes (diameter: 15, 20, 25 and 30cm) in two different CT scanners (SOMATOM Definition Flash, Siemens, and LightSpeed VCT, GE). The contrast-to-noise ratios (CNRs) are measured using the developed phantom. It is found that the CNR increases with decreasing phantom size at constant SSDE, although the increase ratio is smaller than that of the constant CT dose index. This result indicates that the image characteristics differ even when the patient dose received from the CT examination is equivalent for each patient size. In the case of AEC use, the CNR results of the Siemens scanner exhibit a similar trend to those obtained for constant SSDE, for each phantom size. This suggests that the AEC technique that maintains a constant image quality (CARE Dose 4D) for each patient size corresponds well to the image quality obtained for constant SSDE. These findings facilitate further understanding of the relationship between image quality and exposure CT dose depending on patient size.
引用
收藏
页码:246 / 251
页数:6
相关论文
共 50 条
  • [1] Size-Specific Dose Estimates in Fetal Computed Tomography
    Kobayashi, Masanao
    Nishihara, Yusei
    Haba, Tomonobu
    Matsunaga, Yuta
    Minami, Kazuyuki
    Asada, Yasuki
    RADIATION PROTECTION DOSIMETRY, 2022, 198 (06) : 339 - 348
  • [2] SIMPLE METHOD OF SIZE-SPECIFIC DOSE ESTIMATES CALCULATION FROM PATIENT WEIGHT ON COMPUTED TOMOGRAPHY
    Iriuchijima, Akiko
    Fukushima, Yasuhiro
    Nakajima, Takahito
    Tsushima, Yoshito
    Ogura, Akio
    RADIATION PROTECTION DOSIMETRY, 2018, 178 (02) : 208 - 212
  • [3] Using a Somatosensory Controller to Assess Body Size for Size-Specific Dose Estimates in Computed Tomography
    Wu, Jay
    Han, Ruo-Ping
    Liu, Yan-Lin
    BIOMED RESEARCH INTERNATIONAL, 2018, 2018
  • [4] A Segmentation-based Automated Calculation of Patient Size and Size-specific Dose Estimates in Pediatric Computed Tomography Scans
    Abdulkadir, Muhammad Kabir
    Osman, Noor Diyana
    Achuthan, Anusha
    Nasirudin, Radin A.
    Ahmad, Muhammad Zabidi
    Zain, Noor Hasyima Mat
    Shuaib, Ibrahim Lutfi
    JOURNAL OF MEDICAL PHYSICS, 2024, 49 (03) : 456 - 463
  • [5] Size-specific dose estimates: Localizer or transverse abdominal computed tomography images?
    Pourjabbar, Sarvenaz
    Singh, Sarabjeet
    Padole, Atul
    Saini, Akshay
    Blake, Michael A.
    Kalra, Mannudeep K.
    WORLD JOURNAL OF RADIOLOGY, 2014, 6 (05): : 210 - 217
  • [6] Multivariate Analysis of Effective Dose and Size-Specific Dose Estimates for Thorax and Abdominal Computed Tomography
    Shah, Mudasir Ashraf
    Ahmad, Mehtab
    Khalid, Saifullah
    Qaseem, Syed M. Danish
    Siddiqui, Shaista
    Talib, Sayema
    Rather, Sajad Ahmed
    Firdous, Arfat
    JOURNAL OF MEDICAL PHYSICS, 2023, 48 (02) : 210 - 218
  • [7] Size-specific dose estimates (SSDEs) for computed tomography and influencing factors on it: a systematic review
    Satharasinghe, D. M.
    Jeyasugiththan, J.
    Wanninayake, W. M. N. M. B.
    Pallewatte, A. S.
    JOURNAL OF RADIOLOGICAL PROTECTION, 2021, 41 (04) : R108 - R124
  • [8] Development of size-specific dose estimates for common computed tomography examinations: a study in Ghana
    Ohene-Botwe, Benard
    Anim-Sampong, Samuel
    Nkansah, Josephine
    JOURNAL OF RADIOLOGICAL PROTECTION, 2023, 43 (01)
  • [9] Size-Specific Dose Estimates for Thoracic and Abdominal Computed Tomography Examinations at Two Moroccan Hospitals
    El Mansouri, M'hamed
    Choukri, Abdelmajid
    Semghouli, Slimane
    Talbi, Mohammed
    Eddaoui, Khalida
    Saga, Zouhir
    JOURNAL OF DIGITAL IMAGING, 2022, 35 (06) : 1648 - 1653
  • [10] Accuracy of patient dose estimation in cone beam computed tomography in breast irradiation by size-specific dose estimates with position correction
    Ueno, Hiroyuki
    Matsubara, Kosuke
    Bou, Sayuri
    Hizume, Masato
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2022, 23 (12):