Wrist blood pressure overestimates blood pressure measured at the upper arm

被引:32
|
作者
Palatini, P [1 ]
Longo, D [1 ]
Toffanin, G [1 ]
Bertolo, O [1 ]
Zaetta, V [1 ]
Pessina, AC [1 ]
机构
[1] Univ Padua, Dept Clin & Expt Med, Padua, Italy
关键词
wrist; forearm; upper arm; blood pressure; measurement; distension;
D O I
10.1097/00126097-200404000-00004
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
Objective Whether blood pressure (BP) measured at the wrist differs from blood pressure measured at the arm is not well known. The aim of this study was to compare the BP readings obtained at the arm with those obtained at the forearm and to assess whether the wrist-arm discrepancies were related to subjects' clinical characteristics. Methods We measured blood pressure at the forearm and at the upper arm in 85 subjects using conventional sphygmomanometry. Wrist-arm blood pressure discrepancies were assessed in relation to gender, age, body mass index, skin-fold thickness, arm size, blood pressure level, and arterial compliance measured with the HDI/Pulsewave CR-2000. Results Blood pressure measured at the wrist consistently overestimated blood pressure taken at the arm with a mean ( +/- SD) discrepancy of 8.2 +/- 9.7/9.2 +/- 6.4 mmHg. The systolic blood pressure differences were greater in men than in women (p=0.006) and, among the men, varied according to arm adiposity (positive association, p=0.01). In men, diastolic blood pressure differences correlated with diastolic blood pressure level (negative association, p=0.01). Among the women, only age (p=0.04) was a significant Positive independent predictor of the wrist-arm diastolic BP differences. Conclusions These results indicate that forearm blood pressure measurement markedly overestimates upper arm blood pressure and that the between-site difference may vary from subject to subject. Wrist blood pressure measurement is not a valid alternative to traditional measurement at the arm and its use should be discouraged. (C) 2004 Lippincott Williams Wilkins.
引用
收藏
页码:77 / 81
页数:5
相关论文
共 50 条
  • [21] Comparison of femoral arterial blood pressure with radial arterial blood pressure and noninvasive upper arm blood pressure in the reperfusion period during liver transplantation
    Shin, B. S.
    Kim, G. S.
    Ko, J. S.
    Gwak, M. S.
    Yang, M.
    Kim, C. S.
    Hahm, T. S.
    Lee, S.-K.
    TRANSPLANTATION PROCEEDINGS, 2007, 39 (05) : 1326 - 1328
  • [22] Influence of Age on Upper Arm Cuff Blood Pressure Measurement
    Picone, Dean S.
    Schultz, Martin G.
    Otahal, Petr
    Black, J. Andrew
    Bos, Willem J.
    Chen, Chen-Huan
    Cheng, Hao-Min
    Cremer, Antoine
    Dwyer, Nathan
    Fonseca, Ricardo
    Hughes, Alun D.
    Kim, Hack-Lyoung
    Lacy, Peter S.
    Laugesen, Esben
    Ohte, Nobuyuki
    Omboni, Stefano
    Ott, Christian
    Pereira, Telmo
    Pucci, Giacomo
    Roberts-Thomson, Philip
    Rossen, Niklas B.
    Schmieder, Roland E.
    Sueta, Daisuke
    Takazawa, Kenji
    Wang, Jiguang
    Weber, Thomas
    Westerhof, Berend E.
    Williams, Bryan
    Yamada, Hirotsugu
    Yamamoto, Eiichiro
    Sharman, James E.
    HYPERTENSION, 2020, 75 (03) : 844 - 850
  • [23] An evaluation of non-invasive blood pressure (NIBP) monitoring on the wrist: Comparison with upper arm NIBP measurement
    Emerick, DR
    ANAESTHESIA AND INTENSIVE CARE, 2002, 30 (01) : 43 - 47
  • [24] Relationship between blood pressure obtained from the upper arm with a cuff-type sphygmomanometer and central blood pressure measured with a catheter-tipped micromanometer
    Ohte, Nobuyuki
    Saeki, Tomoaki
    Miyabe, Hiromichi
    Sakata, Seichiro
    Mukai, Seiji
    Hayano, Junichiro
    Niki, Kiyomi
    Sugawara, Motoaki
    Kimura, Genjiro
    HEART AND VESSELS, 2007, 22 (06) : 410 - 415
  • [25] Relationship between blood pressure obtained from the upper arm with a cuff-type sphygmomanometer and central blood pressure measured with a catheter-tipped micromanometer
    Nobuyuki Ohte
    Tomoaki Saeki
    Hiromichi Miyabe
    Seichiro Sakata
    Seiji Mukai
    Junichiro Hayano
    Kiyomi Niki
    Motoaki Sugawara
    Genjiro Kimura
    Heart and Vessels, 2007, 22 : 410 - 415
  • [26] Correlation between electronic wrist blood pressure with auscultatory blood pressure in children
    Oztorun, Z. Yilmaz
    EUROPEAN REVIEW FOR MEDICAL AND PHARMACOLOGICAL SCIENCES, 2023, 27 (16) : 7693 - 7699
  • [27] INDIRECT SPHYGMOMANOMETRY OVERESTIMATES BLOOD-PRESSURE IN THE ELDERLY
    HLA, KM
    VOKATY, KA
    FEUSSNER, JR
    CLINICAL RESEARCH, 1983, 31 (05): : A832 - A832
  • [28] Blood pressure measurements in the ankle are not equivalent to blood pressure measurements in the arm
    Goldstein, L. N.
    Wells, M.
    Sliwa, K.
    SAMJ SOUTH AFRICAN MEDICAL JOURNAL, 2014, 104 (12): : 869 - 869
  • [29] Blood pressure measurement Ramifications from how blood pressure is measured?
    Leopold, David
    BRITISH MEDICAL JOURNAL, 2011, 342
  • [30] COULD PLACEMENT OF THE BLOOD PRESSURE CUFF OVER THE ELBOW RATHER THAN THE UPPER ARM INCREASE THE ACCURACY OF AUTOMATIC BLOOD PRESSURE MACHINES IN THE DETECTION OF SYSTOLIC BLOOD PRESSURE?
    Pollard, Charlotte
    Naveed, Umna
    Lewis, Philip S.
    Holland, Maureen
    HEART, 2015, 101 : A84 - A84