Assessment of a Virtual Functional Prototyping Process for the Rapid Manufacture of Passive-Dynamic Ankle-Foot Orthoses

被引:53
|
作者
Schrank, Elisa S. [1 ]
Hitch, Lester [2 ]
Wallace, Kevin [2 ]
Moore, Richard [2 ]
Stanhope, Steven J. [3 ]
机构
[1] Univ Delaware, Dept Kinesiol & Appl Physiol, Newark, DE 19711 USA
[2] US Army Edgewood Chem Biol Ctr, Adv Design & Mfg Div, Edgewood, MD 21010 USA
[3] Univ Delaware, Dept Kinesiol & Appl Physiol, Biomech & Movement Sci Interdisciplinary Program, Dept Mech Engn,Dept Biomed Engn, Newark, DE 19711 USA
基金
美国国家科学基金会;
关键词
ankle; computer-aided design; customization; finite element analysis; fused deposition modeling; orthosis; orthotics; rehabilitation; bending stiffness; virtual functional prototyping; PROSTHETIC SOCKET; MECHANICAL PROPERTY; STIFFNESS; ORTHOTICS; CUSTOMIZATION; POLYPROPYLENE; FABRICATION; TECHNOLOGY; ACCURACY; DESIGN;
D O I
10.1115/1.4024825
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Passive-dynamic ankle-foot orthosis (PD-AFO) bending stiffness is a key functional characteristic for achieving enhanced gait function. However, current orthosis customization methods inhibit objective premanufacture tuning of the PD-AFO bending stiffness, making optimization of orthosis function challenging. We have developed a novel virtual functional prototyping (VFP) process, which harnesses the strengths of computer aided design (CAD) model parameterization and finite element analysis, to quantitatively tune and predict the functional characteristics of a PD-AFO, which is rapidly manufactured via fused deposition modeling (FDM). The purpose of this study was to assess the VFP process for PD-AFO bending stiffness. A PD-AFO CAD model was customized for a healthy subject and tuned to four bending stiffness values via VFP. Two sets of each tuned model were fabricated via FDM using medical-grade polycarbonate (PC-ISO). Dimensional accuracy of the fabricated orthoses was excellent (average 0.51 +/- 0.39 mm). Manufacturing precision ranged from 0.0 to 0.74 Nm/deg (average 0.30 +/- 0.36 Nm/deg). Bending stiffness prediction accuracy was within 1 Nm/deg using the manufacturer provided PC-ISO elastic modulus (average 0.48 +/- 0.35 Nm/deg). Using an experimentally derived PC-ISO elastic modulus improved the optimized bending stiffness prediction accuracy (average 0.29 +/- 0.57 Nm/deg). Robustness of the derived modulus was tested by carrying out the VFP process for a disparate subject, tuning the PD-AFO model to five bending stiffness values. For this disparate subject, bending stiffness prediction accuracy was strong (average 0.20 +/- 0.14 Nm/deg). Overall, the VFP process had excellent dimensional accuracy, good manufacturing precision, and strong prediction accuracy with the derived modulus. Implementing VFP as part of our PD-AFO customization and manufacturing framework, which also includes fit customization, provides a novel and powerful method to predictably tune and precisely manufacture orthoses with objectively customized fit and functional characteristics.
引用
收藏
页数:7
相关论文
共 50 条
  • [21] Experimental comparisons of passive and powered ankle-foot orthoses in individuals with limb reconstruction
    Esposito, Elizabeth Russell
    Schmidtbauer, Kelly A.
    Wilken, Jason M.
    [J]. JOURNAL OF NEUROENGINEERING AND REHABILITATION, 2018, 15
  • [22] Experimental comparisons of passive and powered ankle-foot orthoses in individuals with limb reconstruction
    Elizabeth Russell Esposito
    Kelly A. Schmidtbauer
    Jason M. Wilken
    [J]. Journal of NeuroEngineering and Rehabilitation, 15
  • [23] Digitizing the manual plaster casting process to produce ankle-foot orthoses
    Wang, Joyce
    Lillia, Jonathon
    Bi, Lei
    Kim, Jinman
    Burns, Joshua
    Cheng, Tegan
    [J]. JOURNAL OF THE PERIPHERAL NERVOUS SYSTEM, 2021, 26 (03) : 319 - 319
  • [24] Gait assessment of fixed ankle-foot orthoses in children with spastic diplegia
    Abel, MF
    Juhl, GA
    Vaughan, CL
    Damiano, DL
    [J]. ARCHIVES OF PHYSICAL MEDICINE AND REHABILITATION, 1998, 79 (02): : 126 - 133
  • [25] Dimensional accuracy of ankle-foot orthoses constructed by rapid customization and manufacturing framework
    Schrank, Elisa S.
    Stanhope, Steven J.
    [J]. JOURNAL OF REHABILITATION RESEARCH AND DEVELOPMENT, 2011, 48 (01): : 31 - 42
  • [26] Effects of dynamic ankle-foot orthoses on standing in children with severe spastic diplegia
    Naslund, Annika
    Jesinkey, Kristina
    Hirschfeld, Helga
    von Wendt, Lennart
    Sundelin, Gunnevi
    [J]. INTERNATIONAL JOURNAL OF THERAPY AND REHABILITATION, 2005, 12 (05): : 200 - 206
  • [27] Design and manufacture of a custom ankle-foot orthoses using traditional manufacturing and fused deposition modeling
    Abdalsadah, Falah Hasan
    Hasan, Faisal
    Murtaza, Qasim
    Khan, Abid Ali
    [J]. PROGRESS IN ADDITIVE MANUFACTURING, 2021, 6 (03) : 555 - 570
  • [28] A modified passive-dynamic ankle–foot orthosis: can it prevent amputation and arthrodesis in patients with ankle–foot trauma?
    Niels Jonkergouw
    Loes G. M. de Kruijff
    Rogier E. G. Bongers
    Michiel W. Swaan
    Herman R. Holtslag
    Alfred van der Meer
    Peter van der Wurff
    [J]. Archives of Orthopaedic and Trauma Surgery, 2022, 142 : 2719 - 2726
  • [29] Passive-dynamic ankle-foot orthosis replicates soleus but not gastrocnemius muscle function during stance in gait: Insights for orthosis prescription
    Arch, Elisa S.
    Stanhope, Steven J.
    Higginson, Jill S.
    [J]. PROSTHETICS AND ORTHOTICS INTERNATIONAL, 2016, 40 (05) : 606 - 616
  • [30] Passive-dynamic ankle-foot orthosis improves medium-term clinical outcomes after severe lower extremity trauma
    Ladlow, Peter
    Bennett, N.
    Phillip, R.
    Dharm-Datta, S.
    McMenemy, L.
    Bennett, A. N.
    [J]. JOURNAL OF THE ROYAL ARMY MEDICAL CORPS, 2019, 165 (05) : 330 - 337