Validation of the A&D BP UA-1200BLE device for home blood pressure measurement according to the ISO 81060-2:2013 standard

被引:3
|
作者
Kario, Kazuomi [1 ]
Saito, Kimiyo [1 ]
Sato, Keiko [1 ]
Hamasaki, Haruna [1 ]
Suwa, Hiromi [1 ]
Okura, Ayako [1 ]
Hoshide, Satoshi [1 ]
机构
[1] Jichi Med Univ, Sch Med, Div Cardiovasc Med, Dept Med, 3311-1 Yakushiji, Shimotsuke, Tochigi 3290498, Japan
关键词
blood pressure; device validation; home blood pressure monitoring; self-blood pressure monitoring; wireless monitoring; HYPERTENSION PRACTICE GUIDELINES; RANDOMIZED CONTROLLED-TRIAL; EUROPEAN-SOCIETY; MANAGEMENT; CARE;
D O I
10.1097/MBP.0000000000000347
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
ObjectiveThe aim of this study was to validate the UA-1200BLE device using the ISO 81060-2:2013 standard.Materials and methodsIndividuals aged more than 12 years were recruited from among the outpatients and volunteers of Jichi Medical University (Tochigi, Japan). Eighty-five patients were included in the final analyses. Validation was performed using the same-arm sequence protocol as per the ISO 81060-2:2013 standard.ResultsThe device performed well against the standard; the mean and SD values for the differences between device and observed systolic and diastolic blood pressure values fulfilled both criterion 1 and criterion 2 of the standard (criterion 1: -2.626.89 and -0.90 +/- 6.63mmHg, respectively; criterion 2: -2.65 +/- 5.36 and -0.91 +/- 6.12mmHg, respectively). The Bland-Altman plots did not show any systematic variation in the error.ConclusionThe UA-1200BLE device had a high level of accuracy and fulfilled all ISO 81060-2:2013 standard requirements in adults with a wide range of arm circumference measurements. It is therefore suitable as a wireless blood pressure monitoring tool in patients with hypertension.
引用
收藏
页码:312 / 314
页数:3
相关论文
共 50 条
  • [1] Validation of the TM-2441 ambulatory blood pressure measurement device according to the ISO 81060-2:2013 standard
    Kario, Kazuomi
    Hoshide, Satoshi
    Saito, Kimiyo
    Sato, Keiko
    Hamasaki, Haruna
    Suwa, Hiromi
    Tomitani, Naoko
    BLOOD PRESSURE MONITORING, 2019, 24 (01) : 38 - 41
  • [2] Validation of the A&D UM-212BLE monitor according to ISO 81060-2, 2018: a device with clinically important programmability
    Alpert, Bruce S.
    BLOOD PRESSURE MONITORING, 2023, 28 (02) : 113 - 115
  • [3] Validation of the A&D UA-651 Plus/UA-651SL Plus automated sphygmomanometer according to the ISO 81060-2, 2018 and ISO 81060-2 Amendment 1, 2020, which resulted in the currently pending Amendment to Amendment 1
    Alpert, Bruce S.
    JOURNAL OF CLINICAL HYPERTENSION, 2022, 24 (04): : 513 - 518
  • [4] Validation of the Aktiia blood pressure cuff for clinical use according to the ANSI/AAMI/ISO 81060-2:2013 protocol
    Alexandre, Jeremy
    Tan, Kevin
    Almeida, Tiago P.
    Sola, Josep
    Alpert, Bruce S.
    Shah, Jay
    BLOOD PRESSURE MONITORING, 2023, 28 (02) : 109 - 112
  • [5] Validation of the A&D BP UA-651 device for home blood pressure measurement according to the European Society of Hypertension International Protocol revision 2010
    Benetti, Elisabetta
    Fania, Claudio
    Palatini, Paolo
    BLOOD PRESSURE MONITORING, 2014, 19 (01) : 50 - 53
  • [6] Validation of A&D UA-85X device for blood pressure measurement
    Bonso, Elisa
    Ragazzo, Fabio
    Palatini, Paolo
    BLOOD PRESSURE MONITORING, 2008, 13 (06) : 339 - 341
  • [7] Validation of the RisingSun RS-651 Blood Pressure Monitor Based on Auscultation in Adults According to the ANSI/AAMI/ISO 81060-2:2013 Standard
    She, Jin
    Guan, Xizhou
    Liu, Yanyong
    Xiang, Haiyan
    JOURNAL OF CLINICAL HYPERTENSION, 2016, 18 (12): : 1279 - 1283
  • [8] Validation of a smartphone auscultatory blood pressure kit Accutension XYZ-110 in adults according to the ANSI/AAMI/ISO 81060-2:2013 standard
    Chu, Guang
    Zhang, Zhi
    Xu, Mengdan
    Huang, Daini
    Dai, Qiuyan
    BLOOD PRESSURE MONITORING, 2017, 22 (05) : 290 - 294
  • [9] Validation of the ABPMpro ambulatory blood pressure monitor in the general population according to AAMI/ESH/ISO Universal Standard (ISO 81060-2:2018)
    Roth, Bernhard
    Bothe, Tomas Lucca
    Patzak, Andreas
    Pilz, Niklas
    BLOOD PRESSURE MONITORING, 2023, 28 (03) : 158 - 162
  • [10] Validation of A&D UA-767 device for the self-measurement of blood pressure
    Rogoza, AN
    Pavlova, TS
    Sergeeva, MV
    BLOOD PRESSURE MONITORING, 2000, 5 (04) : 227 - 231