Comparison of Goldmann Applanation Tonometry and Dynamic Contour Tonometry in the Measurement of Intraocular Pressure in Eyes with Different Corneal Thicknesses

被引:8
|
作者
Roszkowska, A. M. [1 ]
De Grazia, L. [1 ]
Cirone, M. [1 ]
Ferreri, G. [1 ]
机构
[1] Univ Messina, Ophthalmol Clin, Dept Surg Specialties, Messina, Italy
关键词
Intraocular pressure; Goldmann applanation tonometer; Dynamic contour tonometer; Corneal thickness; LASER PHOTOREFRACTIVE KERATECTOMY; IN-SITU KERATOMILEUSIS; HUMAN CADAVER EYES; GLAUCOMA; PATIENT; MYOPIA;
D O I
10.1159/000209253
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To compare Goldmann applanation tonometry (GAT) and dynamic contour tonometry (DCT) in the evaluation of intraocular pressure (IOP) in eyes with different corneal thicknesses. Methods: GAT and DCT were performed in 70 eyes of 35 subjects aged between 21 and 59 years. Thirty eyes had normal, nonoperated corneas and 40 eyes had reduced central corneal thickness (CCT) following myopic photorefractive keratectomy. The GAT and DCT values were correlated to CCT and age and the agreement between both tonometers was determined by the Bland and Altman method. Results: GAT correlated strongly with CCT (p = 0.0002), but there was no correlation between DCT and CCT (p = 0.4388). No correlation was found between IOP values obtained with both devices and age. The Bland and Altman plot evidenced a lack of agreement between both tonometers with 95% limits of agreement between 1.6 and -5.6 mm Hg. Conclusions: GAT gives significantly lower values in the eyes with thin corneas and DCT is not influenced by CCT. DCT could be recommended to evaluate IOP in eyes with out-of-range corneal thickness. Copyright (C) 2009 S. Karger AG, Basel
引用
收藏
页码:244 / 249
页数:6
相关论文
共 50 条
  • [21] Comparison of dynamic contour tonometry with Goldmann applanation tonometry in glaucoma practice
    Halkiadakis, Ioannis
    Patsea, Eleni
    Chatzimichali, Katerina
    Skouriotis, Sotiris
    Chalkidou, Sonia
    Amariotakis, Georgios
    Papakonstadinou, Dimitrios
    Theodossiadis, George
    Amariotakis, Apostolos
    Georgopoulos, Gerasimos
    ACTA OPHTHALMOLOGICA, 2009, 87 (03) : 323 - 328
  • [22] Goldmann applanation tonometry and dynamic contour tonometry in eyes with elevated intraocular pressure (IOP): comparison in the same eyes after subsequent medical normalization of IOP
    Chungkwon Yoo
    Young Sub Eom
    Yong Yeon Kim
    Graefe's Archive for Clinical and Experimental Ophthalmology, 2010, 248 : 1611 - 1616
  • [23] Comparisons of intraocular pressure measurements: Goldmann applanation tonometry, noncontact tonometry, Tono-Pen tonometry, and dynamic contour tonometry
    S-Y Hsu
    M-M Sheu
    A-H Hsu
    K-Y Wu
    J-I Yeh
    J-N Tien
    R-K Tsai
    Eye, 2009, 23 : 1582 - 1588
  • [24] Comparison of Goldmann applanation tonometry and Pascal dynamic contour tonometry in relation to central corneal thickness and corneal curvature
    Andreanos, Konstantinos
    Koutsandrea, Chryssanthi
    Papaconstantinou, Dimitris
    Diagourtas, Andreas
    Kotoulas, Andreas
    Dimitrakas, Panagiotis
    Moschos, Marilita M.
    CLINICAL OPHTHALMOLOGY, 2016, 10 : 2477 - 2484
  • [25] The relative effects of corneal thickness and age on Goldmann applanation tonometry and dynamic contour tonometry
    Kotecha, A
    White, ET
    Shewry, JM
    Garway-Heath, DF
    BRITISH JOURNAL OF OPHTHALMOLOGY, 2005, 89 (12) : 1572 - 1575
  • [26] Effects of corneal thickness, corneal curvature, and intraocular pressure level on Goldmann applanation tonometry and dynamic contour tonomet
    Francis, Brian A.
    Hsieh, Amy
    Lai, Mei-Ying
    Chopra, Vikas
    Pena, Fernando
    Azen, Stanley
    Varma, Rohit
    OPHTHALMOLOGY, 2007, 114 (01) : 20 - 26
  • [27] Tonometry in corneal edema after cataract surgery: dynamic contour tonometry versus Goldmann applanation tonometry
    Herr, Anne
    Remky, Andreas
    Hirsch, Thalia
    Rennings, Corinna
    Plange, Niklas
    CLINICAL OPHTHALMOLOGY, 2013, 7 : 815 - 819
  • [28] Effect of Corneal Morphometry on Dynamic Contour and Goldmann Applanation Tonometry
    Saenz-Frances, Federico
    Janez, Luis
    Borrego-Sanz, Lara
    Maria Martinez-de-la-Casa, Jose
    Jerez-Fidalgo, Maria
    Garcia-Sanchez, Julian
    Garcia-Feijoo, Julian
    JOURNAL OF GLAUCOMA, 2013, 22 (05) : 380 - 383
  • [29] The relative effects of corneal thickness and age on Goldmann applanation tonometry and dynamic contour tonometry
    Kotecha, A
    White, ET
    Shewry, JM
    Garway-Heath, DF
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2005, 46
  • [30] Comparison of dynamic contour tonometry and Goldmann applanation tonometry following penetrating keratoplasty
    Kandarakis, Artemios
    Soumplis, Vasileios
    Pitsas, Christos
    Kandarakis, Stylianos
    Halikias, Jiannis
    Karagiannis, Dimitrios
    CANADIAN JOURNAL OF OPHTHALMOLOGY-JOURNAL CANADIEN D OPHTALMOLOGIE, 2010, 45 (05): : 489 - 493