Validation of the A&D wrist-cuff UB-511 (UB-512) device for self-measurement of blood pressure

被引:4
|
作者
Angeli, Fabio [1 ]
Sardone, Mariagrazia [1 ]
Angeli, Enrica [1 ]
Repaci, Salvatore [1 ]
Gattobigio, Roberto [1 ]
Verdecchia, Paolo [1 ]
机构
[1] R Silvestrini Hosp, Dept Cardiol, Clin Res Unit Prevent Cardiol, I-06156 Perugia, Italy
关键词
hypertension; home blood pressure measurement; wrist; validation; device;
D O I
10.1097/01.mbp.0000217995.12215.f2
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
Objectives To determine the accuracy of the A&D UB-511 (UB-512) oscillometric wrist-cuff device for self-measurement of blood pressure, the only difference between the two devices being the size of storage memory. Methods Device evaluation was performed according to the modified British Hypertension Society protocol released in 1993. Eighty-five study participants with characteristics outlined in the British Hypertension Society protocol were recruited among those attending our out-patient clinic. The device was evaluated according to the various steps of the protocol. The non-dominant arm was used for blood pressure measurement. To maintain the wrist at cardiac level during validation, the arm was kept horizontal at the mid-sternum level and supported by a soft table. The wrist was kept extended. Sequential readings were taken for the main validation test. Outcome was classified according to the criteria of the British Hypertension Society recommendations, which are based on four strata of accuracy differing from the mercury standard by 5, 10 and 15 mmHg, or more. Results The device achieved a British Hypertension Society grade B for systolic and a grade B for diastolic blood pressure. The device tended to overestimate arm blood pressure, the mean difference (+/- 1 SD) between device and observers being 4.3 +/- 8.7 mmHg for systolic blood pressure and 3.7 +/- 8.1 mmHg for diastolic blood pressure for observer 2, and 4.4 +/- 8.6 mmHg for systolic blood pressure and 3.8 +/- 7.9 mmHg for diastolic blood pressure for observer 1. In a logistic regression analysis, age Was the sole predictor of an achieved difference between device and mercury column by 5 mmHg or less (hazard ratio 1.020; 95% confidence interval 1.003-1.04; P= 0.024). Conclusions These data show that the A&D UB-511 (UB-512) device satisfies the British Hypertension Society recommendations with a grade B/B. The device tends to overestimate cuff blood pressure and its accuracy increases with age.
引用
收藏
页码:349 / 354
页数:6
相关论文
共 39 条
  • [31] Validation of the A&D UM-201 device for office blood pressure measurement according to the European Society of Hypertension International Protocol Revision 2010
    Fania, Claudio
    Albertini, Federica
    Palatini, Paolo
    BLOOD PRESSURE MONITORING, 2017, 22 (04) : 234 - 237
  • [32] Validation of the A&D BP UA-651 device for home blood pressure measurement according to the European Society of Hypertension International Protocol revision 2010
    Benetti, Elisabetta
    Fania, Claudio
    Palatini, Paolo
    BLOOD PRESSURE MONITORING, 2014, 19 (01) : 50 - 53
  • [33] Validation of the A&D BP UA-1200BLE device for home blood pressure measurement according to the ISO 81060-2:2013 standard
    Kario, Kazuomi
    Saito, Kimiyo
    Sato, Keiko
    Hamasaki, Haruna
    Suwa, Hiromi
    Okura, Ayako
    Hoshide, Satoshi
    BLOOD PRESSURE MONITORING, 2018, 23 (06) : 312 - 314
  • [34] VALIDATION OF THE VITA-D1 NONINVASIVE CONTINUOUS BLOOD PRESSURE MONITOR WITH PHOTOPLETHYSMOGRAPHY-ONLY TECHNOLOGY FOR SELF-MEASUREMENT AND CLINIC USE
    Ang, Jianning Zh
    Tong, Xinzhu
    Qiu, Yumin
    Tao, Jun
    JOURNAL OF HYPERTENSION, 2021, 39 : E114 - E114
  • [35] Validation of the Visomat Comfort 20/40 oscillometric device designed for self-home blood pressure measurement using a single cuff according to the International Protocol
    Stergiou, G.
    Tzamouranis, D.
    Nasothimiou, E.
    Protogerou, A.
    JOURNAL OF HYPERTENSION, 2008, 26 : S488 - S488
  • [36] Validation of an automatic device for the self-measurement of blood pressure in sitting and supine positions according to the ANSI/AAMI/ISO81060-2:2013 guidelines: the Omron HEM-9700T
    Kuwabara, Mitsuo
    Harada, Kanako
    Hishiki, Yukiko
    Kario, Kazuomi
    BLOOD PRESSURE MONITORING, 2019, 24 (03) : 146 - 150
  • [37] Validation of two automatic devices, Omron HEM-6232T and HEM-6181, for self-measurement of blood pressure at the wrist according to the ANSI/AAMI/ISO 81060-2:2013 protocol and the European Society of Hypertension International Protocol revision 2010
    Saito, Kanako
    Hishiki, Yukiko
    Takahashi, Hakuo
    VASCULAR HEALTH AND RISK MANAGEMENT, 2019, 15 : 47 - 55
  • [38] Validation of the Omron HEM-7320-LA, upper arm blood pressure monitor with Intelli Wrap Technology Cuff HEM-FL1 for self-measurement and clinic use according to the European Society of Hypertension International Protocol revision 2010 in the Mexican population
    Grover-Paez, Fernando
    Cardona-Munoz, Ernesto G.
    Cardona-Mueller, David
    Guzman-Saldivar, Victor H.
    Rodriguez-De la Cerda, Mariana
    Jimenez-Cazarez, Mayra B.
    Totsuka-Sutto, Sylvia E.
    Alanis-Sanchez, Guillermo A.
    Ramos-Becerra, Carlos G.
    BLOOD PRESSURE MONITORING, 2017, 22 (06) : 375 - 378